Resonance frequency measurements of implant stability in vivo - A cross-sectional and longitudinal study of resonance frequency measurements on implants in the edentulous and partially dentate maxilla

被引:348
|
作者
Meredith, N
Book, K
Friberg, B
Jemt, T
Sennerby, L
机构
[1] Dept. of Oral and Dental Science, University of Bristol Dental School, Bristol BS1 2LY, Lower Maudlin Street
[2] Dept. of Biomaterials/Handicap Res., Institute for Surgical Sciences, Göteborg University, Gothenburg
[3] Branemark Clinic, Public Dental Health, University of Göteborg, Gothenburg
关键词
dental; implant; osseointegration; resonance frequency; stability;
D O I
10.1034/j.1600-0501.1997.080309.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The aim of this investigation was to evaluate the use of resonance frequency measurements in the clinical measurement of implant stability. Resonance frequency measurements are undertaken by measuring the response of a small transducer attached to an implant fixture or abutment. Two groups of patients were selected for study. Group A comprised 9 patients who had a total of 56 implants placed. Resonance frequency measurements were made at fixture installation and repeated 8 months later at abutment connection. The resonance frequency of the implant/transducer system increased for 50 out of the 56 implants from a mean value of 7473 Hz +/- 127 Hz (P<0.05) to a mean of 7915 Hz +/- 112 Hz (P<0.05). Two implants had failed to integrate and the resonance frequency of these had fallen. Group B comprised 9 patients who had been provided with fixed prostheses and had a total of 52 implants placed. They were examined 5 years after fixture placement and the prostheses removed. All implants were judged clinically to be osseointegrated. The level of the marginal bone around each implant was calculated by measuring the number of exposed threads on intraoral periapical radiographs and added to the length of each abutment to give a value termed the effective implant length (EIL). Measurements indicated a correlation (R = -0.78, P<0.01) between EIL and resonance frequency. The results support the hypothesis that the resonance frequency of an implant/ transducer system is related to the height of the implant not surrounded by bone and the stability of the implant/tissue interface as determined by the absence of clinical mobility.
引用
收藏
页码:226 / 233
页数:8
相关论文
共 38 条
  • [31] Analysis of implant stability changes in immediate loading using a laser displacement sensor in vivo and comparison of its sensitivity with that of resonance frequency analysis
    Murakami, Kazuhiro
    Yamamoto, Kazuhiko
    Ishida, Junichi
    Tsutsumi, Sadami
    Kirita, Tadaaki
    CLINICAL ORAL IMPLANTS RESEARCH, 2021, 32 (11) : 1341 - 1356
  • [32] In vitro comparison of resonance frequency analysis devices to evaluate implant stability of narrow diameter implants at varying drilling speeds in dense artificial bone blocks
    Romanos, Georgios E.
    Bastardi, Daniel J.
    Kakar, Apoory
    Moore, Rachel
    Delgado-Ruiz, Rafael A.
    Javed, Fawad
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2019, 21 (05) : 1023 - 1027
  • [33] Osseointegration of magnesium-incorporated sand-blasted acid-etched implant in the dog mandible: Resonance frequency measurements and histomorphometric analysis
    Song, Won-Wook
    Heo, Jin-Ho
    Lee, Jeong-Han
    Park, Young-Min
    Kim, Yong-Deok
    TISSUE ENGINEERING AND REGENERATIVE MEDICINE, 2016, 13 (02) : 191 - 199
  • [34] Can resonance frequency analysis detect narrow marginal bone defects around dental implants? An ex vivo animal pilot study
    Yao, C. J.
    Ma, L.
    Mattheos, N.
    AUSTRALIAN DENTAL JOURNAL, 2017, 62 (04) : 433 - 439
  • [35] Clinical study on the insertion torque and wireless resonance frequency analysis in the assessment of torque capacity and stability of self-tapping dental implants
    Kahraman, S.
    Bal, B. T.
    Asar, N. V.
    Turkyilmaz, I.
    Tozum, T. F.
    JOURNAL OF ORAL REHABILITATION, 2009, 36 (10) : 755 - 761
  • [36] Histologic and Resonance Frequency Analysis of Peri-Implant Bone Healing After Low-Level Laser Therapy: An In Vivo Study
    Mayer, Luciano
    Gomes, Fernando Vacilotto
    Carlsson, Lennart
    Gerhardt-Oliveira, Marilia
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2015, 30 (05) : 1028 - 1035
  • [37] The effect of platelet-rich plasma upon implant stability measured by resonance frequency analysis in the lower anterior mandibles - A pilot study
    Monov, G
    Fuerst, G
    Tepper, G
    Watzak, G
    Zechner, W
    Watzek, G
    CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (04) : 461 - 465
  • [38] Comparative radiographic and resonance frequency analyses of the peri-implant tissue after dental implants placement using flap and flapless techniques: An experimental study on domestic pigs
    Vlahovic, Zoran
    Mihailovic, Branko
    Lazic, Zoran
    Golubovic, Mileta
    VOJNOSANITETSKI PREGLED, 2013, 70 (06) : 586 - 594