Combined local impedance and contact force for radiofrequency ablation assessment

被引:38
|
作者
Garrott, Kara [1 ]
Laughner, Jacob [1 ]
Gutbrod, Sarah [1 ]
Sugrue, Alan [2 ]
Shuros, Allan [1 ]
Sulkin, Matt [1 ]
Yasin, Omar [2 ]
Bush, Jamie [1 ]
Pottinger, Nathan [1 ]
Meyers, Jason [3 ]
Kapa, Suraj [2 ]
机构
[1] Boston Sci Corp, St Paul, MN USA
[2] Mayo Clin, Dept Cardiovasc Med, Coll Med, 200 First St SW, Rochester, MN 55905 USA
[3] Iowa Heart Ctr, Heart Rhythm Ctr, W Des Moines, IA USA
关键词
Ablation; Contact force; Electroanatomic mapping; Impedance; Radiofrequency; CATHETER ABLATION; LESION FORMATION; TEMPERATURE; SAFETY; MODEL; SIZE; FLOW;
D O I
10.1016/j.hrthm.2020.03.016
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND The combination of contact force (CF) and local impedance (LI) may improve tissue characterization and lesion prediction during radiofrequency (RF) ablation. OBJECTIVE The purpose of this study was to evaluate the utility of LI combined with CF in assessing RF ablation efficacy. METHODS An LI catheter with CF sensing was evaluated in swine (n = 11) and in vitro (n = 14). The relationship between LI and CF in different tissue types was evaluated in vivo. Discrete lesions were created in vitro and in vivo at a range of forces, powers, and durations. Finally, an intercaval line was created in 3 groups at 30 W: 30s, Delta 20 Omega, and Delta 30 Omega. In the Delta 20 Omega and Delta 30 Omega groups, the user ablated until a 20 or 30 Omega LI drop. In the 30s group, the user was blinded to LI. RESULTS In vivo, distinction in LI was found between the blood pool and the myocardium (blood pool: 122 +/- 7.02 Omega; perpendicular contact: 220 +/- 29 Omega; parallel contact: 207 +/- 31 Omega). LI drop correlated with lesion depth both in vitro (R = 0.84) and in vivo (R = 0.79), informing sufficient lesion creation (LI drop >20 Omega) and warning of excessive heating (LI drop >65 Omega). When creating an intercaval line, the total RF time was significantly reduced when using LI guidance (6.4 +/- 2 minutes in Delta 20 Omega and 8.1 +/- 1 minutes in Delta 30 Omega) compared with a standard 30-second workflow (18 +/- 7 minutes). Acute conduction block was achieved in all Delta 30 Omega and 30s lines. CONCLUSION The addition of LI to CF provides feedback on both electrical and mechanical loads. This provides information on tissue type and catheter-tissue coupling; provides feedback on whether volumetric tissue heating is inadequate, sufficient, or excessive; and reduces ablation time.
引用
收藏
页码:1371 / 1380
页数:10
相关论文
共 50 条
  • [21] Optimal local impedance drops for an effective radiofrequency ablation during cavo-tricuspid isthmus ablation
    Sasaki, Takehito
    Nakamura, Kohki
    Inoue, Mitsuho
    Minami, Kentaro
    Miki, Yuko
    Goto, Koji
    Take, Yutaka
    Kaseno, Kenichi
    Yamashita, Eiji
    Koyama, Keiko
    Naito, Shigeto
    JOURNAL OF ARRHYTHMIA, 2020, 36 (05) : 905 - 911
  • [22] When local impedance meets contact force: preliminary experience from the CHARISMA registry
    Francesco Solimene
    Valerio De Sanctis
    Ruggero Maggio
    Maurizio Malacrida
    Luca Segreti
    Matteo Anselmino
    Vincenzo Schillaci
    Massimo Mantica
    Marco Scaglione
    Antonio Dello Russo
    Filippo Maria Cauti
    Gianluca Zingarini
    Claudio Pandozi
    Marco Cavaiani
    Anna Ferraro
    Giampiero Maglia
    Giuseppe Stabile
    Journal of Interventional Cardiac Electrophysiology, 2022, 63 : 749 - 758
  • [23] When local impedance meets contact force: preliminary experience from the CHARISMA registry
    Solimene, Francesco
    De Sanctis, Valerio
    Maggio, Ruggero
    Malacrida, Maurizio
    Segreti, Luca
    Anselmino, Matteo
    Schillaci, Vincenzo
    Mantica, Massimo
    Scaglione, Marco
    Dello Russo, Antonio
    Cauti, Filippo Maria
    Zingarini, Gianluca
    Pandozi, Claudio
    Cavaiani, Marco
    Ferraro, Anna
    Maglia, Giampiero
    Stabile, Giuseppe
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2022, 63 (03) : 749 - 758
  • [24] Insight into contact force local impedance technology for predicting effective pulmonary vein isolation
    Lepillier, Antoine
    Maggio, Ruggero
    De Sanctis, Valerio
    Malacrida, Maurizio
    Stabile, Giuseppe
    Zakine, Cyril
    Champ-Rigot, Laure
    Anselmino, Matteo
    Segreti, Luca
    Dell'Era, Gabriele
    Garnier, Fabien
    Mascia, Giuseppe
    Pandozi, Claudio
    Dello Russo, Antonio
    Scaglione, Marco
    Cosaro, Giuseppe
    Ferraro, Anna
    Paziaud, Olivier
    Maglia, Giampiero
    Solimene, Francesco
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2023, 10
  • [25] Role of Catheter Contact Force on Biophysical Properties of the Ablation Lesion Formation in Radiofrequency Catheter Cardiac Ablation
    Masnok, Kriengsak
    Watanabe, Nobuo
    2021 IEEE REGION 10 SYMPOSIUM (TENSYMP), 2021,
  • [26] Contact-Force Recovery Can Predict Cardiac Perforation during Radiofrequency Ablation
    Nazeri, Alireza
    Ganapathy, Anand
    Massumi, Ali
    Massumi, Mehran
    Constantine, Garth
    Raz, Shaul
    Razavi, Mehdi
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2014, 37 (09): : 1129 - 1132
  • [27] The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study
    Reddy, Vivek Y.
    Shah, Dipen
    Kautzner, Josef
    Schmidt, Boris
    Saoudi, Nadir
    Herrera, Claudia
    Jais, Pierre
    Hindricks, Gerhard
    Peichl, Petr
    Yulzari, Aude
    Lambert, Hendrik
    Neuzil, Petr
    Natale, Andrea
    Kuck, Karl-Heinz
    HEART RHYTHM, 2012, 9 (11) : 1789 - 1795
  • [28] Cardiac radiofrequency ablation tracking using electrical impedance tomography
    Nguyen, Duc M.
    Qian, Pierre
    Barry, Tony
    McEwan, Alistair
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2020, 6 (01):
  • [29] Importance of Catheter Contact Force During Irrigated Radiofrequency Ablation: Evaluation in a Porcine Ex Vivo Model Using a Force-Sensing Catheter
    Thiagalingam, Aravinda
    D'Avila, Andre
    Foley, Lori
    Guerrero, J. Luis
    Lambert, Hendrik
    Leo, Giovanni
    Ruskin, Jeremy N.
    Reddy, Vivek Y.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2010, 21 (07) : 806 - 811
  • [30] Contact Sensing Provides a Highly Accurate Means to Titrate Radiofrequency Ablation Lesion Depth
    Holmes, Douglas
    Fish, Jeffrey M.
    Byrd, Israel A.
    Dando, Jeremy D.
    Fowler, Steven J.
    Cao, Hong
    Jensen, James A.
    Puryear, Harry A.
    Chinitz, Larry A.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2011, 22 (06) : 684 - 690