In vivo and in situ ischemic tissue characterization using electrical impedance spectroscopy

被引:75
作者
Casas, O [1 ]
Bragós, R
Riu, PJ
Rosell, J
Tresànchez, M
Warren, M
Rodriguez-Sinovas, A
Carreño, A
Cinca, J
机构
[1] Univ Politecn Cataluna, Dept Elect Engn, Div Instrumentacio & Bioengn, ES-08034 Barcelona, Spain
[2] Univ Vall Hebron, Gen Hosp, Serv Cardiol, Lab Cardiol Expt, Barcelona 08035, Spain
来源
ELECTRICAL BIOIMPEDANCE METHODS: APPLICATIONS TO MEDICINE AND BIOTECHNOLOGY | 1999年 / 873卷
关键词
D O I
10.1111/j.1749-6632.1999.tb09448.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The investigation of processes of ischemia in different organ tissues is very important for the development of methods of protection and preservation during surgical procedures. Electrical impedance spectroscopy was used to distinguish between different tissues and their degree of ischemia. We describe mathematical methods used to adjust experimental data to Cole-Cole models for one-circle and two-circle impedance loci and a study of the main parameters for representing the behavior of ischemia in time. In vivo and in situ postmortem measurements of different tissues from pigs are shown in the 100 Hz to 1 MHz range. The Cole parameters that best characterize the ischemia are R-0 and f(c).
引用
收藏
页码:51 / 58
页数:8
相关论文
共 13 条
[1]  
BRAGOS R, 1996, 14 C AN SOC ESP ING, P97
[2]  
BRAGOS R, 1996, 18 ANN INT C IEEE EN, P414
[3]  
CINCA J, 1998, IN PRESS CARDIOVASC
[4]   The dielectric properties of biological tissues .1. Literature survey [J].
Gabriel, C ;
Gabriel, S ;
Corthout, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2231-2249
[5]   The dielectric properties of biological tissues .2. Measurements in the frequency range 10 Hz to 20 GHz [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2251-2269
[6]   The dielectric properties of biological tissues .3. Parametric models for the dielectric spectrum of tissues [J].
Gabriel, S ;
Lau, RW ;
Gabriel, C .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2271-2293
[7]  
GERSING E, 1995, INNOV TECH BIOL MED, V16, P672
[8]  
LINHART HG, 1995, 9 INT C EL BIOIMP HE, P207
[9]   Parametric modelling for electrical impedance spectroscopy system [J].
Lu, L ;
Hamzaoui, L ;
Brown, BH ;
Rigaud, B ;
Smallwood, RH ;
Barber, DC ;
Morucci, JP .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1996, 34 (02) :122-126
[10]  
McAdams E., 1995, INNOV TECH BIOL MED, V16, P662