3D image-guided robotic needle positioning system for small animal interventions

被引:9
作者
Bax, Jeffrey S. [1 ,2 ]
Waring, Christopher S. R. [1 ,2 ]
Sherebrin, Shi [2 ]
Stapleton, Shawn [3 ]
Hudson, Thomas J. [2 ]
Jaffray, David A. [3 ]
Lacefield, James C. [1 ,2 ]
Fenster, Aaron [1 ,2 ]
机构
[1] Univ Western Ontario, Biomed Engn Grad Program, London, ON N6A 5B9, Canada
[2] Univ Western Ontario, Robarts Res Inst, London, ON N6A 5K8, Canada
[3] Univ Toronto, Radiat Med Program, Princess Margaret Hosp, Toronto, ON M5G 2M9, Canada
基金
加拿大自然科学与工程研究理事会; 加拿大健康研究院;
关键词
medical robotics; image guidance; ULTRASOUND; GUIDANCE; SURGERY; MICE; INJECTIONS; ASSISTANT; BIOPSY;
D O I
10.1118/1.4771958
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This paper presents the design of a micro-CT guided small animal robotic needle positioning system. In order to simplify the robotic design and maintain a small targeting error, a novel implementation of the remote center of motion is used in the system. The system has been developed with the objective of achieving a mean targeting error of <200 mu m while maintaining a high degree of user friendliness. Methods: The robot is compact enough to operate within a 25 cm diameter micro-CT bore. Small animals can be imaged and an intervention performed without the need to transport the animal from one workspace to another. Not requiring transport of the animal reduces opportunities for targets to shift from their localized position in the image and simplifies the workflow of interventions. An improved method of needle calibration is presented that better characterizes the calibration using the position of the needle tip in photographs rather than the needle axis. A calibration fixture was also introduced, which dramatically reduces the time requirements of calibration while maintaining calibration accuracy. Two registration modes have been developed to correspond the robot coordinate system with the coordinate system of the micro-CT scanner. The two registration modes offer a balance between the time required to complete a registration and the overall registration accuracy. The development of slow high accuracy and fast low accuracy registration modes provides users with a degree of flexibility in selecting a registration mode best suited for their application. Results: The target registration error (TRE) of the higher accuracy primary registration was TREprimary = 31 +/- 12 mu m. The error in the lower accuracy combined registration was TREcombined = 139 +/- 63 mu m. Both registration modes are therefore suitable for small-animal needle interventions. The targeting accuracy of the robotic system was characterized using targeting experiments in tissue-mimicking gelatin phantoms. The results of the targeting experiments were combined with the known calibration and needle deflection errors to provide a more meaningful measure of the needle positioning accuracy of the system. The combined targeting errors of the system were 149 +/- 41 mu m and 218 +/- 38 mu m using the primary and combined registrations, respectively. Finally, pilot in vivo experiments were successfully completed to demonstrate the performance of the system in a biomedical application. Conclusions: The device was able to achieve the desired performance with an error of <200 mu m and improved repeatability when compared to other designs. The device expands the capabilities of image-guided interventions for preclinical biomedical applications. (C) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4771958]
引用
收藏
页数:15
相关论文
共 30 条
[1]   LEAST-SQUARES FITTING OF 2 3-D POINT SETS [J].
ARUN, KS ;
HUANG, TS ;
BLOSTEIN, SD .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1987, 9 (05) :699-700
[2]   In vivo small-animal imaging using micro-CT and digital subtraction angiography [J].
Badea, C. T. ;
Drangova, M. ;
Holdsworth, D. W. ;
Johnson, G. A. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (19) :R319-R350
[3]   Small animal computed tomography Imaging [J].
Bartling, Soenke H. ;
Stiller, Wolfram ;
Semmler, Wolfhard ;
Kiessling, Fabian .
CURRENT MEDICAL IMAGING, 2007, 3 (01) :45-59
[4]   Mechanically assisted 3D ultrasound guided prostate biopsy system [J].
Bax, Jeffrey ;
Cool, Derek ;
Gardi, Lori ;
Knight, Kerry ;
Smith, David ;
Montreuil, Jacques ;
Sherebrin, Shi ;
Romagnoli, Cesare ;
Fenster, Aaron .
MEDICAL PHYSICS, 2008, 35 (12) :5397-5410
[5]   Design of a Parallel Robot for Needle-Based Interventions on Small Animals [J].
Bebek, Ozkan ;
Hwang, Myun Joong ;
Cavusoglu, M. Cenk .
IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2013, 18 (01) :62-73
[6]   An alternative method for intrathymic injections in mice [J].
Blair-Handon, Robin ;
Mueller, Kristen ;
Hoogstraten-Miller, Shelley .
LAB ANIMAL, 2010, 39 (08) :248-252
[7]   A robotic system for 18F-FMISO PET-guided intratumoral pO2 measurements [J].
Chang, Jenghwa ;
Wen, Bixiu ;
Kazanzides, Peter ;
Zanzonico, Pat ;
Finn, Ronald D. ;
Fichtinger, Gabor ;
Ling, C. Clifton .
MEDICAL PHYSICS, 2009, 36 (11) :5301-5309
[8]   Robotic surgery in oral and maxillofacial, craniofacial and head and neck surgery: A systematic review of the literature [J].
De Ceulaer, J. ;
De Clercq, C. ;
Swennen, G. R. J. .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2012, 41 (11) :1311-1324
[9]   Refinement of intrathymic injection in mice [J].
de la Cueva, Teresa ;
Naranjo, Angel ;
de la Cueva, Ernesto ;
Rubio, Daniel .
LAB ANIMAL, 2007, 36 (05) :27-32
[10]   Islet Cells Contribute to Pancreatic Carcinogenesis in an Animal Model [J].
El-Ghamari, Mohamed ;
Bergmann, Frank ;
Schmied, Bruno M. ;
Weitz, Juergen ;
Ulrich, Alexis .
PANCREAS, 2011, 40 (02) :242-246