Evaluation of the ArcCHECK QA system for IMRT and VMAT verification

被引:89
作者
Li, Guangjun [1 ,2 ]
Zhang, Yingjie [1 ,2 ]
Jiang, Xiaoqin [1 ,2 ]
Bai, Sen [1 ,2 ]
Peng, Guang [1 ,2 ]
Wu, Kui [1 ,2 ]
Jiang, Qingfeng [1 ,2 ]
机构
[1] Sichuan Univ, West China Hosp, Ctr Canc, Radiat Phys Ctr, Chengdu 610064, Sichuan, Peoples R China
[2] Sichuan Univ, West China Hosp, State Key Lab Biotherapy, Chengdu 610064, Sichuan, Peoples R China
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2013年 / 29卷 / 03期
基金
中国国家自然科学基金;
关键词
IMRT; VMAT; Diode array; Quality assurance; VOLUMETRIC MODULATED ARC; RADIATION-THERAPY COMMITTEE; DIODE-ARRAY DOSIMETER; STEP-AND-SHOOT; QUALITY-ASSURANCE; IONIZATION-CHAMBER; RADIOTHERAPY; CALIBRATION; DELIVERY; PHANTOM;
D O I
10.1016/j.ejmp.2012.04.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purposes of this study were to perform tests for the ArcCHECK QA system, and to evaluate the suitability of this system for IMRT and VMAT verification. The device was tested for short term reproducibility, dose linearity, dose rate dependence, dose per pulse dependence, field size dependence, out of field dependence and directional dependence. Eight simple plans that each used four beams of different field sizes as well as IMRT and VMAT plans for various organs of 10 patients were measured by ArcCHECK. The phantom data was then compared with ion chamber measurements and planned results. The ArcCHECK diodes performed well for all tests except directional dependence, which varies from a minimum of -4.9% (seen only when the beam is incident on the diode at 180 degrees degrees) to a maximum of 9.1% (approximately at 105 degrees). For simple plan verification, the absolute dose pass rates of gamma index (3%/3 mm) were almost identical. They had an average pass rate of 94.6% +/- 1.3% when the field size was <= 20 cm in the X direction (right to left direction), but the pass rate fell rapidly when the field size was >20 cm in the X direction. For all patient-specific IMRT and VMAT QA, the pass rates exceeded 95% and 93%, respectively, and high reproducibility of these results has been observed from week to week. The comparative measurements show that the ArcCHECK QA system is completely suitable for clinical IMRT and VMAT verification. (C) 2012 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:295 / 303
页数:9
相关论文
共 24 条
[1]   Dosimetric characterization of a large area pixel-segmented ionization chamber [J].
Amerio, S ;
Boriano, A ;
Bourhaleb, F ;
Cirio, R ;
Donetti, M ;
Fidanzio, A ;
Garelli, E ;
Giordanengo, S ;
Madon, E ;
Marchetto, F ;
Nastasi, U ;
Peroni, C ;
Piermattei, A ;
Freire, CJS ;
Sardo, A ;
Trevisiol, E .
MEDICAL PHYSICS, 2004, 31 (02) :414-420
[2]  
[Anonymous], 2004, TECHN REP SER TRS, V430
[3]   Evaluation of the Delta4 phantom for IMRT and VMAT verification [J].
Bedford, James L. ;
Lee, Young K. ;
Wai, Philip ;
South, Christopher P. ;
Warrington, Alan P. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) :N167-N176
[4]   COMMISSIONING OF VOLUMETRIC MODULATED ARC THERAPY (VMAT) [J].
Bedford, James L. ;
Warrington, Alan P. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (02) :537-545
[5]   Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee [J].
Ezzell, GA ;
Galvin, JM ;
Low, D ;
Palta, JR ;
Rosen, I ;
Sharpe, MB ;
Xia, P ;
Xiao, Y ;
Xing, L ;
Yu, CX .
MEDICAL PHYSICS, 2003, 30 (08) :2089-2115
[6]   Evaluation of a new VMAT QA device, or the "X" and "O" array geometries [J].
Feygelman, Vladimir ;
Zhang, Geoffrey ;
Stevens, Craig ;
Nelms, Benjamin E. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02) :146-168
[7]   Evaluation of a biplanar diode array dosimeter for quality assurance of step-and-shoot IMRT [J].
Feygelman, Vladimir ;
Forster, Kenneth ;
Opp, Daniel ;
Nilsson, Gorgen .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2009, 10 (04) :64-78
[8]   American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning [J].
Fraass, B ;
Doppke, K ;
Hunt, M ;
Kutcher, G ;
Starkschall, G ;
Stern, R ;
Van Dyke, J .
MEDICAL PHYSICS, 1998, 25 (10) :1773-1829
[9]   Is a single arc sufficient in volumetric-modulated arc therapy (VMAT) for complex-shaped target volumes? [J].
Guckenberger, Matthias ;
Richter, Anne ;
Krieger, Thomas ;
Wilbert, Juergen ;
Baier, Kurt ;
Flentje, Michael .
RADIOTHERAPY AND ONCOLOGY, 2009, 93 (02) :259-265
[10]   A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery [J].
Jursinic, PA ;
Nelms, BE .
MEDICAL PHYSICS, 2003, 30 (05) :870-879