Optimization of Measurement Arrangements for Magnetic Detection Electrical Impedance Tomography

被引:10
|
作者
Hao, Liling [1 ,2 ]
Li, Gang [3 ]
Lin, Ling [4 ]
机构
[1] Tianjin Univ, Coll Precis Instrument & Optoelect Engn, Tianjin 300072, Peoples R China
[2] Northeastern Univ, Dept SINO DUTCH Biomed & Informat Engn, Shenyang 110819, Liaoning, Peoples R China
[3] Tianjin Univ, Tianjin Key Lab Biomed Detecting Tech & Instrumen, Tianjin 300072, Peoples R China
[4] Tianjin Univ, State Key Lab Precis Measurement Technol & Instru, Tianjin 300072, Peoples R China
基金
中国国家自然科学基金;
关键词
Magnetic detection electrical impedance tomography (MDEIT); optimum measurement arrangement; redundancy reduction; region of interest (RoI); singular value analysis (SVA); IMAGE-RECONSTRUCTION;
D O I
10.1109/TBME.2013.2280632
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Magnetic detection electrical impedance tomography (MDEIT) is an imaging modality that aims to compute the cross-sectional distribution of the conductivity inside a volume. The current is injected into the volume by the surface electrodes and the resulting magnetic fields surrounding the object are detected by coils. The image resolution and contrast in MDEIT image reconstruction are affected by the parameters such as the numbers and locations of electrodes and measurements, and the finite-element mesh resolution. This paper addresses the numerical experiment applied to the singular value analysis (SVA) of the sensitivity matrix in the presence of noisy measurements, subsequently suggesting the optimal electrode and detector configurations for the whole imaging object region. For the region of interest (RoI), the combined SVA and redundancy reduction is used to obtain the optimum measurement arrangement. Finally, the optimum design is confirmed by examining the image reconstructions of the simulated data acquired with different measurement arrangements. The results indicate that properly increasing the number of current injections and the number of measurement circles, and locating preferentially the electrodes and detectors on the region nearest to the RoI produce more useful singular values and better reconstructed images. These results provide guidelines for the design of the MDEIT experimental system.
引用
收藏
页码:444 / 452
页数:9
相关论文
共 50 条
  • [21] A convolutional neural network algorithm for breast tumor detection with magnetic detection electrical impedance tomography
    Chen, Ruijuan
    Zhao, Songsong
    Wu, Weiwei
    Sun, Zhihui
    Wang, Jinhai
    Wang, Huiquan
    Han, Guang
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2021, 92 (06):
  • [22] On the measurement of conductivity distribution of the human head using magnetic resonance electrical impedance tomography
    Gao, N
    Zhu, SN
    He, B
    PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2004, 26 : 4443 - 4446
  • [23] Brain haemorrhage detection using a SVM classifier with electrical impedance tomography measurement frames
    McDermott, Barry
    O'Halloran, Martin
    Porter, Emily
    Santorelli, Adam
    PLOS ONE, 2018, 13 (07):
  • [24] Measurement of ventilation and cardiac related impedance changes with electrical impedance tomography
    Caroline A Grant
    Trang Pham
    Judith Hough
    Thomas Riedel
    Christian Stocker
    Andreas Schibler
    Critical Care, 15
  • [25] Measurement of ventilation and cardiac related impedance changes with electrical impedance tomography
    Grant, Caroline A.
    Trang Pham
    Hough, Judith
    Riedel, Thomas
    Stocker, Christian
    Schibler, Andreas
    CRITICAL CARE, 2011, 15 (01):
  • [26] A Hybrid Device for Electrical Impedance Tomography and Bioelectrical Impedance Spectroscopy Measurement
    Michalikova, Marketa
    Prauzek, Michal
    2014 IEEE 27TH CANADIAN CONFERENCE ON ELECTRICAL AND COMPUTER ENGINEERING (CCECE), 2014,
  • [27] A New Electrode Mode for Magnetic Detection Electrical Impedance Tomography: Computer Simulation Study
    Li, Gang
    Hao, Liling
    Chen, Ruijuan
    Lin, Ling
    IEEE TRANSACTIONS ON MAGNETICS, 2012, 48 (10) : 2543 - 2550
  • [28] Research on Grounding Grids Imaging Reconstruction Based on Magnetic Detection Electrical Impedance Tomography
    Liu Kai
    Yang Fan
    Zhang Songyang
    Zhu Liwei
    Hu Jiayuan
    Wang Xiaoyu
    Ullah, Irfan
    IEEE TRANSACTIONS ON MAGNETICS, 2018, 54 (03)
  • [29] Theoretical simulation of the measurement process of electrical impedance tomography
    Ma, Y
    Miedema, SA
    Vlasblom, WJ
    SYSTEM SIMULATION AND SCIENTIFIC COMPUTING (SHANGHAI), VOLS I AND II, 2002, : 261 - 265
  • [30] THE IMPORTANCE OF PHASE MEASUREMENT IN ELECTRICAL-IMPEDANCE TOMOGRAPHY
    GRIFFITHS, H
    PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (11): : 1435 - 1444