Falsely elevated serum vancomycin concentrations in hemodialysis patients

被引:32
作者
Follin, SL
Mueller, BA
Scott, MK
Carfagna, MA
Kraus, MA
机构
[1] PURDUE UNIV,SCH PHARM & PHARMACAL SCI,DEPT PHARM PRACTICE,W LAFAYETTE,IN 47907
[2] INDIANA UNIV,SCH MED,DEPT PHARMACOL,INDIANAPOLIS,IN 46202
[3] INDIANA UNIV,SCH MED,DEPT TOXICOL,INDIANAPOLIS,IN 46202
[4] INDIANA UNIV,MED CTR,DEPT MED,DIV NEPHROL,INDIANAPOLIS,IN 46202
关键词
vancomycin; chronic kidney failure; enzyme multiplied immunoassay technique; fluorescence polarization immunoassay; drug monitoring;
D O I
10.1016/S0272-6386(96)90032-3
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Fluorescence polarization immunoassay (FPIA) is the most widely used clinical vancomycin assay in the United States, Questions exist regarding the accuracy of this polyclonal assay in patients with end-stage renal disease (ESRD). While several studies have reported discrepancies in vancomycin serum concentrations determined by FPIA compared with other vancomycin assays, no study has investigated the accuracy of vancomycin serum concentrations determined by FPIA in patients with ESRD undergoing maintenance hemodialysis, Therefore, we compared the assay performance of FPIA and enzyme multiplied immunoassay technique (EMIT) in six subjects with ESRD receiving high-efficiency hemodialysis, Subjects underwent 6 consecutive weeks of hemodialysis treatment with a cellulose acetate dialyzer (CA210) and received 1 g vancomycin intravenously once weekly during the last hour of dialysis. Vancomycin serum concentrations were determined by both EMIT and FPIA methodologies, From the serum concentration results of both assays, vancomycin dosing recommendations were calculated to achieve a desired steady-state peak concentration of 35 mg/L and trough concentration of 10 mg/L. Overall, vancomycin serum concentrations reported by FPIA were significantly higher than those reported by EMIT, The mean difference between assays in the peak serum concentrations at weeks 1, 4, and 6 was 7.5, 11.5, and 11.2 mg/L, respectively. The mean difference in trough serum concentrations at weeks 1, 4, and 6 was 4.2, 6.2, and 5.2 mg/L, respectively, The FPIA overestimation of the EMIT values (calculated as FPIA - EMIT) varied widely among study subjects with a range of 0.0 mg/L to 27.0 mg/L for peak serum concentrations and 0.0 mg/L to 12.8 mg/L for trough serum concentrations, The mean doses calculated based on FPIA results were significantly lower than the EMIT-derived doses, No significant difference was observed in the calculated dosing intervals, These results demonstrate that FPIA significantly overestimates vancomycin serum concentrations compared with EMIT in patients with ESRD undergoing high-efficiency hemodialysis, The overestimation by FPIA may result in significantly different vancomycin dosing recommendations, leading to underdosing and the potential for therapeutic failures. Due to the unpredictability of the overestimation by FPIA, we were unable to formulate vancomycin dosing guidelines for institutions that use FPIA. Therefore, we recommend that the EMIT vancomycin assay be used in patients with ESRD to ensure appropriate dosing. (C) 1996 by the National Kidney Foundation, Inc.
引用
收藏
页码:67 / 74
页数:8
相关论文
共 24 条
[1]  
ANNE L, 1989, THER DRUG MONIT, V11, P585
[2]  
BASTANI B, 1988, DIALYSIS TRANSPLANT, V17, P527
[3]  
CANTU TG, 1994, CLIN INFECT DIS, V18, P533
[4]  
COMSTOCK TJ, 1988, CLIN PHARMACOL THER, V43, P172
[5]  
DESOI CA, 1992, AM J KIDNEY DIS, V20, P354
[6]  
FREEMAN CD, 1993, ANN PHARMACOTHER, V27, P594
[7]   STRUCTURE OF THE GLYCOPEPTIDE ANTIBIOTIC VANCOMYCIN - EVIDENCE FOR AN ASPARAGINE RESIDUE IN THE PEPTIDE [J].
HARRIS, CM ;
HARRIS, TM .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1982, 104 (15) :4293-4295
[8]   VANCOMYCIN - STRUCTURE AND TRANSFORMATION TO CDP-I [J].
HARRIS, CM ;
KOPECKA, H ;
HARRIS, TM .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1983, 105 (23) :6915-6922
[9]   MEASUREMENT OF VANCOMYCIN IN RENALLY IMPAIRED PATIENT SAMPLES USING A NEW HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY METHOD WITH VITAMIN-B12 INTERNAL STANDARD - COMPARISON OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, EMIT, AND FLUORESCENCE POLARIZATION IMMUNOASSAY METHODS [J].
HU, MW ;
ANNE, L ;
FORNI, T ;
GOTTWALD, K .
THERAPEUTIC DRUG MONITORING, 1990, 12 (06) :562-569
[10]   MARKEDLY INCREASED CLEARANCE OF VANCOMYCIN DURING HEMODIALYSIS USING POLYSULFONE DIALYZERS [J].
LANESE, DM ;
ALFREY, PS ;
MOLITORIS, BA .
KIDNEY INTERNATIONAL, 1989, 35 (06) :1409-1412