Influence of imaging source and panel position uncertainties on the accuracy of 2D/3D image registration of cranial images

被引:5
作者
Warmerdam, Guy [1 ,3 ]
Steininger, Philipp [4 ]
Neuner, Markus [4 ]
Sharp, Gregory [1 ,2 ]
Winey, Brian [1 ,2 ]
机构
[1] Massachusetts Gen Hosp, Boston, MA 02114 USA
[2] Harvard Univ, Sch Med, Boston, MA 02115 USA
[3] Eindhoven Univ Technol, Sch Med Phys & Engn, NL-5600 MB Eindhoven, Netherlands
[4] Paracelsus Med Univ, Inst Res & Dev Adv Radiat Technol, A-5020 Salzburg, Austria
关键词
2D/3D registration; image calibration; CONE-BEAM CT; COMPUTED-TOMOGRAPHY; GEOMETRIC CALIBRATION; QUALITY-ASSURANCE; IMPLEMENTATION; GUIDANCE; SYSTEMS;
D O I
10.1118/1.4742866
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To determine the effects of imager source and panel positioning uncertainties on the accuracy of dual intensity-based 2D/3D image registration of cranial images. Methods: An open source 2D/3D image registration algorithm has been developed for registration of two orthogonal x-rays to a 3D volumetric image. The initialization files of the algorithm allow for nine degrees of freedom system calibration including,v, y, z positions of the source and panel, and three rotational degrees of freedom of the panel about each of the three translational axes. A baseline system calibration was established and a baseline 2D/3D registration between two orthogonal x-rays and the volumetric image was determined. The calibration file was manipulated to insert errors into each of the nine calibration variables of both imager geometries. Rigid six degrees of freedom registrations were iterated for each panel or source positional error over a range of predetermined calibration errors to determine the resulting error in the registration versus the baseline registration due to the manipulated error of the panel or source calibration. Results: Panel and source translational errors orthogonal to the imager/panel axis introduced the greatest errors in the registration accuracy (4.0 mm geometric error results in up to 2.7 mm registration error). Panel rotation about the imaging direction also resulted in errors of the registration (2.0 degrees geometric error results in up to 1.7 degrees registration error). Differences in magnification and panel tilt and roll, i.e., source and/or panel translation along the imaging direction and panel rotations about the orthogonal axes had minimal effects on the registration accuracy (below 0.3 mm and 0.2 degrees registration error). Conclusions: While five of the nine imaging system variables were found to have a considerable effect on 2D/3D registration accuracy of cranial images, the other four variables showed minimal effects. Vendors typically provide simplified calibration procedures which aim to remove encountered geometric uncertainties by accounting for two panel translations. This study shows that at least the five relevant positional variables should be separately calibrated, if accurate alignment is required for 2D/3D registration. (c) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4742866]
引用
收藏
页码:5547 / 5556
页数:10
相关论文
共 33 条
  • [1] Evaluation of the effects of sagging shifts on isocenter accuracy and image quality of cone-beam CT from kV on-board imagers
    Ali, Imad
    Ahmad, Salahuddin
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (03): : 180 - 194
  • [2] Rigid 2D/3D slice-to-volume registration and its application on fluoroscopic CT images
    Birkfellner, Wolfgang
    Figl, Michael
    Kettenbach, Joachim
    Hummel, Johann
    Homolka, Peter
    Schernthaner, Ruediger
    Nau, Thomas
    Bergmann, Helmar
    [J]. MEDICAL PHYSICS, 2007, 34 (01) : 246 - 255
  • [3] Quality assurance for the geometric accuracy of cone-beam CT guidance in radiation therapy
    Bissonnette, Jean-Pierre
    Moseley, Doug
    White, Elizabeth
    Sharpe, Michael
    Purdie, Tom
    Jaffray, David A.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01): : S57 - S61
  • [4] Automatic 3D-to-2D registration for CT and dual-energy digital radiography for calcification detection
    Chen, Xiang
    Gilkeson, Robert C.
    Fei, Baowei
    [J]. MEDICAL PHYSICS, 2007, 34 (12) : 4934 - 4943
  • [5] EVALUATION OF RADIATION DOSE AND IMAGE QUALITY FOR THE VARIAN CONE BEAM COMPUTED TOMOGRAPHY SYSTEM
    Cheng, Harry C. Y.
    Wu, Vincent W. C.
    Liu, Eva S. F.
    Kwong, Dora L. W.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (01): : 291 - 300
  • [6] Accurate technique for complete geometric calibration of cone-beam computed tomography systems
    Cho, YB
    Moseley, DJ
    Siewerdsen, JH
    Jaffray, DA
    [J]. MEDICAL PHYSICS, 2005, 32 (04) : 968 - 983
  • [7] Use of an amorphous silicon EPID for measuring MLC calibration at varying gantry angle
    Clarke, M. F.
    Budgell, G. J.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (02) : 473 - 485
  • [8] Claus BEH., 2006, PHYS MED IMAGING, V6142, P823
  • [9] Geometric calibration of a mobile C-arm for intraoperative cone-beam CT
    Daly, M. J.
    Siewerdsen, J. H.
    Cho, Y. B.
    Jaffray, D. A.
    Irish, J. C.
    [J]. MEDICAL PHYSICS, 2008, 35 (05) : 2124 - 2136
  • [10] Commissioning and clinical implementation of a mega-voltage cone beam CT system for treatment localization
    Gayou, Olivier
    Miften, Moyed
    [J]. MEDICAL PHYSICS, 2007, 34 (08) : 3183 - 3192