Detection of Salmonella Typhimurium with Gold Nanoparticles Using Quartz Crystal Microbalance Biosensor

被引:9
作者
Min, Hyun Jung [1 ]
Mina, Hansel A. [2 ]
Deering, Amanda J. [2 ]
Robinson, J. Paul [3 ,4 ]
Bae, Euiwon [1 ]
机构
[1] Purdue Univ, Sch Mech Engn, Appl Opt Lab, W Lafayette, IN 47907 USA
[2] Purdue Univ, Dept Food Sci, W Lafayette, IN 47907 USA
[3] Purdue Univ, Dept Basic Med Sci, W Lafayette, IN 47907 USA
[4] Purdue Univ, Weldon Sch Biomed Engn, W Lafayette, IN 47907 USA
关键词
quartz crystal microbalance; Salmonella Typhimurium; gold nanoparticles; frequency shifts; SENSITIVE DETECTION; QCM IMMUNOSENSOR; SENSOR; DNA; ANTIBODIES; FREQUENCY; SURFACES; SPR;
D O I
10.3390/s22228928
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Demonstration of the Salmonella Typhimurium detection system was shown utilizing a quartz crystal microbalance (QCM) biosensor and signal enhancement by gold nanoparticles. In this study, a benchtop system of a QCM biosensor was utilized for the detection of Salmonella Typhimurium. It was designed with a peristaltic pump system to achieve immobilization of antibodies, detection of Salmonella, and the addition of gold nanoparticles to the sensor. As a series of biochemical solutions were introduced to the surface, the proposed system was able to track the changes in the resonant frequency which were proportional to the variations of mass on the sensor. For antibody immobilization, polyclonal antibodies were immobilized via self-assembled monolayers to detect Salmonella O-antigen. Subsequently, Salmonella Typhimurium was detected by antibodies and the average frequency before and after detecting Salmonella was compared. The highest frequency shifts were -26.91 Hz for 10(9) CFU/mL while the smallest frequency shift was -3.65 Hz corresponding to 103 CFU/mL. For the specificity tests, non-Salmonella samples such as E. coli, Listeria, and Staphylococcus resulted in low cross-reactivity. For signal amplification, biotinylated antibodies reacted to Salmonella followed by streptavidin-100 nm AuNPs through biotin-avidin interaction. The frequency shifts of 103 CFU/mL showed -28.04 Hz, and consequently improved the limit of detection.
引用
收藏
页数:14
相关论文
共 49 条
[1]  
Alexander Todd E, 2019, Cell Surf, V5, P100024, DOI 10.1016/j.tcsw.2019.100024
[2]  
Andrews WH., 2011, Bacteriological Analytical Manual (BAM)
[3]   Capillary aging of the contacts between glass spheres and a quartz resonator surface [J].
D'Amour, JN ;
Stålgren, JJR ;
Kanazawa, KK ;
Frank, CW ;
Rodahl, M ;
Johannsmann, D .
PHYSICAL REVIEW LETTERS, 2006, 96 (05)
[4]   Rapid sensing of hepatitis B virus using QCM in the thickness shear mode [J].
Dultsev, Fedor N. ;
Tronin, Andrei V. .
SENSORS AND ACTUATORS B-CHEMICAL, 2015, 216 :1-5
[5]   Recent biomedical applications of gold nanoparticles: A review [J].
Elahi, Narges ;
Kamali, Mehdi ;
Baghersad, Mohammad Hadi .
TALANTA, 2018, 184 :537-556
[6]   Label Free QCM Immunobiosensor for AFB1 Detection Using Monoclonal IgA Antibody as Recognition Element [J].
Ertekin, Ozlem ;
Ozturk, Selma ;
Ozturk, Zafer Ziya .
SENSORS, 2016, 16 (08)
[7]   QCM immunosensor with nanoparticle amplification for detection of Escherichia coli O157:H7 [J].
Fei Liu ;
Yanbin Li ;
Xiao-Li Su ;
Michael F. Slavik ;
Yibin Ying ;
Jianping Wang .
Sensing and Instrumentation for Food Quality and Safety, 2007, 1 (4) :161-168
[8]  
Feng P., 2002, ENUMERATION ESCHERIC
[9]   The detection of influenza A and B viruses in clinical specimens using a quartz crystal microbalance [J].
Hewa, Thamara M. Peduru ;
Tannock, Gregory A. ;
Mainwaring, David E. ;
Harrison, Sally ;
Fecondo, John V. .
JOURNAL OF VIROLOGICAL METHODS, 2009, 162 (1-2) :14-21
[10]  
Hitchins A D., 2004, BAM Chapter 10: Detection of Listeria monocytogenes in foods and environmental samples, and enumeration of Listeria monocytogenes in foods