Evaluation of a novel device that maintains the balance of a cardiopulmonary resuscitation performer in a moving ambulance to improve chest compression quality

被引:0
作者
Feng, Shun Yi [1 ]
Song, Ya Qi [1 ]
Zhang, Yu Lin [1 ]
Li, Yong [1 ]
机构
[1] Cangzhou Cent Hosp, Emergency Dept, Cangzhou, Hebei, Peoples R China
关键词
ambulance; cardiopulmonary resuscitation; chest compression; instability; quality; HOSPITAL CARDIAC-ARREST; BLOOD-FLOW; CPR; EMERGENCY;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
INTRODUCTION According to the findings of some studies, instability due to inertia during changes in speed may negatively impact the quality of chest compressions performed during cardiopulmonary resuscitation (CPR) in a moving environment. This study thus aimed to introduce a simple device that maintains the balance of a person performing CPR in a moving environment, such as an ambulance. We also sought to evaluate the effectiveness of this device in the improvement of the quality of chest compressions. METHODS The experiment comprised a total of 40 simulated cardiopulmonary arrest scenes (20 in the experimental group and 20 in the control), in which CPR was conducted by eight paramedics. Each simulation involved two paramedics randomly selected from the eight. The ambulance took the same route from the simulated site to the hospital, and continuous CPR was performed on a manikin in the ambulance with or without the aid of our proposed novel device. RESULTS The average number of chest compressions per simulation in the experimental and control groups was 1330.75 and 1266.60, respectively (p = 0.095). The percentage of chest compressions with adequate depth achieved in the experimental and control groups was 72% +/- 4% and 50% +/- 3%, respectively (p < 0.0001). CONCLUSION By maintaining the balance of the CPR performer, our proposed novel device can offset the negative impact that instability (due to a moving environment) has on chest compressions. The device may also lead to an increase in the percentage of chest compressions that achieve adequate depth.
引用
收藏
页码:645 / 648
页数:4
相关论文
共 15 条
[1]   RELATIONSHIP OF BLOOD-PRESSURE AND FLOW DURING CPR TO CHEST COMPRESSION AMPLITUDE - EVIDENCE FOR AN EFFECTIVE COMPRESSION THRESHOLD [J].
BABBS, CF ;
VOORHEES, WD ;
FITZGERALD, KR ;
HOLMES, HR ;
GEDDES, LA .
ANNALS OF EMERGENCY MEDICINE, 1983, 12 (09) :527-532
[2]   CORONARY BLOOD-FLOW DURING CARDIOPULMONARY RESUSCITATION IN SWINE [J].
BELLAMY, RF ;
DEGUZMAN, LR ;
PEDERSEN, DC .
CIRCULATION, 1984, 69 (01) :174-180
[3]  
EISENBERG MS, 1980, LANCET, V1, P812
[4]   Continuation of CPR on admission to emergency department after out-of-hospital cardiac arrest. Occurrence, characteristics and outcome [J].
Herlitz, J ;
Ekstrom, L ;
Axelsson, A ;
Bang, A ;
Wennerblom, B ;
Waagstein, L ;
Dellborg, M ;
Holmberg, S .
RESUSCITATION, 1997, 33 (03) :223-231
[5]   Changes of aortic dimensions as evidence of cardiac pump mechanism during cardiopulmonary resuscitation in humans [J].
Hwang, SO ;
Lee, KH ;
Cho, JH ;
Yoon, J ;
Choe, KH .
RESUSCITATION, 2001, 50 (01) :87-93
[6]   PREDICTING THE OUTCOME OF UNSUCCESSFUL PREHOSPITAL ADVANCED CARDIAC LIFE-SUPPORT [J].
KELLERMANN, AL ;
HACKMAN, BB ;
SOMES, G .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1993, 270 (12) :1433-1436
[7]   A randomized, controlled comparison of cardiopulmonary resuscitation performed on the floor and on a moving ambulance stretcher [J].
Kim, John A. ;
Vogel, Douglas ;
Guimond, Guy ;
Hostler, David ;
Wang, Henry E. ;
Menegazzi, James J. .
PREHOSPITAL EMERGENCY CARE, 2006, 10 (01) :68-70
[8]   Quality of out-of-hospital cardiopulmonary resuscitation with real time automated feedback: A prospective interventional study [J].
Kramer-Johansen, Jo ;
Myklebust, Helge ;
Wik, Lars ;
Fellows, Bob ;
Svensson, Leif ;
Sorebo, Hallstein ;
Steen, Petter Andreas .
RESUSCITATION, 2006, 71 (03) :283-292
[9]   The effect of transport on quality of cardiopulmonary resuscitation in out-of-hospital cardiac arrest [J].
Odegaard, Silje ;
Olasveengen, Theresa ;
Steen, Petter Andreas ;
Kramer-Johansen, Jo .
RESUSCITATION, 2009, 80 (08) :843-848
[10]   MECHANISMS OF BLOOD-FLOW DURING CARDIOPULMONARY RESUSCITATION [J].
RUDIKOFF, MT ;
MAUGHAN, WL ;
EFFRON, M ;
FREUND, P ;
WEISFELDT, ML .
CIRCULATION, 1980, 61 (02) :345-352