Dosimetric and radiobiological comparison of TG-43 and Monte Carlo calculations in 192Ir breast brachytherapy applications

被引:16
作者
Peppa, V. [1 ]
Pappas, E. P. [1 ]
Karaiskos, P. [1 ]
Major, T. [2 ]
Polgar, C. [2 ]
Papagiannis, P. [1 ]
机构
[1] Univ Athens, Sch Med, Med Phys Lab, Athens, Greece
[2] Natl Inst Oncol, Budapest, Hungary
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2016年 / 32卷 / 10期
关键词
TG-43; MBDCA; Ir-192; Breast; Brachytherapy; Dosimetry; AIDED DOSIMETRY; PARAMETERS; IRRADIATION; ACCURACY;
D O I
10.1016/j.ejmp.2016.09.020
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the clinical significance of introducing model based dose calculation algorithms (MBDCAs) as an alternative to TG-43 in Ir-192 interstitial breast brachytherapy. Materials and methods: A 57 patient cohort was used in a retrospective comparison between TG-43 based dosimetry data exported from a treatment planning system and Monte Carlo (MC) dosimetry performed using MCNP v. 6.1 with plan and anatomy information in DICOM-RT format. Comparison was performed for the target, ipsilateral lung, heart, skin, breast and ribs, using dose distributions, dose-volume histograms (DVH) and plan quality indices clinically used for plan evaluation, as well as radiobiological parameters. Results: TG-43 overestimation of target DVH parameters is statistically significant but small (less than 2% for the target coverage indices and 4% for homogeneity indices, on average). Significant dose differences (>5%) were observed close to the skin and at relatively large distances from the implant leading to a TG-43 dose overestimation for the organs at risk. These differences correspond to low dose regions (<50% of the prescribed dose), being less than 2% of the prescribed dose. Detected dosimetric differences did not induce clinically significant differences in calculated tumor control probabilities (mean absolute difference <0.2%) and normal tissue complication probabilities. Conclusion: While TG-43 shows a statistically significant overestimation of most indices used for plan evaluation, differences are small and therefore not clinically significant. Improved MBDCA dosimetry could be important for re-irradiation, technique inter-comparison and/or the assessment of secondary cancer induction risk, where accurate dosimetry in the whole patient anatomy is of the essence. (C) 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1245 / 1251
页数:7
相关论文
共 31 条
[1]   Beta versus gamma dosimetry close to Ir-192 brachytherapy sources [J].
Baltas, D ;
Karaiskos, P ;
Papagiannis, P ;
Sakelliou, L ;
Loeffler, E ;
Zamboglou, N .
MEDICAL PHYSICS, 2001, 28 (09) :1875-1882
[2]   Morbidity related to axillary irradiation in the treatment of breast cancer [J].
Bentzen, SM ;
Dische, S .
ACTA ONCOLOGICA, 2000, 39 (03) :337-347
[3]   Concepts for critical organ dosimetry in three-dimensional image-based breast brachytherapy [J].
Berger, Daniel ;
Kauer-Dorner, Daniela ;
Seitz, Wolfgang ;
Poetter, Richard ;
Kirisits, Christian .
BRACHYTHERAPY, 2008, 7 (04) :320-326
[4]   Comparison of three accelerated partial breast irradiation techniques: Treatment effectiveness based upon biological models [J].
Bovi, Joseph ;
Qj, X. Sharon ;
White, Julia ;
Li, X. Allen .
RADIOTHERAPY AND ONCOLOGY, 2007, 84 (03) :226-232
[5]   Calculation of integrated biological response in brachytherapy [J].
Dale, RG ;
Coles, IP ;
Deehan, C ;
ODonoghue, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03) :633-642
[6]   Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source [J].
Daskalov, GM ;
Loffler, E ;
Williamson, JF .
MEDICAL PHYSICS, 1998, 25 (11) :2200-2208
[7]   A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: Report of AAPM Task Group No. 138 and GEC-ESTRO [J].
DeWerd, Larry A. ;
Ibbott, Geoffrey S. ;
Meigooni, Ali S. ;
Mitch, Michael G. ;
Rivard, Mark J. ;
Stump, Kurt E. ;
Thomadsen, Bruce R. ;
Venselaar, Jack L. M. .
MEDICAL PHYSICS, 2011, 38 (02) :782-801
[8]   Analysis of a large number of clinical studies for breast cancer radiotherapy: estimation of radiobiological parameters for treatment planning [J].
Guerrero, M ;
Li, XA .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (20) :3307-3326
[9]   Review of clinical brachytherapy uncertainties: Analysis guidelines of GEC-ESTRO and the AAPM [J].
Kirisits, Christian ;
Rivard, Mark J. ;
Baltas, Dimos ;
Ballester, Facundo ;
De Brabandere, Marisol ;
van der Laarse, Rob ;
Niatsetski, Yury ;
Papagiannis, Panagiotis ;
Hellebust, Taran Paulsen ;
Perez-Calatayud, Jose ;
Tanderup, Kari ;
Venselaar, Jack L. M. ;
Siebert, Frank-Andre .
RADIOTHERAPY AND ONCOLOGY, 2014, 110 (01) :199-212
[10]   Validation of the Oncentra Brachy Advanced Collapsed cone Engine for a commercial 192Ir source using heterogeneous geometries [J].
Ma, Yunzhi ;
Lacroix, Frederic ;
Lavallee, Marie-Claude ;
Beaulieu, Luc .
BRACHYTHERAPY, 2015, 14 (06) :939-952