Performance Evaluation of a Synthetic Aperture Real-Time Ultrasound System

被引:0
作者
Stuart, M. B. [1 ]
Tomov, B. G. [1 ]
Jensen, J. A. [1 ]
机构
[1] Tech Univ Denmark, Dept Elect Engn, Ctr Fast Ultrasound Imaging, DK-2800 Lyngby, Denmark
来源
15TH NORDIC-BALTIC CONFERENCE ON BIOMEDICAL ENGINEERING AND MEDICAL PHYSICS (NBC 2011) | 2011年 / 34卷
关键词
Ultrasound; hardware; performance; beamformer; scanner;
D O I
暂无
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
This paper evaluates the signal-to-noise ratio, the time stability, and the phase difference of the sampling in the experimental ultrasound scanner SARUS: A synthetic aperture, real-time ultrasound system. SARUS has 1024 independent transmit and receive channels and is capable of handling 2D probes for 3D ultrasound imaging. It samples at 12 bits per sample and has a sampling rate of 70 MHz with the possibility of decimating the sampling frequency at the input. SARUS is capable of advanced real-time computations such as synthetic aperture imaging. The system is built using field-programmable gate arrays (FPGAs) making it very flexible and allowing implementation of other real-time ultrasound processing methods in the future. For conventional B-mode imaging, a penetration depth around 7 cm for a 7 MHz transducer is obtained (signal-to-noise ratio of 0 dB), which is comparable to commercial ultrasound scanners. Furthermore, the jitter between successive acquisitions for flow estimation is around 1.41 ps with a standard deviation of 48.3 Ps. This has a negligible impact (0.03%) on the flow measurement. Additionally, for the phase of the sampling, it is shown that the small differences between different channels (on average 111 ps for a 70 MHz sampling clock) are deterministic and can therefore be compensated for.
引用
收藏
页码:160 / 163
页数:4
相关论文
共 13 条
  • [1] Precise Time-of-Flight Calculation For 3-D Synthetic Aperture Focusing
    Andresen, Henrik
    Nikolov, Svetoslav Ivanov
    Jensen, Jorgen Arendt
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2009, 56 (09) : 1880 - 1887
  • [2] Preliminary invivo results for spatially coded synthetic transmit aperture ultrasound based on frequency division
    Gran, Fredrik
    Hansen, Kristoffer Lindskov
    Nielsen, Michael Bachmann
    Jensen, Jorgen Arendt
    [J]. 2006 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-5, PROCEEDINGS, 2006, : 1087 - +
  • [3] ANALYSIS OF WORST-CASE PHASE QUANTIZATION SIDELOBES IN FOCUSED BEAMFORMING
    HOLM, S
    KRISTOFFERSEN, K
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1992, 39 (05) : 593 - 599
  • [4] Ultrasound research scanner for real-time synthetic aperture data acquisition
    Jensen, JA
    Holm, O
    Jensen, LJ
    Bendsen, H
    Nikolov, SI
    Tomov, BG
    Munk, P
    Hansen, M
    Salomonsen, K
    Hansen, J
    Gormsen, K
    Pedersen, HM
    Gammelmark, KL
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2005, 52 (05) : 881 - 891
  • [5] Directional synthetic aperture flow imaging
    Jensen, JA
    Nikolov, SI
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2004, 51 (09) : 1107 - 1118
  • [6] Jensen JA, 1999, ULTRASON, P1595, DOI 10.1109/ULTSYM.1999.849300
  • [7] Estimation of velocity vectors in synthetic aperture ultrasound imaging
    Jensen, Jorgen Arendt
    Oddershede, Niels
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2006, 25 (12) : 1637 - 1644
  • [8] High frame rate imaging system for limited diffraction array beam imaging with square-wave aperture weightings
    Lu, Jian-yu
    Cheng, Jiqi
    Wang, Jing
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006, 53 (10) : 1796 - 1812
  • [9] Misaridis TX, 2000, ULTRASON, P1689, DOI 10.1109/ULTSYM.2000.921647
  • [10] Nikolov S, 2002, IEEE P ULTR S, P1545