An automated workstation for forced degradation of active pharmaceutical ingredients

被引:10
作者
Sims, JL [1 ]
Roberts, JK [1 ]
Bateman, AG [1 ]
Carreira, JA [1 ]
Hardy, MJ [1 ]
机构
[1] GlaxoSmithKline Pharmaceut R&D, Dept Analyt Chem, Harlow CM19 5AW, Essex, England
关键词
automated workstation; LC/MS; SK233; drug substance degradation; stability indicating methods; HPLC; active pharmaceutical ingredient;
D O I
10.1002/jps.10055
中图分类号
R914 [药物化学];
学科分类号
100701 ;
摘要
This article describes a system capable of performing and analysing multiple degradation experiments on drug substances as part of the process of developing stability indicating separations methods. Qualitative data are generated on the significant primary degradation processes of the drug of interest. Ten samples are refluxed with stirring in a single heating block. The robot arm is equipped with a sampling device capable of removing aliquots, during the reflux experiment, and transferring them to an HPLC injector. On-line analysis using fast HPLC with diode array and electrospray mass spectrometric detection allows identification of degradants. The methods described offer a significant time savings compared to the previously applied manual approach, and also provide data from multiple time points for each reaction. This increased knowledge about the progress of the reaction aids us in focussing efficiently on the primary degradation processes. (C) 2002 Wiley-Liss, Inc.
引用
收藏
页码:884 / 892
页数:9
相关论文
共 15 条
[1]   A versatile and cost-effective approach to automated laboratory organic synthesis [J].
Armitage, MA ;
Smith, GE ;
Veal, KT .
ORGANIC PROCESS RESEARCH & DEVELOPMENT, 1999, 3 (03) :189-195
[2]   A practical approach to accelerated process screening and optimisation [J].
Emiabata-Smith, DF ;
Crookes, DL ;
Owen, MR .
ORGANIC PROCESS RESEARCH & DEVELOPMENT, 1999, 3 (04) :281-288
[3]   KINETIC MONITORING OF ENZYMATIC-REACTIONS IN REAL-TIME BY QUANTITATIVE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
HSIEH, FYL ;
TONG, X ;
WACHS, T ;
GANEM, B ;
HENION, J .
ANALYTICAL BIOCHEMISTRY, 1995, 229 (01) :20-25
[4]  
Hsu LC, 1999, LAB ROBOTICS AUTOMAT, V11, P186, DOI 10.1002/(SICI)1098-2728(1999)11:4<186::AID-LRA2>3.0.CO
[5]  
2-N
[6]   Fully automated 96-well liquid-liquid extraction for analysis of biological samples by liquid chromatography with tandem mass spectrometry [J].
Peng, SX ;
Branch, TM ;
King, SL .
ANALYTICAL CHEMISTRY, 2001, 73 (03) :708-714
[7]   Automated 96-well SPE and LC-MS-MS for determination of protease inhibitors in plasma and cartilage tissues [J].
Peng, SX ;
King, SL ;
Bornes, DM ;
Foltz, DJ ;
Baker, TR ;
Natchus, MG .
ANALYTICAL CHEMISTRY, 2000, 72 (08) :1913-1917
[8]   USE OF LIQUID-CHROMATOGRAPHY NUCLEAR-MAGNETIC-RESONANCE SPECTROSCOPY FOR THE IDENTIFICATION OF IMPURITIES IN DRUG SUBSTANCES [J].
ROBERTS, JK ;
SMITH, RJ .
JOURNAL OF CHROMATOGRAPHY A, 1994, 677 (02) :385-389
[9]  
Sauvagnat B, 1998, RAPID COMMUN MASS SP, V12, P1034, DOI 10.1002/(SICI)1097-0231(19980815)12:15<1034::AID-RCM264>3.0.CO
[10]  
2-Y