Patient Radiation Dose Reduction during Transarterial Chemoembolization Using a Novel X-Ray Imaging Platform

被引:5
|
作者
Kohlbrenner, Ryan [1 ]
Kolli, Pallav [1 ]
Taylor, Andrew G. [1 ]
Kohi, Maureen P. [1 ]
Fidelman, Nicholas [1 ]
LaBerge, Jeanne M. [1 ]
Kerlan, Robert K. [1 ]
Agarwal, Vishal K. [1 ]
Lehrman, Evan D. [1 ]
Nanavati, Sujal [1 ]
Avrin, David E. [1 ]
Gould, Robert [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiol, San Francisco, CA 94143 USA
关键词
INTERVENTIONAL RADIOLOGY PROCEDURES; DIGITAL-SUBTRACTION-ANGIOGRAPHY; AREA PRODUCT; FLUOROSCOPY; TECHNOLOGY; QUALITY; TIME; IR;
D O I
10.1016/j.jvir.2015.06.016
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate radiation dose reduction in patients undergoing transarterial chemoembolization with the use of a new image acquisition and processing platform. Materials and 'Methods: Radiation-dose data were obtained from 176 consecutive chemoembolization procedures in 135 patients performed in a single angiography suite. From January 2013 through October 2013, 85 procedures were, performed by using our institution's standard fluoroscopic settings. After upgrading the x-ray fluoroscopy system with an image acquisition and processing platform designed to reduce image noise and reduce skin entrance dose, 91 chemoembolization procedures were performed from November 2013 through December 2014. Cumulative dose area product (CDAP), cumulative air kerma (CAK), and total fluoroscopy time were recorded for each procedure. Image quality was assessed by three interventional radiologists blinded to the x-ray acquisition platform used. Results: Patient radiation dose indicators were significantly lower for chemoembolization procedures perfauned with the novel imaging platform. Mean CDAP decreased from 3,033.2 dGy.cm(2) (range, 600.3-9,404.1 dGy.cm(2)) to 1,640.1 dGy.cm(2) (range, 278.6-6,779.9 dGy.cm(2); 45.9% reduction; P<.00001). Mean CAK 'decreased from 1,445.4 mGy (range, 303.6-5,233.7 mGy) to 971.7 mGy (range, 144.2-3,512.0 mGy; 32.8% reduction; P<.0001). A 20.3% increase in mean total fluoroscopy time was noted after upgrading the imaging platform, but blinded analysis of the image quality revealed no significant degradation. Conclusions: Although a small increase in fluoroscopy time was observed, a significant reduction in patient radiation dose was achieved by using the optimized imaging platform, without image quality degradation.
引用
收藏
页码:1331 / 1338
页数:8
相关论文
共 50 条
  • [1] Novel X-Ray Imaging Technology Allows Substantial Patient Radiation Reduction without Image Quality Impairment in Repetitive Transarterial Chemoembolization for Hepatocellular Carcinoma
    Wen, Xiaofei
    Jiang, Xianxian
    Li, Renfei
    Zhang, Junya
    Yang, Po
    Shen, Baozhong
    ACADEMIC RADIOLOGY, 2015, 22 (11) : 1361 - 1367
  • [2] The Impact of Novel X-Ray Systems and X-Ray System Optimization on Patient Radiation Dose Administered During Cardiac Catheterization
    Vemmou, Evangelia
    Nikolakopoulos, Ilias
    Xenogiannis, Iosif
    Megaly, Michael
    Omer, Mohamed A.
    Stanberry, Larissa
    Garberich, Ross
    Rangan, Bavana, V
    Baran, Kenneth W.
    Gossl, Mario
    Sorajja, Paul
    Chavez, Ivan
    Mooney, Michael
    Traverse, Jay
    Wang, Yale
    Garcia, Santiago
    Poulose, Anil
    Burke, M. Nicholas
    Brilakis, Emmanouil S.
    JOURNAL OF INVASIVE CARDIOLOGY, 2020, 32 (06) : 218 - 221
  • [3] Reduced Patient Radiation Exposure during Neurodiagnostic and Interventional X-Ray Angiography with a New Imaging Platform
    van der Marel, K.
    Vedantham, S.
    van der Bom, I. M. J.
    Howk, M.
    Narain, T.
    Ty, K.
    Karellas, A.
    Gounis, M. J.
    Puri, A. S.
    Wakhloo, A. K.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2017, 38 (03) : 442 - 449
  • [4] Radiation Dose Reduction during Uterine Fibroid Embolization Using an Optimized Imaging Platform
    Kohlbrenner, Ryan
    Kolli, K. Pallav
    Taylor, Andrew G.
    Kohi, Maureen P.
    Lehrman, Evan D.
    Fidelman, Nicholas
    Conrad, Miles
    LaBerge, Jeanne M.
    Kerlan, Robert K.
    Gould, Robert
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2017, 28 (08) : 1129 - 1135
  • [5] Advanced Robotic Angiography Systems for Image Guidance During Conventional Transarterial Chemoembolization Impact on Radiation Dose and Image Quality
    Vogl, Thomas J.
    Alizadeh, Leona S.
    Maeder, Richard
    Naguib, Nagy N.
    Herrmann, Eva
    Bickford, Matthew W.
    Burck, Iris
    Albrecht, Moritz H.
    INVESTIGATIVE RADIOLOGY, 2019, 54 (03) : 153 - 159
  • [6] Dose optimization in cardiac x-ray imaging
    Gislason-Lee, Amber J.
    McMillan, Catherine
    Cowen, Arnold R.
    Davies, Andrew G.
    MEDICAL PHYSICS, 2013, 40 (09)
  • [7] X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
    Burion, Steve
    Funk, Tobias
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2011, (55):
  • [8] Dose optimization in pediatric cardiac x-ray imaging
    Gislason, Amber J.
    Davies, Andrew G.
    Cowen, Arnold R.
    MEDICAL PHYSICS, 2010, 37 (10) : 5258 - 5269
  • [9] Patient-specific region-of-interest fluoroscopy device for X-ray dose reduction
    Xu, T
    Le, HQ
    Molloi, S
    RADIOLOGY, 2003, 226 (02) : 585 - 592
  • [10] A new angiographic imaging platform reduces radiation exposure for patients with liver cancer treated with transarterial chemoembolization
    Schernthaner, Ruediger E.
    Duran, Rafael
    Chapiro, Julius
    Wang, Zhijun
    Geschwind, Jean-Francois H.
    Lin, MingDe
    EUROPEAN RADIOLOGY, 2015, 25 (11) : 3255 - 3262