MONCH detector enables fast and low-dose free-propagation phase-contrast computed tomography of in situ mouse lungs

被引:9
作者
Dullin, Christian [1 ,2 ]
Albers, Jonas [1 ]
Tromba, Giuliana [2 ]
Andra, Marie [3 ]
Ramilli, Marco [3 ]
Bergamaschi, Anna [3 ]
机构
[1] Univ Med Ctr, Inst Diagnost & Intervent Radiol, Robert Koch Str 40, D-37075 Gottingen, Lower Saxony, Germany
[2] Elettra Sincrotrone Trieste, Str Statale 14,Km 163-5 AREA Sci Pk, I-34149 Trieste, Friuli Venezia, Italy
[3] Paul Scherrer Inst, CH-5232 Villigen, Switzerland
关键词
mouse-lung imaging; low dose phase-contrast computed tomography; charge-integration detector; MICRO-CT; QUANTIFICATION; EXPOSURE; MODELS;
D O I
10.1107/S160057751701668X
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Due to the complexity of the underlying pathomechanism, in vivo mouse lung-disease models continue to be of great importance in preclinical respiratory research. Longitudinal studies following the cause of a disease or evaluating treatment efficacy are of particular interest but challenging due to the small size of the mouse lung and the fast breathing rate. Synchrotron-based in-line phase-contrast computed tomography imaging has been successfully applied in lung research in various applications, but mostly at dose levels that forbid longitudinal in vivo studies. Here, the novel charge-integrating hybrid detector MONCH is presented, which enables imaging of mouse lungs at a pixel size of 25 mu m, in less than 10 s and with an entrance dose of about 70 mGy, which therefore will allow longitudinal lung disease studies to be performed in mouse models.
引用
收藏
页码:565 / 569
页数:5
相关论文
共 24 条
[1]  
Back M., 1955, FUNDAMENTALS RADIOBI, P389
[2]   Retrospective motion Gating in small animal CT of mice and rats [J].
Bartling, Soenke H. ;
Stiller, Wolfram ;
Grasruck, Michael ;
Schmidt, Bernhard ;
Peschke, Peter ;
Semmler, Wolfhard ;
Kiessling, Fabian .
INVESTIGATIVE RADIOLOGY, 2007, 42 (10) :704-714
[3]   Performance of a single photon counting microstrip detector for strip pitches down to 10 μm [J].
Bergamaschi, A. ;
Broennimann, Ch. ;
Dinapoli, R. ;
Eikenberry, E. ;
Gozzo, F. ;
Henrich, B. ;
Kobas, M. ;
Kraft, P. ;
Patterson, B. ;
Schmitt, B. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 591 (01) :163-166
[4]   Looking at single photons using hybrid detectors [J].
Bergamaschi, A. ;
Cartier, S. ;
Dinapoli, R. ;
Greiffenberg, D. ;
Jungmann-Smith, J. H. ;
Mezza, D. ;
Mozzanica, A. ;
Schmitt, B. ;
Shi, X. ;
Tinti, G. .
JOURNAL OF INSTRUMENTATION, 2015, 10
[5]   High-Speed Single-Breath-Hold Micro-Computed Tomography of Thoracic and Abdominal Structures in Mice Using a Simplified Method for Intubation [J].
Boll, Hanne ;
Bag, Simona ;
Schambach, Sebastian J. ;
Doyon, Fabian ;
Nittka, Stefanie ;
Kramer, Martin ;
Groden, Christoph ;
Brockmann, Marc A. .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2010, 34 (05) :783-790
[6]  
Brun F, 2017, ADV STRUCT CHEM IMAG, V3, DOI 10.1186/s40679-016-0036-8
[7]   Micrometer-resolution imaging using MoNCH: towards G2-less grating interferometry [J].
Cartier, Sebastian ;
Kagias, Matias ;
Bergamaschi, Anna ;
Wang, Zhentian ;
Dinapoli, Roberto ;
Mozzanica, Aldo ;
Ramilli, Marco ;
Schmitt, Bernd ;
Bruckner, Martin ;
Frojdh, Erik ;
Greiffenberg, Dominic ;
Mayilyan, Davit ;
Mezza, Davide ;
Redford, Sophie ;
Ruder, Christian ;
Schadler, Lukas ;
Shi, Xintian ;
Thattil, Dhanya ;
Tinti, Gemma ;
Zhang, Jiaguo ;
Stampanoni, Marco .
JOURNAL OF SYNCHROTRON RADIATION, 2016, 23 :1462-1473
[8]   Age-related effects of X-ray irradiation on mouse hippocampus [J].
Casciati, Arianna ;
Dobos, Katalin ;
Antonelli, Francesca ;
Benedek, Anett ;
Kempf, Stefan J. ;
Belles, Montserrat ;
Balogh, Andrea ;
Tanori, Mirella ;
Heredia, Luis ;
Atkinson, Michael J. ;
von Toerne, Christine ;
Azimzadeh, Omid ;
Saran, Anna ;
Safrany, Geza ;
Benotmane, Mohammed A. ;
Victoria Linares-Vidal, M. ;
Tapio, Soile ;
Lumniczky, Katalin ;
Pazzaglia, Simonetta .
ONCOTARGET, 2016, 7 (19) :28040-28058
[9]   X-ray dose delivered during a longitudinal micro-CT study has no adverse effect on cardiac and pulmonary tissue in C57BL/6 mice [J].
Detombe, Sarah A. ;
Dunmore-Buyze, Joy ;
Petrov, Ivailo E. ;
Drangova, Maria .
ACTA RADIOLOGICA, 2013, 54 (04) :435-441
[10]   MONCH, a small pitch, integrating hybrid pixel detector for X-ray applications [J].
Dinapoli, R. ;
Bergamaschi, A. ;
Cartier, S. ;
Greiffenberg, D. ;
Johnson, I. ;
Jungmann, J. H. ;
Mezza, D. ;
Mozzanica, A. ;
Schmitt, B. ;
Shi, X. ;
Tinti, G. .
JOURNAL OF INSTRUMENTATION, 2014, 9