Comparison of Intraocular Pressure before and after Laser In Situ Keratomileusis Refractive Surgery Measured with Perkins Tonometry, Noncontact Tonometry, and Transpalpebral Tonometry

被引:15
作者
Cacho, Isabel [1 ]
Sanchez-Naves, Juan [1 ]
Batres, Laura [2 ]
Pintor, Jesus [3 ]
Carracedo, Gonzalo [2 ,3 ]
机构
[1] Inst Balear Oftalmol, Palma De Mallorca 07011, Spain
[2] Univ Complutense Madrid, Fac Opt & Optomet, Dept Opt Optomet & Vis 2, Madrid 28037, Spain
[3] Univ Complutense Madrid, Fac Opt & Optomet, Dept Bioquim & Biol Mol 4, Madrid 28037, Spain
关键词
DYNAMIC CONTOUR TONOMETRY; GOLDMANN APPLANATION TONOMETRY; CORNEAL BIOMECHANICAL PROPERTIES; GLAUCOMA; TGDC-01; MYOPIA; EYES; THICKNESS; LASIK;
D O I
10.1155/2015/683895
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To compare the intraocular pressure (IOP) before and after Laser In Situ Keratomileusis (LASIK), measured by Diaton, Perkins, and noncontact air pulse tonometers. Methods. Fifty-seven patients with a mean age of 34.88 were scheduled for myopia LASIK treatment. Spherical equivalent refraction (SER), corneal curvature (K), and central corneal thickness (CCT) and superior corneal thickness (SCT) were obtained before and after LASIK surgery. IOP values before and after surgery were measured using Diaton, Perkins, and noncontact air pulse tonometers. Results. The IOP values before and after LASIK surgery using Perkins tonometer and air tonometers were statistically significant (p < 0.05). However, no significant differences were found (p > 0.05) for IOP values measured with Diaton tonometer. CCT decreases significantly after surgery (p < 0.05) but no statistical differences were found in SCT (p = 0.08). Correlations between pre- and postsurgery were found for all tonometers used, with p = 0.001 and p = 0.434 for the air pulse tonometer, p = 0.008 and r = 0.355 for Perkins, and p < 0.001 and p = 0.637 for Diaton. Conclusion. Transpalpebral tonometry may be useful for measuring postsurgery IOP after myopic LASIK ablation because this technique is not influenced by the treatment.
引用
收藏
页数:6
相关论文
共 32 条
  • [1] Agudelo LM, 2002, J REFRACT SURG, V18, P472
  • [2] Bhan A, 2002, INVEST OPHTH VIS SCI, V43, P1389
  • [3] Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of Intraocular pressure
    Broman, Aimee Teo
    Congdon, Nathan G.
    Bandeen-Roche, Karen
    Quigley, Harry A.
    [J]. JOURNAL OF GLAUCOMA, 2007, 16 (07) : 581 - 588
  • [4] Comparison of Goldmann applanation tonometry and dynamic contour tonometry in healthy and glaucomatous eyes
    Ceruti, P.
    Morbio, R.
    Marraffa, M.
    Marchini, G.
    [J]. EYE, 2009, 23 (02) : 262 - 269
  • [5] Ocular blood-flow hemodynamics before and after application of a laser in situ keratomileusis ring
    Conway, Miriam L.
    Wevill, Mark
    Benavente-Perez, Alexandra
    Hosking, Sarah L.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (02) : 268 - 272
  • [6] Systematic Review of the Agreement of Tonometers with Goldmann Applanation Tonometry
    Cook, Jonathan Alistair
    Botello, Adriana Paola
    Elders, Andrew
    Ali, Alia Fathi
    Azuara-Blanco, Augusto
    McCormack, Kirsty
    Burr, Jennifer Margaret
    [J]. OPHTHALMOLOGY, 2012, 119 (08) : 1552 - 1557
  • [7] De Moraes C.G. V., 2008, J OPTOM, V1, P43, DOI [10.3921/joptom.2008.43, DOI 10.3921/JOPTOM.2008.43]
  • [8] Diaton tonometry: an assessment of validity and preference against Goldmann tonometry
    Doherty, Mark D.
    Carrim, Zia I.
    O'Neill, Damian P.
    [J]. CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 (04) : E171 - E175
  • [9] El Danasoury MA, 2001, J REFRACT SURG, V17, P97
  • [10] A correction formula for the real intraocular pressure after LASIK for the correction of myopic astigmatism
    Kohlhaas, M
    Spoerl, E
    Boehm, AG
    Pollack, K
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2006, 22 (03) : 263 - 267