An analytical application of the electrocatalysis of the iodate reduction at tungsten oxide films

被引:25
作者
da Rocha, JRC [1 ]
Ferreira, TL [1 ]
Torresi, RM [1 ]
Bertotti, M [1 ]
机构
[1] Univ Sao Paulo, Inst Quim, BR-05599970 Sao Paulo, SP, Brazil
基金
巴西圣保罗研究基金会;
关键词
electrochemical sensors; iodate; tungsten oxide; flow injection; amperometric detection; salt samples;
D O I
10.1016/j.talanta.2005.09.010
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Films of non-stoichiometric tungsten oxides have been deposited onto glassy carbon surfaces by electrodeposition from acidic W(VI) solutions and the chemical stability of these oxides was investigated by using the electrochemical quartz crystal microbalance. At these modified surfaces, rotating disc electrode voltammetric experiments indicated that iodate is electrocatalytically reduced in a mass-transport controlled process. The influence of the film thickness on the response to iodate was investigated and the results indicated a linear relationship between catalytic current and film thickness for relatively thin oxide layers. The modified electrode was employed successfully as an amperometric sensor for iodate in a flow injection apparatus. The linear response of the developed method is extended from 5 mu mol L-1 to 5 mmol L-1 iodate with a limit of detection (signal-to-noise = 3) of 1.2 mu mol L-1. The repeatability of the method for 41 injections of a 1 mmol L-1 iodate solution was 0.8% and the throughput was determined as 123 h(-1). Interference from other oxidant anions such as nitrate and nitrite was not noticeable, whereas bromate and chlorate interfere at slight levels. The method was used in the determination of the iodate content in table salt samples. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:148 / 153
页数:6
相关论文
共 34 条
[1]  
BASSET J, 1989, VOGEL TXB QUANTITATI, P384
[2]   Anodic oxidation of oxalic acid using WOx based anodes [J].
Bock, C ;
Smith, A ;
MacDougall, B .
ELECTROCHIMICA ACTA, 2002, 48 (01) :57-67
[3]   The electrochemical oxidation of organics using tungsten oxide based electrodes [J].
Bock, C ;
MacDougall, B .
ELECTROCHIMICA ACTA, 2002, 47 (20) :3361-3373
[4]   Electrochemical and spectroscopic characterization of a tungsten electrode as a sensitive amperometric sensor of small inorganic ions [J].
Casella, IG ;
Contursi, M .
ELECTROCHIMICA ACTA, 2005, 50 (20) :4146-4154
[5]   Potentiometric detection of carboxylic acids by flow injection analysis using a tungsten oxide electrode [J].
Chen, ZL ;
Alexander, PW .
ANALYTICA CHIMICA ACTA, 1996, 332 (2-3) :187-192
[6]   ELECTROCHROMISM IN ANODIC WO3 FILMS .2. OPTICAL AND ELECTROCHROMIC PROPERTIES OF COLORED DISTORTED HEXAGONAL FILMS [J].
DELICHERE, P ;
FALARAS, P ;
HUGOTLEGOFF, A .
THIN SOLID FILMS, 1988, 161 :47-58
[7]   ELECTROCHROMISM OF WO3 IN ACID-SOLUTIONS - AN ELECTROCHEMICAL, OPTICAL AND ELECTROGRAVIMETRIC STUDY [J].
DETORRESI, SIC ;
GORENSTEIN, A ;
TORRESI, RM ;
VAZQUEZ, MV .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1991, 318 (1-2) :131-144
[8]   Germanium dioxide as internal standard for simplified trace determination of bromate, bromide, iodate and iodide by on-line coupling ion chromatography-inductively coupled plasma mass spectrometry [J].
Eickhorst, T ;
Seubert, A .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1050 (01) :103-109
[9]   Sequential flow injection determination of iodate and periodate with spectrophotometric detection [J].
Ensafi, AA ;
Chamjangali, MA .
SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2003, 59 (12) :2897-2903
[10]   CALIBRATION OF THE ELECTROCHEMICAL QUARTZ CRYSTAL MICROBALANCE [J].
GABRIELLI, C ;
KEDDAM, M ;
TORRESI, R .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1991, 138 (09) :2657-2660