Evaluating the effect of reduced entrance surface dose on neonatal chest imaging using subjective image quality evaluation

被引:12
作者
Conradie, Anita [1 ]
Herbst, Charles Petrus [1 ]
机构
[1] Univ Free State, Dept Med Phys, ZA-9300 Bloemfontein, South Africa
来源
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS | 2016年 / 32卷 / 10期
关键词
Neonatal; Chest X-rays; Dose; Rabbit phantom; Image quality; INTENSIVE-CARE-UNIT; RADIATION-EXPOSURE; RADIOGRAPHY; OPTIMIZATION; REDUCTION;
D O I
10.1016/j.ejmp.2016.07.005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: As hospitalized neonates are radiosensitive and often require numerous X-rays, institutions should investigate the optimal beam parameter combinations to deliver diagnostically acceptable quality images at the lowest possible entrance surface dose (ESD). Using a subjective approach, this study evaluated the effect of different beam parameter combinations on image quality. Methods and materials: Five rabbits simulated the neonatal chest. The ESD was reduced using a variation of voltage, tube current and filtration. Eight radiology registrars, blinded to the dose parameter information, ranked the digital X-ray images of three anatomical regions from best to worst using a variation of the multiple rank order method. T-tests compared the average values, obtained from the scores assigned by each observer to images acquired at different ESDs, for each region. The calculated intraclass correlation coefficient (ICC) assessed observer agreement. Results: Results showed that a 64% dose reduction was achievable by altering the beam parameters. Large variation among the observers was confirmed by an ICC value <0.5. A 95% confidence interval could not conclude that different ESD values, resulting in a 50-77% dose reduction compared to current practice, would result in different overall observed image quality. This was noted for all three regions indicating that no preference existed towards an image acquired with a specific beam parameter combination. Conclusions: The large variation in observers' opinion of acceptable image quality emphasizes the importance of subjective image quality evaluation in the clinical environment. The rabbit phantom and multiple rank order method are considered appropriate for these evaluations. (C) 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1368 / 1374
页数:7
相关论文
共 23 条
  • [11] Groenewald A, 2014, GLOBAL J SCI FRONT R, V14, P33
  • [12] An optimisation strategy in a digital environment applied to neonatal chest imaging
    Hansson, J
    Båth, M
    Håkansson, M
    Grundin, H
    Bjurklint, E
    Orvestad, P
    Kjellström, A
    Boström, H
    Jönsson, M
    Jonsson, K
    Månsson, LG
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 278 - 285
  • [13] Hart D., 2000, Reference Doses and Patient Size in Paediatric Radiology
  • [14] Neonatal chest and abdominal radiation dosimetry: a comparison of two radiographic techniques
    Jones, NF
    Palarm, TW
    Negus, IS
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2001, 74 (886) : 920 - 925
  • [15] Månsson LG, 2000, RADIAT PROT DOSIM, V90, P89, DOI 10.1093/oxfordjournals.rpd.a033149
  • [16] Radiology in the neonatal intensive care unit: Dose reduction and image quality
    McParland, BJ
    Gorka, W
    Lee, R
    Lewall, DB
    Omojola, MF
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1996, 69 (826) : 929 - 937
  • [17] Radiation Exposure to Premature Infants in a Neonatal Intensive Care Unit in Turkey
    Olgar, Turan
    Onal, Esra
    Bor, Dogan
    Okumus, Nurullah
    Atalay, Yildiz
    Turkyilmaz, Canan
    Ergenekon, Ebru
    Koc, Esin
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2008, 9 (05) : 416 - 419
  • [18] Neonatal doses from X ray examinations by birth weight in a neonatal intensive care unit
    Ono, K
    Akahane, K
    Aota, T
    Hada, M
    Takano, Y
    Kai, M
    Kusama, T
    [J]. RADIATION PROTECTION DOSIMETRY, 2003, 103 (02) : 155 - 162
  • [19] Removal of Comfort Pads underneath Babies: A Method of Reducing Radiation Exposure to Neonates
    Rattan, Amit S.
    Cohen, Mervyn D.
    [J]. ACADEMIC RADIOLOGY, 2013, 20 (10) : 1297 - 1300
  • [20] Radiation exposure and image quality in chest CT examinations
    Ravenel, J
    Scalzetti, EM
    Huda, W
    Garrisi, W
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (02) : 279 - 284