Charnley low-frictional torque arthroplasty of the hip - 20-to 30-year results

被引:100
作者
Wroblewski, BM [1 ]
Fleming, PA [1 ]
Siney, PD [1 ]
机构
[1] Wrightington Hosp, John Charnley Res Inst, Wigan WN6 9EP, Lancs, England
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1999年 / 81B卷 / 03期
关键词
D O I
10.1302/0301-620X.81B3.9521
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We reviewed 261 patients with 320 Charnley low-friction arthroplasties who had a mean follow-up of 22 years 10 months (20 to 30), Of these, 93.9% considered the operation to be a success; 82.3% were free from pain and 11.6% had occasional discomfort. Satisfactory function was achieved in 59.6% and 62% had an excellent range of movement. The clinical results did not correlate well with the radiological appearance; radiologically loose components did not affect the clinical outcome. The main long-term problem was wear and loosening of the UHMWPE cup. Our findings suggest that the radiological appearance of the arthroplasty is a more reliable indication of the state of the arthroplasty than the clinical results.
引用
收藏
页码:427 / 430
页数:4
相关论文
共 11 条