Is it safe to cut pacing wires flush with the skin instead of removing them?

被引:31
作者
Shaikhrezai, Kasra [1 ]
Khorsandi, Maziar [1 ]
Patronis, Marios [1 ]
Prasad, Sai [1 ]
机构
[1] Royal Infirm Edinburgh NHS Trust, Dept Cardiothorac Surg, Edinburgh EH16 4SA, Midlothian, Scotland
关键词
Epicardial pacing wire; Cardiac pacemaker; Complications; Migration; MIGRATION; FRAGMENT;
D O I
10.1093/icvts/ivs397
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A best evidence topic in cardiac surgery was written according to a structured protocol. The question addressed was whether it is safe to cut the temporary epicardial pacing wires (TEPWs) flush with the patient's skin surface prior to discharge. Altogether 105 relevant papers were identified of which 13 case reports represented the best evidence to answer the question. The author, journal, date, country of publication, complications, the culprit TEPW and relevant outcomes are tabulated. All case reports demonstrated a wide spectrum of complications. Complications from a retained TEPW mainly arise after a long dormant period. A recent case report has demonstrated the herniation of intra-abdominal contents through a diaphragmatic defect created by the abandoned epicardial pacing wires after a few decades. In multiple case reports, the migration of TEPW was the culprit of serious complications. In two case reports, the TEPWs attached to the right chambers of the heart had migrated to the pulmonary artery via the right atrium and then the right ventricle. In one case report, a similar migration of the right ventricular TEPW to the right ventricular outflow track was observed. The TEPW migration was not limited to the right side of the heart, as in one case report the right atrial TEPW had migrated to the right carotid artery via the ascending aorta. A distant extravascular migration of TEPWs to the skin surface and intraperitoneal and pelvic cavities has also been reported. Retained TEPWs have also been reported to inflict complications locally. One case report has shown a large right-sided para-cardiac mass caused by a right atrial TEPW. In two other case reports, the bronchocutaneous fistula, lobar consolidation and bronchiectasis were the manifestations of a retained TEPW. We conclude that the retention of TEPW after cardiac surgery is not necessarily safe and may cause severe complications. We recommend that TEPWs should be completely removed when possible. If TEPWs are retained, this should be appropriately documented and the surgeon should be mindful of this when the patient presents with complications postoperatively.
引用
收藏
页码:1047 / 1051
页数:5
相关论文
共 19 条
[1]   Discovery and Management of Diaphragmatic Hernia Related to Abandoned Epicardial Pacemaker Wires in a Pregnant Woman with {S,L,L} Transposition of the Great Arteries [J].
Benson, Craig C. ;
Valente, Anne M. ;
Economy, Katherine E. ;
Hoffman-Sage, Yael ;
Bevilacqua, Laura M. ;
Podovei, Mihaela ;
Opotowsky, Alexander R. .
CONGENITAL HEART DISEASE, 2012, 7 (02) :183-188
[2]   Determining the utility of temporary pacing wires after coronary artery bypass surgery [J].
Bethea, BT ;
Salazar, JD ;
Grega, MA ;
Doty, JR ;
Fitton, TP ;
Alejo, DE ;
Borowicz, LM ;
Gott, VL ;
Sussman, MS ;
Baumgartner, WA .
ANNALS OF THORACIC SURGERY, 2005, 79 (01) :104-107
[3]  
Carroll K C, 1998, Am J Crit Care, V7, P444
[4]  
Del Nido P, 1989, J Card Surg, V4, P99, DOI 10.1111/j.1540-8191.1989.tb00262.x
[5]  
Dunning Joel, 2003, Interact Cardiovasc Thorac Surg, V2, P405, DOI 10.1016/S1569-9293(03)00191-9
[6]  
García-Bengochea J, 2007, TEX HEART I J, V34, P376
[7]   COMPLICATIONS OF RETAINED EPICARDIAL FACING WIRES - AN UNUSUAL BRONCHIAL FOREIGN-BODY [J].
GENTRY, WH ;
HASSAN, AA .
ANNALS OF THORACIC SURGERY, 1993, 56 (06) :1391-1393
[8]  
Hodam R P, 1969, Ann Thorac Surg, V8, P506
[9]  
Horng GS, 2008, ANN THORAC SURG, V86, P1352, DOI 10.1016/j.athoracsur.2008.03.013
[10]   Bizarre case of migration of a retained epicardial pacing wire [J].
Juchem, Gerd ;
Golczyk, Karl ;
Kopf, Carsten ;
Reichart, Bruno ;
Lamm, Peter .