Comparison of the retinal image quality with a Hartmann-Shack wavefront sensor and a double-pass instrument

被引:185
作者
Díaz-Doutón, F
Benito, A
Pujol, J
Arjona, M
Güell, JL
Artal, P
机构
[1] Univ Murcia, Dept Fis, Lab Opt, E-30071 Murcia, Spain
[2] Univ Politecn Cataluna, Ctr Desenvolupament Sensor Instrumentacio & Siste, Terrassa, Spain
[3] Univ Autonoma Barcelona, E-08193 Barcelona, Spain
[4] IMO, Barcelona, Spain
关键词
D O I
10.1167/iovs.05-1049
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. Wavefront sensors provide quite Useful information on the optical quality of the eye. However, in eyes where very high-order aberrations and scattered light are prominent, wave-front sensors may overestimate retinal image quality. This study showed that, in those cases, the double-pass technique is complementary tool for better estimation of ocular optical quality. METHODS. A double-pass (DP) instrument was used, based on recording images of a point source in near-infrared light after reflection in the retina and double-pass through the Ocular media. The aberrations were also measured with a prototype of near-infrared Hartmann-Shack (HS) wavefront sensor adapted to the clinical environment. From the wave aberrations, the modulation transfer function (MTF) was calculated (MTF_HS). The mTF was also obtained from the double-pass images (MTF_DP). Both techniques were applied in normal young subjects as the control and in three other groups of eyes: older subjects, after LASIK refractive surgery, and after IOL implantation. RESULTS. The MTFs obtained from DP and HS techniques were compared. In the group of normal eyes with low levels of intraocular scattering, these estimates were quite similar, indicating that both techniques captured well most of the optical degradation. However, in eves where scatter was more predominant (e.g., early cataract, posterior capsular opacification after IOL implantation) the MTF provided by the HS sensor was always higher than the MTF obtained from DP. A single parameter was used to indicate the differences. CONCLUSIONS. In eyes with low scattering, DP and HS techniques provided similar estimates of the retinal image quality. However, in a patient's eye with mild to severe amount Of scatter, wavefront sensors might overestimate image quality, whereas the DP technique produces a more accurate description of the optical quality, better correlated with the quality of vision.
引用
收藏
页码:1710 / 1716
页数:7
相关论文
共 39 条
  • [21] Guirao A, 1999, INVEST OPHTH VIS SCI, V40, P203
  • [22] Measurement of the wave-front aberration of the eye by a fast psychophysical procedure
    He, JC
    Marcos, S
    Webb, RH
    Burns, SA
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1998, 15 (09): : 2449 - 2456
  • [23] Improvement in retinal image quality with dynamic correction of the eye's aberrations
    Hofer, H
    Chen, L
    Yoon, GY
    Singer, B
    Yamauchi, Y
    Williams, DR
    [J]. OPTICS EXPRESS, 2001, 8 (11): : 631 - 643
  • [24] Estimates of the ocular wave aberration from pairs of double-pass retinal images
    Iglesias, I
    Berrio, E
    Artal, P
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1998, 15 (09): : 2466 - 2476
  • [25] Extended source pyramid wave-front sensor for the human eye
    Iglesias, I
    Ragazzoni, R
    Julien, Y
    Artal, P
    [J]. OPTICS EXPRESS, 2002, 10 (09): : 419 - 428
  • [26] Calculation of ocular single-pass modulation transfer function and retinal image simulation from measurements of the polarized double-pass ocular point spread function
    Kobayashi, K
    Shibutani, M
    Takeuchi, G
    Ohnuma, K
    Miyake, Y
    Negishi, K
    Ohno, K
    Noda, T
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2004, 9 (01) : 154 - 161
  • [27] OBJECTIVE MEASUREMENT OF WAVE ABERRATIONS OF THE HUMAN EYE WITH THE USE OF A HARTMANN-SHACK WAVE-FRONT SENSOR
    LIANG, JZ
    GRIMM, B
    GOELZ, S
    BILLE, JF
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1994, 11 (07): : 1949 - 1957
  • [28] Aberrations and retinal image quality of the normal human eye
    Liang, JZ
    Williams, DR
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1997, 14 (11) : 2873 - 2883
  • [29] Comparison of double-pass estimates of the retinal-image quality obtained with green and near-infrared light
    LopezGil, N
    Artal, P
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1997, 14 (05): : 961 - 971
  • [30] Moreno-Barriuso E, 2001, INVEST OPHTH VIS SCI, V42, P1396