Establishment and results of a magnetic resonance quality assurance program for the Pediatric Brain Tumor Consortium

被引:21
作者
Mulkern, Robert V. [1 ]
Forbes, Peter [2 ]
Dewey, Kevin [1 ]
Osganian, Stravoula [2 ]
Clark, Maureen [2 ]
Wong, Sharon [2 ]
Ramamurthy, Uma [3 ]
Kun, Larry [3 ]
Poussaint, Tina Young [1 ]
机构
[1] Childrens Hosp, Dept Radiol, Boston, MA 02115 USA
[2] Childrens Hosp, Clin Res Program, Boston, MA 02115 USA
[3] St Jude Childrens Hosp, Memphis, TN 38105 USA
关键词
magnetic resonance; quality assurance; pediatric brain tumor; American College of Radiology;
D O I
10.1016/j.acra.2008.04.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objectives. Magnetic resonance (MR) imaging is used to assess brain tumor response to therapies, and a MR quality assurance (QA) program is necessary for multicenter clinical trials employing imaging. This study was performed to determine overall variability of quantitative imaging metrics measured with the American College of Radiology (ACR) phantom among 11 sites participating in the Pediatric Brain Tumor Consortium (PBTC) Neuroimaging Center (NIC) MR QA program. Materials and Methods. An MR QA program was implemented among 11 participating PBTC sites and quarterly evaluations of scanner performance for seven imaging metrics defined by the ACR were Sought and subject to statistical evaluation over a 4.5-year period. Overall compliance with the QA program, means, standard deviations, and coefficients of variation (CV). for the quantitative imaging metrics were evaluated. Results. Quantitative measures of the seven imaging metrics were generally within ACR recommended guidelines for all sites. Compliance improved as the study progressed. Intersite variabilities, as gauged by CV for slice thickness and geometric accuracy, imaging parameters that influence size or positioning measurements in tumor studies, were on the order of 10% and 1%, respectively. Conclusions. Although challenging to establish, MR QA programs within the context of PBTC multisite clinical trials when based on the ACR MR phantom program can indicate sites performing below acceptable image quality levels and establish levels of precision through instrumental variabilities that are relevant to quantitative image analyses (eg, tumor volume changes).
引用
收藏
页码:1099 / 1110
页数:12
相关论文
共 13 条
[1]  
[Anonymous], 2005, PHANTOM TEST GUIDANC
[2]   Multicenter trial for the set-up of a MRI quality assurance programme [J].
Colombo, P ;
Baldassarri, A ;
Del Corona, M ;
Mascaro, L ;
Strocchi, S .
MAGNETIC RESONANCE IMAGING, 2004, 22 (01) :93-101
[3]   THE LOSS OF SMALL OBJECTS IN VARIABLE TE IMAGING - IMPLICATIONS FOR FSE, RARE, AND EPI [J].
CONSTABLE, RT ;
GORE, JC .
MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (01) :9-24
[4]   The NIH MRI study of normal brain development [J].
Evans, AC .
NEUROIMAGE, 2006, 30 (01) :184-202
[5]   Perfusion-weighted magnetic resonance imaging of the brain: techniques and application in children [J].
Huisman, TAGM ;
Sorensen, AG .
EUROPEAN RADIOLOGY, 2004, 14 (01) :59-72
[6]   MRI quality control:: six imagers studied using eleven unified image quality parameters [J].
Ihalainen, T ;
Sipilä, O ;
Savolainen, S .
EUROPEAN RADIOLOGY, 2004, 14 (10) :1859-1865
[8]   PARTIAL RF ECHO PLANAR IMAGING WITH THE FAISE METHOD .1. EXPERIMENTAL AND THEORETICAL ASSESSMENT OF ARTIFACT [J].
MELKI, PS ;
JOLESZ, FA ;
MULKERN, RV .
MAGNETIC RESONANCE IN MEDICINE, 1992, 26 (02) :328-341
[9]   CONTRAST MANIPULATION AND ARTIFACT ASSESSMENT OF 2D AND 3D RARE SEQUENCES [J].
MULKERN, RV ;
WONG, STS ;
WINALSKI, C ;
JOLESZ, FA .
MAGNETIC RESONANCE IMAGING, 1990, 8 (05) :557-566
[10]   Correlation of relative permeability and relative cerebral blood volume in high-grade cerebral neoplasms [J].
Provenzale, James M. ;
York, Gerald ;
Moya, Marcelo Galvez ;
Parks, Lauren ;
Choma, Michael ;
Kealey, Susan ;
Cole, Patricia ;
Serajuddin, Humaira .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 187 (04) :1036-1042