Establishment and results of a magnetic resonance quality assurance program for the Pediatric Brain Tumor Consortium

被引:21
作者
Mulkern, Robert V. [1 ]
Forbes, Peter [2 ]
Dewey, Kevin [1 ]
Osganian, Stravoula [2 ]
Clark, Maureen [2 ]
Wong, Sharon [2 ]
Ramamurthy, Uma [3 ]
Kun, Larry [3 ]
Poussaint, Tina Young [1 ]
机构
[1] Childrens Hosp, Dept Radiol, Boston, MA 02115 USA
[2] Childrens Hosp, Clin Res Program, Boston, MA 02115 USA
[3] St Jude Childrens Hosp, Memphis, TN 38105 USA
关键词
magnetic resonance; quality assurance; pediatric brain tumor; American College of Radiology;
D O I
10.1016/j.acra.2008.04.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. Magnetic resonance (MR) imaging is used to assess brain tumor response to therapies, and a MR quality assurance (QA) program is necessary for multicenter clinical trials employing imaging. This study was performed to determine overall variability of quantitative imaging metrics measured with the American College of Radiology (ACR) phantom among 11 sites participating in the Pediatric Brain Tumor Consortium (PBTC) Neuroimaging Center (NIC) MR QA program. Materials and Methods. An MR QA program was implemented among 11 participating PBTC sites and quarterly evaluations of scanner performance for seven imaging metrics defined by the ACR were Sought and subject to statistical evaluation over a 4.5-year period. Overall compliance with the QA program, means, standard deviations, and coefficients of variation (CV). for the quantitative imaging metrics were evaluated. Results. Quantitative measures of the seven imaging metrics were generally within ACR recommended guidelines for all sites. Compliance improved as the study progressed. Intersite variabilities, as gauged by CV for slice thickness and geometric accuracy, imaging parameters that influence size or positioning measurements in tumor studies, were on the order of 10% and 1%, respectively. Conclusions. Although challenging to establish, MR QA programs within the context of PBTC multisite clinical trials when based on the ACR MR phantom program can indicate sites performing below acceptable image quality levels and establish levels of precision through instrumental variabilities that are relevant to quantitative image analyses (eg, tumor volume changes).
引用
收藏
页码:1099 / 1110
页数:12
相关论文
共 50 条
[21]   Vorinostat and isotretinoin with chemotherapy in young children with embryonal brain tumors: A report from the Pediatric Brain Tumor Consortium (PBTC-026) [J].
Leary, Sarah E. S. ;
Lindsay, Kilburn ;
Geyer, J. Russell ;
Mehmet, Kocak ;
Huang, Jie ;
Smith, Kyle S. ;
Hadley, Jennifer ;
Ermoian, Ralph ;
MacDonald, Tobey J. ;
Goldman, Stewart ;
Phillips, Peter ;
Poussaint, Tina Young ;
Olson, James M. ;
Ellison, David W. ;
Dunkel, Ira J. ;
Fouladi, Maryam ;
Onar-Thomas, Arzu ;
Northcott, Paul A. .
NEURO-ONCOLOGY, 2022, 24 (07) :1178-1190
[22]   Cone beam computed tomography for dose calculation quality assurance for magnetic resonance-only radiotherapy [J].
Wyatt, Jonathan J. ;
Pearson, Rachel A. ;
Walker, Christopher P. ;
Brooks, Rachel L. ;
Pilling, Karen ;
McCallum, Hazel M. .
PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2021, 17 :71-76
[23]   Development of a temperature-controlled phantom for magnetic resonance quality assurance of diffusion, dynamic, and relaxometry measurements [J].
Jerome, Neil P. ;
Papoutsaki, Marianthi-Vasiliki ;
Orton, Matthew R. ;
Parkes, Harold G. ;
Winfield, Jessica M. ;
Boss, Michael A. ;
Leach, Martin O. ;
deSouza, Nandita M. ;
Collins, David J. .
MEDICAL PHYSICS, 2016, 43 (06) :2998-3007
[24]   Developing quality assurance tests for simultaneous Positron Emission Tomography - Magnetic Resonance imaging for radiotherapy planning [J].
Wyatt, Jonathan J. ;
McCallum, Hazel M. ;
Maxwell, Ross J. .
PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2022, 22 :28-35
[25]   Quality Assurance for Magnetic Resonance - guided High Intensity Focused Ultrasound at Hospital Universiti Sains Malaysia [J].
Taib, Nur Hartini Mohd ;
Abdullah, Mohd Shafie ;
Haron, Juhara ;
Ahmed, Wan Aireene Wan .
2018 IEEE-EMBS CONFERENCE ON BIOMEDICAL ENGINEERING AND SCIENCES (IECBES), 2018, :146-149
[26]   A novel multimodality anthropomorphic phantom enhances compliance with quality assurance guidelines for magnetic resonance imaging in radiotherapy [J].
Alzahrani, Meshal ;
Broadbent, David A. ;
Teh, Irvin ;
Al-Qaisieh, Bashar ;
Johnstone, Emily ;
Speight, Richard .
PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2025, 33
[27]   EUREQUO: establishment of an ophthalmic quality assurance system - first year results from the hospital of St. John of God in Linz [J].
Luebke, Britta ;
Reinelt, Peter ;
Jirak, Paul ;
Schoenherr, Ulrich .
SPEKTRUM DER AUGENHEILKUNDE, 2011, 25 (06) :364-367
[28]   Quality assurance of magnetic resonance imaging for a polymer gel dosimeter using a 3D-printed phantom [J].
Tachibana, Hidenobu ;
Hoshino, Yoshihiko ;
Watanabe, Yusuke ;
Usui, Keisuke ;
Mizukami, Shinya ;
Shibukawa, Shuhei ;
Kodama, Takashi ;
Tachibana, Rie .
RADIATION PHYSICS AND CHEMISTRY, 2025, 226
[29]   Quality assurance issues for computed tomography-, ultrasound-, and magnetic resonance imaging-guided brachytherapy [J].
Cormack, Robert A. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01) :S136-S141
[30]   Magnetic resonance spectroscopy (MRS) in the evaluation of pediatric brain tumors .1. Introduction to MRS [J].
Byrd, SE ;
Tomita, T ;
Palka, PS ;
Darling, CF ;
Norfray, JP ;
Fan, J .
JOURNAL OF THE NATIONAL MEDICAL ASSOCIATION, 1996, 88 (10) :649-654