Can we use pulsed fluoroscopy to decrease the radiation dose during video fluoroscopic feeding studies in children?

被引:28
作者
Cohen, M. D. [1 ]
机构
[1] Indiana Univ, Dept Radiol, James Whitcomb Riley Hosp Children, Sch Med, Indianapolis, IN 46202 USA
关键词
EXPOSURE;
D O I
10.1016/j.crad.2008.07.011
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
AIM: To investigate whether it is possible to reduce the radiation dose during video fluoroscopic feeding studies below the current 30 frames/s (continuous fluoroscopy). METHODS: Ten consecutive children who had supraglottic penetration while swallowing barium were evaluated as part of a video fluoroscopic feeding study. All fluoroscopic studies were performed with a pulse rate of 30 frames/s. Frame by frame analysis was performed of the first episode of penetration in each patient to determine on how many image frames the penetration could be detected. RESULTS: Supraglottic penetration occurred very rapidly. In seven of the 10 patients, full-depth penetration was only seen on one image frame. In no patient was the full-depth penetration seen in greater than two imaging frames. CONCLUSION: Decreasing the fluoroscopic pulse rate cannot be used as a method of decreasing radiation dose during performance of video fluoroscopic studies because it will potentially result in non-detection of episodes of supraglottic penetration of liquid barium. (C) 2008 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:70 / 73
页数:4
相关论文
共 11 条
[1]  
Amis E Stephen Jr, 2007, J Am Coll Radiol, V4, P272, DOI 10.1016/j.jacr.2007.03.002
[2]   Optimization of a fluoroscope to reduce radiation exposure in pediatric imaging [J].
Brown, PH ;
Thomas, RD ;
Silberberg, PJ ;
Johnson, LM .
PEDIATRIC RADIOLOGY, 2000, 30 (04) :229-235
[3]   Are we doing enough to minimize fluoroscopic radiation exposure in children? [J].
Cohen, Mervyn .
PEDIATRIC RADIOLOGY, 2007, 37 (10) :1020-1024
[4]   RADIOLOGIC ASSESSMENT OF ABNORMAL ORAL AND PHARYNGEAL PHASES OF SWALLOWING [J].
DODDS, WJ ;
LOGEMANN, JA ;
STEWART, ET .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1990, 154 (05) :965-974
[5]   PHYSIOLOGY AND RADIOLOGY OF THE NORMAL ORAL AND PHARYNGEAL PHASES OF SWALLOWING [J].
DODDS, WJ ;
STEWART, ET ;
LOGEMANN, JA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1990, 154 (05) :953-963
[6]  
Hahn H., 1997, Medicamundi, V41, P12
[7]   Reduction of radiation dose in pediatric patients using pulsed fluoroscopy [J].
Hernandez, RJ ;
Goodsitt, MM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1996, 167 (05) :1247-1253
[8]  
Nagel H. D., 1997, Medicamundi, V41, P43
[9]   CLINICAL AND VIDEOFLUOROSCOPIC EVALUATION OF SWALLOWING DISORDERS [J].
OTT, DJ ;
PIKNA, LA .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1993, 161 (03) :507-513
[10]   Management of pediatric radiation dose using Philips fluoroscopy systems DoseWise: perfect image, perfect sense [J].
Stueve, Dick .
PEDIATRIC RADIOLOGY, 2006, 36 (Suppl 2) :216-220