Context sensitive cardiac x-ray imaging: a machine vision approach to x-ray dose control

被引:4
作者
Kengyelics, Stephen M. [1 ]
Gislason-Lee, Amber J. [1 ]
Keeble, Claire [1 ]
Magee, Derek R. [2 ]
Davies, Andrew G. [1 ]
机构
[1] Univ Leeds, Leeds Inst Cardiovasc & Metab Med, Div Biomed Imaging, Leeds LS2 9JT, W Yorkshire, England
[2] Univ Leeds, Sch Comp, Leeds LS2 9JT, W Yorkshire, England
关键词
cardiac; x-ray; contrast; machine vision; PERCUTANEOUS CORONARY INTERVENTION;
D O I
10.1117/1.JEI.24.5.051002
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Modern cardiac x-ray imaging systems regulate their radiation output based on the thickness of the patient to maintain an acceptable signal at the input of the x-ray detector. This approach does not account for the context of the examination or the content of the image displayed. We have developed a machine vision algorithm that detects iodine-filled blood vessels and fits an idealized vessel model with the key parameters of contrast, diameter, and linear attenuation coefficient. The spatio-temporal distribution of the linear attenuation coefficient samples, when appropriately arranged, can be described by a simple linear relationship, despite the complexity of scene information. The algorithm was tested on static anthropomorphic chest phantom images under different radiographic factors and 60 dynamic clinical image sequences. It was found to be robust and sensitive to changes in vessel contrast resulting from variations in system parameters. The machine vision algorithm has the potential of extracting real-time context sensitive information that may be used for augmenting existing dose control strategies. (C) 2015 SPIE and IS&T
引用
收藏
页数:9
相关论文
共 9 条
[1]  
[Anonymous], 2014, MATLAB VERS 8 3 0 53
[2]   Cancer risk related to low-dose ionizing radiation from cardiac imaging in patients after acute myocardial infarction [J].
Eisenberg, Mark J. ;
Afilalo, Jonathan ;
Lawler, Patrick R. ;
Abrahamowicz, Michal ;
Richard, Hugues ;
Pilote, Louise .
CANADIAN MEDICAL ASSOCIATION JOURNAL, 2011, 183 (04) :430-436
[3]  
Frangi AF, 1998, LECT NOTES COMPUT SC, V1496, P130, DOI 10.1007/BFb0056195
[4]   Dose optimization in cardiac x-ray imaging [J].
Gislason-Lee, Amber J. ;
McMillan, Catherine ;
Cowen, Arnold R. ;
Davies, Andrew G. .
MEDICAL PHYSICS, 2013, 40 (09)
[5]   Percutaneous coronary intervention. I: History and development [J].
Grech, ED .
BRITISH MEDICAL JOURNAL, 2003, 326 (7398) :1080-1082
[6]   ABC of interventional cardiology - Percutaneous coronary intervention. II: The procedure [J].
Grech, ED .
BRITISH MEDICAL JOURNAL, 2003, 326 (7399) :1137-1140
[7]  
Ludman P., 2011, NATL AUDIT PERCUTANE
[8]  
*WHO, 2014, TOP 10 CAUS DEATH 10
[9]   Patient peak skin doses from cardiac interventional procedures [J].
Zontar, D. ;
Kuhelj, D. ;
Skrk, D. ;
Zdesar, U. .
RADIATION PROTECTION DOSIMETRY, 2010, 139 (1-3) :262-265