Performance of mechanical ventilators at the patient's home: a multicentre quality control study

被引:32
作者
Farre, R
Navajas, D
Prats, E
Marti, S
Guell, R
Montserrat, JM
Tebe, C
Escarrabill, J
机构
[1] Univ Barcelona, Fac Med, Unitat Biofis & Bioengn, IDIBAPS, Barcelona 08036, Spain
[2] Hosp Univ Bellvitge, UFISS Resp, Barcelona, Spain
[3] Hosp Gen Valle Hebron, Barcelona, Spain
[4] Hosp Santa Creu & Sant Pau, E-08025 Barcelona, Spain
[5] Hosp Clin Barcelona, E-08036 Barcelona, Spain
[6] Agencia Avaluacio Tecnol & Recerca Med Catalunya, Barcelona, Spain
关键词
D O I
10.1136/thx.2005.052647
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: Quality control procedures vary considerably among the providers of equipment for home mechanical ventilation (HMV). Methods: A multicentre quality control survey of HMV was performed at the home of 300 patients included in the HMV programmes of four hospitals in Barcelona. It consisted of three steps: (1) the prescribed ventilation settings, the actual settings in the ventilator control panel, and the actual performance of the ventilator measured at home were compared; (2) the different ventilator alarms were tested; and (3) the effect of differences between the prescribed settings and the actual performance of the ventilator on non-programmed readmissions of the patient was determined. Results: Considerable differences were found between actual, set, and prescribed values of ventilator variables; these differences were similar in volume and pressure preset ventilators. The percentage of patients with a discrepancy between the prescribed and actual measured main ventilator variable (minute ventilation or inspiratory pressure) of more than 20% and 30% was 13% and 4%, respectively. The number of ventilators with built in alarms for power off, disconnection, or obstruction was 225, 280 and 157, respectively. These alarms did not work in two (0.9%), 52 (18.6%) and eight (5.1%) ventilators, respectively. The number of non-programmed hospital readmissions in the year before the study did not correlate with the index of ventilator error. Conclusions: This study illustrates the current limitations of the quality control of HMV and suggests that improvements should be made to ensure adequate ventilator settings and correct ventilator performance and ventilator alarm operation.
引用
收藏
页码:400 / 404
页数:5
相关论文
共 12 条
[1]   Performance characteristics of 10 home mechanical ventilators in pressure-support mode - A comparative bench study [J].
Battisti, A ;
Tassaux, D ;
Janssens, JP ;
Michotte, JB ;
Jaber, S ;
Jolliet, P .
CHEST, 2005, 127 (05) :1784-1792
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Home mechanical ventilatory support in patients with restrictive ventilatory disorders: A 48-year experience [J].
Duiverman, ML ;
Bladder, G ;
Meinesz, AF ;
Wijkstra, PJ .
RESPIRATORY MEDICINE, 2006, 100 (01) :56-65
[4]   Effects of different ventilator settings on sleep and inspiratory effort in patients with neuromuscular disease [J].
Fanfulla, F ;
Delmastro, M ;
Berardinelli, A ;
Lupo, ND ;
Nava, S .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2005, 172 (05) :619-624
[5]   Quality control of equipment in home mechanical ventilation: a European survey [J].
Farre, R ;
Lloyd-Owen, SJ ;
Ambrosino, N ;
Donaldson, G ;
Escarrabill, J ;
Fauroux, B ;
Robert, D ;
Schoenhofer, B ;
Simonds, A ;
Wedzicha, JA .
EUROPEAN RESPIRATORY JOURNAL, 2005, 26 (01) :86-94
[6]   Quality control of mechanical ventilation at the patient's home [J].
Farré, R ;
Giró, E ;
Casolivé, V ;
Navajas, D ;
Escarrabill, J .
INTENSIVE CARE MEDICINE, 2003, 29 (03) :484-486
[7]   Patterns of home mechanical ventilation use in Europe: results from the Eurovent survey [J].
Lloyd-Owen, SJ ;
Donaldson, GC ;
Ambrosino, N ;
Escarabill, J ;
Farre, R ;
Fauroux, B ;
Robert, D ;
Schoenhofer, B ;
Simonds, AK ;
Wedzicha, JA .
EUROPEAN RESPIRATORY JOURNAL, 2005, 25 (06) :1025-1031
[8]   Inaccuracy of tidal volume delivered by home mechanical ventilators [J].
Lofaso, F ;
Fodil, R ;
Lorino, H ;
Leroux, K ;
Quintel, A ;
Leroy, A ;
Harf, A .
EUROPEAN RESPIRATORY JOURNAL, 2000, 15 (02) :338-341
[9]   Frequency, causes, and outcome of home ventilator failure [J].
Srinivasan, S ;
Doty, SM ;
White, TR ;
Segura, VH ;
Jansen, MT ;
Ward, SLD ;
Keens, TG .
CHEST, 1998, 114 (05) :1363-1367
[10]   Rationale for the use of non-invasive ventilation in chronic ventilatory failure [J].
Turkington, PM ;
Elliott, MW .
THORAX, 2000, 55 (05) :417-423