Mechanical properties and applications of custom-built gold AFM cantilevers

被引:8
|
作者
Kolchuzhin, Vladimir A. [1 ]
Sheremet, Evgeniya [2 ]
Bhattacharya, Kunal [3 ]
Rodriguez, Raul D. [3 ]
Paul, Soumya Deep [2 ]
Mehner, Jan [1 ]
Hietschold, Michael [2 ]
Zahn, Dietrich R. T. [3 ]
机构
[1] Tech Univ Chemnitz, Microsyst & Biomed Engn, D-09107 Chemnitz, Germany
[2] Tech Univ Chemnitz, Solid Surfaces Anal Grp, D-09107 Chemnitz, Germany
[3] Tech Univ Chemnitz, Semiconduct Phys, D-09107 Chemnitz, Germany
关键词
Cantilevers; scanning probe microscopy (SPM); Atomic force microscopy (AFM); Tip-enhanced Raman spectroscopy (TERS); Current-sensing AFM (CSAFM); Kelvin Probe force microscopy (KPFM); Finite element method simulations (FEM simulations); ATOMIC-FORCE MICROSCOPE; SPRING CONSTANTS;
D O I
10.1016/j.mechatronics.2016.05.015
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Silicon and silicon nitride metal-coated cantilevers have been in use for a long time in several scanning probe microscopy applications that require electrically conductive probes. However, conventional metalized cantilevers present several drawbacks such as limited life-time of the metal layer due to wear, and increased tip radius. This work focuses on monolithic metallic cantilevers developed in order to overcome the limitations of conventional metalized atomic force microscopy (AFM) probes. These custom-made cantilevers are designed for several applications such as in current-sensing AFM (CSAFM), Kelvin probe force microscopy (KPFM), and tip-enhanced Raman spectroscopy (TERS). Determination of the dynamic and static mechanical properties of these cantilevers in a non-destructive way is reported here. Key parameters under investigation are the cantilever spring constant and the frequency response using finite element method (FEM). Gold cantilevers are selected for this study, which allow optimizing the design and the process of developing these metallic cantilevers with parameters engineered for the applications mentioned above. This work contributes to the establishment and applicability of custom-made probes in advanced scanning probe microscopy methods and their performance understanding using computer simulations. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:281 / 286
页数:6
相关论文
共 50 条
  • [1] Custom-built integrated circuits for sophisticated applications
    Hedlund, Bjorn
    Ingman, Klas
    ABB Review, 1994, (03): : 19 - 22
  • [2] Custom-built solutions
    不详
    CHEMISTRY & INDUSTRY, 2006, (16) : 36 - 36
  • [3] CUSTOM-BUILT BULLETIN BOARDS
    KEAVENEY, C
    PERSONAL COMPUTING, 1987, 11 (08): : 91 - &
  • [4] Custom-built noise enclosures
    不详
    JOURNAL OF THE SOCIETY OF LEATHER TECHNOLOGISTS AND CHEMISTS, 1996, 80 (03): : 98 - 99
  • [5] CUSTOM-BUILT PRACTICE PREMISES
    KERR, IWF
    BRITISH MEDICAL JOURNAL, 1969, 1 (5646): : 783 - &
  • [6] MATRICES WITH CUSTOM-BUILT EIGENSPACES
    GALVIN, WP
    AMERICAN MATHEMATICAL MONTHLY, 1984, 91 (05): : 308 - 309
  • [7] Custom-built integral dosing package
    Water Services, 1996, 100 (1200):
  • [8] Custom-built system controls NOx
    不详
    CHEMICAL PROCESSING, 2001, 64 (03): : 62 - 62
  • [9] CUSTOM-BUILT DEER TAKE TO THE HILLS
    TUDGE, C
    NEW SCIENTIST, 1987, 114 (1555) : 28 - 28
  • [10] Characterising a Custom-Built Radio Frequency PECVD Reactor to Vary the Mechanical Properties of TMDSO Films
    Radjef, Racim
    Jarvis, Karyn L.
    Hall, Colin
    Ang, Andrew
    Fox, Bronwyn L.
    McArthur, Sally L.
    MOLECULES, 2021, 26 (18):