Computerized scheme for evaluating mammographic phantom images

被引:8
作者
Asahara, Masaki [1 ,2 ]
Kodera, Yoshie [1 ]
机构
[1] Nagoya Univ, Grad Sch Med Sci, Higashi Ku, Nagoya, Aichi 4618673, Japan
[2] Kagawa Univ Hosp, Dept Radiol, Miki, Kagawa 7610793, Japan
关键词
mammography; quality control; ACR phantom; image quality; image analysis; ACR ACCREDITATION; AMERICAN-COLLEGE; INTERVAL CHANGE; QUALITY; INCONSISTENCIES; SUBTRACTION; CAMPI;
D O I
10.1118/1.3687159
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The authors developed a computer algorithm to automatically evaluate images of the American College of Radiology (ACR) mammography accreditation phantom. Methods: The developed algorithm consist of the edge detection of wax insert, nonuniformity correction of background, and correction for magnification and also calculate the cross-correlation coefficient by image matching technique. The algorithm additionally evaluates target shape for fibers, target contrast for speck groups, and target circularity for masses. To obtain an ideal template image without noise and spatial resolution loss, the wax insert containing the embedded test pattern was extracted from the phantom and radiographed. Two template images and ten test phantom images were prepared for this study. The results of evaluation using the algorithm outputs were compared with the averaged results of observer studies by six skilled observers. Results: In comparing the results from the algorithm outputs with the results of observers, the authors found that the computer outputs were well correlated with the evaluations by observers, and they indicate the quality of the phantom image. The correlation coefficients between results of observer studies and two outputs of computer algorithm, i.e., the cross-correlation coefficient by template matching and indices of target shape for fibers, were 0.89 (95% confidence interval, 0.82-0.93; hereinafter the same) and 0.85 (0.76-0.91). The correlation coefficients between observer's results and two outputs: the cross-correlation coefficient and indices of target contrast for speck groups, were 0.83 (0.79-0.86) and 0.85 (0.81-0.88) and between observer's results and two outputs: the cross-correlation coefficient and indices of target circularity for masses, were 0.90 (0.84-0.94) and 0.87 (0.77-0.92). Conclusions: Image evaluation using the ACR phantom is indispensable in quality control of a mammography system. The proposed algorithm is useful for quality control and image evaluation of mammography units. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.3687159]
引用
收藏
页码:1609 / 1617
页数:9
相关论文
共 20 条
[1]  
Arifin A. Z., P IND JAP JOINT SCI
[2]  
Brooks K. W., 1993, MED PHYS, V20, P881
[3]   SUBJECTIVE EVALUATIONS OF MAMMOGRAPHIC ACCREDITATION PHANTOM IMAGES BY 3 OBSERVER GROUPS [J].
BROOKS, KW ;
TRUEBLOOD, JH ;
KEARFOTT, KJ .
INVESTIGATIVE RADIOLOGY, 1994, 29 (01) :42-47
[4]   Automated analysis of the American College of Radiology mammographic accreditation phantom images [J].
Brooks, KW ;
Trueblood, JH ;
Kearfott, KJ ;
Lawton, DT .
MEDICAL PHYSICS, 1997, 24 (05) :709-723
[5]  
CACAK RK, 1991, RADIOLOGY, V181, P288
[6]   QUANTITATIVE VERSUS SUBJECTIVE EVALUATION OF MAMMOGRAPHY ACCREDITATION PHANTOM IMAGES [J].
CHAKRABORTY, DP ;
ECKERT, MP .
MEDICAL PHYSICS, 1995, 22 (02) :133-143
[7]   Physical measures of image quality in mammography [J].
Chakraborty, DP .
PHYSICS OF MEDICAL IMAGING: MEDICAL IMAGING 1996, 1996, 2708 :179-193
[8]   Computer analysis of mammography phantom images (CAMPI): An application to the measurement of microcalcification image quality of directly acquired digital images [J].
Chakraborty, DP .
MEDICAL PHYSICS, 1997, 24 (08) :1269-1277
[9]   Application of Computer Analysis of Mammography Phantom Images (CAMPI) methodology to the comparison of two digital biopsy machines [J].
Chakraborty, DP ;
Fatouros, PP .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :618-628
[10]  
Destouet Judy M, 2005, J Am Coll Radiol, V2, P585