Real-time monitoring of quality attributes by in-line Fourier transform infrared spectroscopic sensors at ultrafiltration and diafiltration of bioprocess

被引:27
作者
Wasalathanthri, Dhanuka P. [1 ]
Feroz, Hasin [1 ]
Puri, Neha [1 ]
Hung, Jessica [1 ]
Lane, Gregory [2 ]
Holstein, Melissa [1 ]
Chemmalil, Letha [1 ]
Both, Douglas [2 ]
Ghose, Sanchayita [1 ]
Ding, Julia [1 ]
Li, Zheng Jian [1 ]
机构
[1] Bristol Myers Squibb Co, Biol Proc Dev, 38 Jackson Rd, Devens, MA 01434 USA
[2] Bristol Myers Squibb Co, Engn Technol, New Brunswick, NJ USA
关键词
Fourier transform mid-infrared spectroscopy; multivariate data analysis; process analytical technology; real-time monitoring; OPPORTUNITIES; CHALLENGES; PROTEIN; PAT;
D O I
10.1002/bit.27532
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Technologies capable of monitoring product quality attributes and process parameters in real time are becoming popular due to the endorsement of regulatory agencies and also to support the agile development of biotherapeutic pipelines. The utility of vibrational spectroscopic techniques such as Fourier transform mid-infrared (Mid-IR) and multivariate data analysis (MVDA) models allows the prediction of multiple critical attributes simultaneously in real time. This study reports the use of Mid-IR and MVDA model sensors for monitoring of multiple attributes (excipients and protein concentrations) in real time (measurement frequency of every 40 s) at ultrafiltration and diafiltration (UF/DF) unit operation of biologics manufacturing. The platform features integration of fiber optic Mid-IR probe sensors to UF/DF set up at the bulk solution and through a flow cell at the retentate line followed by automated Mid-IR data piping into a process monitoring software platform with pre-loaded partial least square regression (PLS) chemometric models. Data visualization infrastructure is also built-in to the platform so that upon automated PLS prediction of excipients and protein concentrations, the results were projected in a graphical or numerical format in real time. The Mid-IR predicted concentrations of excipients and protein show excellent correlation with the offline measurements by traditional analytical methods. Absolute percent difference values between Mid-IR predicted results and offline reference assay results were <= 5% across all the excipients and the protein of interest; which shows a great promise as a reliable process analytical technology tool.
引用
收藏
页码:3766 / 3774
页数:9
相关论文
共 31 条
[1]  
[Anonymous], 2009, Pharmaceutical development Q8 (R2)
[2]   Effect of Protein and Solution Properties on the Donnan Effect During the Ultrafiltration of Proteins [J].
Bolton, Glen R. ;
Boesch, Austin W. ;
Basha, Jonida ;
LaCasse, Daniel P. ;
Kelley, Brian D. ;
Acharya, Hari .
BIOTECHNOLOGY PROGRESS, 2011, 27 (01) :140-152
[3]   In-column ATR-FTIR spectroscopy to monitor affinity chromatography purification of monoclonal antibodies [J].
Boulet-Audet, Maxime ;
Kazarian, Sergei G. ;
Byrne, Bernadette .
SCIENTIFIC REPORTS, 2016, 6
[4]  
Bruker Optics, 2015, FT IR FT NIR PROC SP
[5]   At-line mid infrared spectroscopy for monitoring downstream processing unit operations [J].
Capito, Florian ;
Skudas, Romas ;
Kolmar, Harald ;
Hunzinger, Christian .
PROCESS BIOCHEMISTRY, 2015, 50 (06) :997-1005
[6]   Matrix effects during monitoring of antibody and host cell proteins using attenuated total reflection spectroscopy [J].
Capito, Florian ;
Skudas, Romas ;
Stanislawski, Bernd ;
Kolmar, Harald .
BIOTECHNOLOGY PROGRESS, 2013, 29 (01) :265-274
[7]  
Challener CA, 2018, BIOPHARM INT, V31, P24
[8]   Mining bioprocess data: opportunities and challenges [J].
Charaniya, Salim ;
Hu, Wei-Shou ;
Karypis, George .
TRENDS IN BIOTECHNOLOGY, 2008, 26 (12) :690-699
[9]  
FDA, 2004, GUID IND PAT A FRAM
[10]  
Garidel P., 2006, BioProcess International, V4, P40