Patient Radiation Dose Reduction during Transarterial Chemoembolization Using a Novel X-Ray Imaging Platform

被引:5
作者
Kohlbrenner, Ryan [1 ]
Kolli, Pallav [1 ]
Taylor, Andrew G. [1 ]
Kohi, Maureen P. [1 ]
Fidelman, Nicholas [1 ]
LaBerge, Jeanne M. [1 ]
Kerlan, Robert K. [1 ]
Agarwal, Vishal K. [1 ]
Lehrman, Evan D. [1 ]
Nanavati, Sujal [1 ]
Avrin, David E. [1 ]
Gould, Robert [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94143 USA
关键词
INTERVENTIONAL RADIOLOGY PROCEDURES; DIGITAL-SUBTRACTION-ANGIOGRAPHY; AREA PRODUCT; FLUOROSCOPY; TECHNOLOGY; QUALITY; TIME; IR;
D O I
10.1016/j.jvir.2015.06.016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate radiation dose reduction in patients undergoing transarterial chemoembolization with the use of a new image acquisition and processing platform. Materials and 'Methods: Radiation-dose data were obtained from 176 consecutive chemoembolization procedures in 135 patients performed in a single angiography suite. From January 2013 through October 2013, 85 procedures were, performed by using our institution's standard fluoroscopic settings. After upgrading the x-ray fluoroscopy system with an image acquisition and processing platform designed to reduce image noise and reduce skin entrance dose, 91 chemoembolization procedures were performed from November 2013 through December 2014. Cumulative dose area product (CDAP), cumulative air kerma (CAK), and total fluoroscopy time were recorded for each procedure. Image quality was assessed by three interventional radiologists blinded to the x-ray acquisition platform used. Results: Patient radiation dose indicators were significantly lower for chemoembolization procedures perfauned with the novel imaging platform. Mean CDAP decreased from 3,033.2 dGy.cm(2) (range, 600.3-9,404.1 dGy.cm(2)) to 1,640.1 dGy.cm(2) (range, 278.6-6,779.9 dGy.cm(2); 45.9% reduction; P<.00001). Mean CAK 'decreased from 1,445.4 mGy (range, 303.6-5,233.7 mGy) to 971.7 mGy (range, 144.2-3,512.0 mGy; 32.8% reduction; P<.0001). A 20.3% increase in mean total fluoroscopy time was noted after upgrading the imaging platform, but blinded analysis of the image quality revealed no significant degradation. Conclusions: Although a small increase in fluoroscopy time was observed, a significant reduction in patient radiation dose was achieved by using the optimized imaging platform, without image quality degradation.
引用
收藏
页码:1331 / 1338
页数:8
相关论文
共 50 条
  • [21] X-ray dose reduction in fluoroscopically guided electrophysiology procedures
    Davies, AG
    Cowen, AR
    Kengyelics, SM
    Moore, J
    Pepper, C
    Cowan, C
    Sivanathan, MU
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (03): : 262 - 271
  • [22] Radiation dose from 3D rotational X-ray imaging: organ and effective dose with conversion factors
    Kim, Sangroh
    Sopko, David
    Toncheva, Greta
    Enterline, David
    Keijzers, Bruno
    Yoshizumi, Terry T.
    RADIATION PROTECTION DOSIMETRY, 2012, 150 (01) : 50 - 54
  • [23] Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studies
    Habibzadeh, M. A.
    Ay, M. R.
    Asl, A. R. Kamali
    Ghadiri, H.
    Zaidi, H.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (03): : 191 - 199
  • [24] Radiation dose reduction with frame rate conversion in X-ray fluoroscopic imaging systems with flat panel detector: basic study and clinical retrospective analysis
    Noriyuki Sakai
    Katsuyuki Tabei
    Jiro Sato
    Toshikazu Imae
    Yuichi Suzuki
    Shigeharu Takenaka
    Keiichi Yano
    Osamu Abe
    European Radiology, 2019, 29 : 985 - 992
  • [25] Radiation dose reduction with frame rate conversion in X-ray fluoroscopic imaging systems with flat panel detector: basic study and clinical retrospective analysis
    Sakai, Noriyuki
    Tabei, Katsuyuki
    Sato, Jiro
    Imae, Toshikazu
    Suzuki, Yuichi
    Takenaka, Shigeharu
    Yano, Keiichi
    Abe, Osamu
    EUROPEAN RADIOLOGY, 2019, 29 (02) : 985 - 992
  • [26] Radiation dose reduction during adrenal vein sampling using a new angiographic imaging technology
    Spink, Clemens
    Avanesov, Maxim
    Lenz, Alexander
    Henes, Frank Oliver
    Well, Lennart
    Schmidt, Thomas
    Adam, Gerhard
    Ittrich, Harald
    Bannas, Peter
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [27] Characteristics of a New X-Ray Imaging System for Interventional Procedures: Improved Image Quality and Reduced Radiation Dose
    Schernthaner, Ruediger E.
    Haroun, Reham R.
    Nguyen, Sonny
    Duran, Rafael
    Sohn, Jae Ho
    Sahu, Sonia
    Chapiro, Julius
    Zhao, Yan
    Radaelli, Alessandro
    van der Bom, Imramsjah M.
    Mauti, Maria
    Hong, Kelvin
    Geschwind, Jean-Francois H.
    Lin, MingDe
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2018, 41 (03) : 502 - 508
  • [28] Image quality based x-ray dose control in cardiac imaging
    Davies, Andrew G.
    Kengyelics, Stephen M.
    Gislason-Lee, Amber J.
    IMAGE PROCESSING: ALGORITHMS AND SYSTEMS XIII, 2015, 9399
  • [29] Significant Radiation Dose Reduction Using a Novel Angiography Platform in Patients Undergoing Cryoballoon Pulmonary Vein Isolation
    Schmidt, Martin
    Tesche, Christian
    Hapfelmeier, Alexander
    Finsterer, Sebastian
    Valassis, Gregorius
    Ebersberger, Ullrich
    Schlueter, Mathis
    Ziegler, Martin
    Kraemer, Bjoern F.
    JOURNAL OF THORACIC IMAGING, 2020, 35 (03) : 167 - 172
  • [30] Radiation dose to staff from medical X-ray scatter in the orthopaedic theatre
    Dorman, T.
    Drever, B.
    Plumridge, S.
    Gregory, K.
    Cooper, M.
    Roderick, A.
    Arruzza, E.
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2023, 33 (07) : 3059 - 3065