Management of humeral nonunion after the failure of locking intramedullary nails

被引:96
作者
McKee, MD
Miranda, MA
Riemer, BL
Blasier, RB
Redmond, BJ
Sims, SH
Waddell, JP
Jupiter, JB
机构
[1] ST MICHAELS HOSP, UPPER EXTREM RECONSTRUCT SERV, TORONTO, ON M5B 1W8, CANADA
[2] ALLEGHENY GEN HOSP, PITTSBURGH, PA 15212 USA
[3] UNIV TORONTO, DEPT SURG, DIV ORTHOPAED, TORONTO, ON, CANADA
[4] MASSACHUSETTS GEN HOSP, ORTHOPAED TRAUMA SERV, BOSTON, MA 02114 USA
关键词
humeral nonunion; interlocking nailing; humeral reconstruction;
D O I
10.1097/00005131-199610000-00008
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We reviewed 21 cases of humeral nonunion following the failure of ''locking'' humeral nails. The nails had been inserted as the primary operative procedure following humeral fracture in fifteen cases or after the failure of closed treatment in six cases. Reconstruction after the failure of these implants was complicated by poor bone stock and difficulty in achieving union. Although technically difficult, open reduction and internal fixation with plating and bone grafting (successful in nine of nine cases) was more consistent than exchange nailing (successful in four of 10 cases) in achieving union (p = 0.01). Two patients refused further surgical intervention. The degree of bone loss associated with a loose nail, the lack of success of exchange nailing, and the insertion site morbidity associated with humeral nail removal differentiate these nonunions from similar lower extremity problems. The degree of bone loss following failed locking nailing of the humerus is a major concern, and exchange nailing alone may not bean acceptable option to deal with this problem.
引用
收藏
页码:492 / 499
页数:8
相关论文
共 24 条
[1]   INTRAMEDULLARY STABILIZATION OF HUMERAL SHAFT FRACTURES IN PATIENTS WITH MULTIPLE TRAUMA [J].
BRUMBACK, RJ ;
BOSSE, MJ ;
POKA, A ;
BURGESS, AR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :960-970
[2]  
Dabezies E J, 1992, J Orthop Trauma, V6, P10
[3]   A BIOMECHANICAL COMPARISON OF INTRAMEDULLARY NAILING SYSTEMS FOR THE HUMERUS [J].
DALTON, JE ;
SALKELD, SL ;
SATTERWHITE, YE ;
COOK, SD .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1993, 7 (04) :367-374
[4]  
DURBIN RA, 1983, CLIN ORTHOP RELAT R, P168
[5]   PREVENTION OF INFECTION IN TREATMENT OF 1000 AND 25 OPEN FRACTURES OF LONG BONES - RETROSPECTIVE AND PROSPECTIVE ANALYSES [J].
GUSTILO, RB ;
ANDERSON, JT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1976, 58 (04) :453-458
[6]   A LOCKING NAIL FOR FRACTURES OF THE HUMERUS [J].
HABERNEK, H ;
ORTHNER, E .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (04) :651-653
[7]  
Henley M B, 1992, J Orthop Trauma, V6, P18
[8]  
HOFFMEYER RE, 1992, J ORTHOP TRAUMA, V6, P14
[9]   LOCKED INTRAMEDULLARY NAILING OF HUMERAL SHAFT FRACTURES - IMPLANT DESIGN, SURGICAL TECHNIQUE, AND CLINICAL-RESULTS [J].
INGMAN, AM ;
WATERS, DA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (01) :23-29