Point-of-care ultrasound versus auscultation in determining the position of double-lumen tube

被引:20
作者
Hu, Wei-Cai [1 ]
Xu, Lei [1 ]
Zhang, Quan [1 ]
Wei, Li [1 ]
Zhang, Wei [2 ]
机构
[1] Henan Prov Peoples Hosp, Dept Thorac Surg, Zhengzhou, Henan, Peoples R China
[2] Henan Prov Peoples Hosp, Dept Anesthesiol, 7 Weiwu Rd, Zhengzhou 450003, Henan, Peoples R China
关键词
anesthesiology; fiberoptic bronchoscopy; point-of-care ultrasound; ENDOTRACHEAL INTUBATION; BRONCHIAL INTUBATION; BEDSIDE ULTRASOUND; CRITICALLY-ILL; MANAGEMENT; PLACEMENT; ULTRASONOGRAPHY; ANESTHESIA; ACCURACY; SURGERY;
D O I
10.1097/MD.0000000000009311
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study was designed to assess the accuracy of point-of-care ultrasound in determining the position of double-lumen tubes (DLTs).A total of 103 patients who required DLT intubation were enrolled into the study. After DLTs were tracheal intubated in the supine position, an auscultation researcher and ultrasound researcher were sequentially invited in the operating room to conduct their evaluation of the DLT. After the end of their evaluation, fiberscope researchers (FRs) were invited in the operating room to evaluate the position of DLT using a fiberscope. After the patients were changed to the lateral position, the same evaluation process was repeated. These 3 researchers were blind to each other when they made their conclusions. Sensitivity, specificity, positive predictive value, negative predictive value, and accuracy were obtained by statistical analysis.When left DLTs (LDLTs) were used, the accuracy of ultrasound (84.2% [72.1%, 92.5%]) was higher than the accuracy of auscultation (59.7% [45.8%, 72.4%]) (P<.01). When right DLTs (RDLTs) were used, the accuracy of ultrasound (89.1% [76.4%, 96.4%]) was higher than the accuracy of auscultation (67.4% [52.0%, 80.5%]) (P<.01). When LDLTs were used in the lateral position, the accuracy of ultrasound (75.4% [62.2%, 85.9%]) was higher than the accuracy of auscultation (54.4% [40.7%, 67.6%]) (P<.05). When RDLT were used, the accuracy of ultrasound (73.9% [58.9%, 85.7%]) was higher than the accuracy of auscultation (47.8% [32.9%, 63.1%]) (P<.05).Assessment via point-of-care ultrasound is superior to auscultation in determining the position of DLTs.
引用
收藏
页数:6
相关论文
共 28 条
  • [1] Direct ultrasound methods: a confirmatory technique for proper endotracheal intubation in the emergency department
    Abbasi, Saeed
    Farsi, Davood
    Zare, Mohammad A.
    Hajimohammadi, Majid
    Rezai, Mahdi
    Hafezimoghadam, Peyman
    [J]. EUROPEAN JOURNAL OF EMERGENCY MEDICINE, 2015, 22 (01) : 10 - 16
  • [2] Alvarez-Díaz N, 2015, REV ESP ANEST REANIM, V62, P305, DOI [10.1016/j.redar.2014.06.005, 10.1016/j.redare.2014.06.001]
  • [3] [Anonymous], 1992, JAMA, V268, P2171
  • [4] Positioning of a double-lumen endobronchial tube without the aid of any instruments: An implication for emergency management
    Bahk, JH
    Lim, YJ
    Kim, CS
    [J]. JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2000, 49 (05): : 899 - 902
  • [5] Point-of-care sonographic detection of left endobronchial main stem intubation and obstruction versus endotracheal intubation
    Blaivas, Michael
    Tsung, James W.
    [J]. JOURNAL OF ULTRASOUND IN MEDICINE, 2008, 27 (05) : 785 - 789
  • [6] Chun Rosaleen, 2004, Prehosp Disaster Med, V19, P366
  • [7] Cohen Edmond, 2004, Curr Opin Anaesthesiol, V17, P1, DOI 10.1097/00001503-200402000-00002
  • [8] An update on anesthesia for thoracoscopic surgery
    Fischer, Gregory W.
    Cohen, Edmond
    [J]. CURRENT OPINION IN ANESTHESIOLOGY, 2010, 23 (01) : 7 - 11
  • [9] Gerscovich EO, 2001, J ULTRAS MED, V20, P597
  • [10] Using lung ultrasound in an infant to detect bronchial intubation not previously identified by auscultation
    Hiruma, Masataka
    Watanabe, Tatsunori
    Baba, Hiroshi
    [J]. CANADIAN JOURNAL OF ANESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2015, 62 (10): : 1121 - 1122