Combined local impedance and contact force for radiofrequency ablation assessment

被引:38
|
作者
Garrott, Kara [1 ]
Laughner, Jacob [1 ]
Gutbrod, Sarah [1 ]
Sugrue, Alan [2 ]
Shuros, Allan [1 ]
Sulkin, Matt [1 ]
Yasin, Omar [2 ]
Bush, Jamie [1 ]
Pottinger, Nathan [1 ]
Meyers, Jason [3 ]
Kapa, Suraj [2 ]
机构
[1] Boston Sci Corp, St Paul, MN USA
[2] Mayo Clin, Dept Cardiovasc Med, Coll Med, 200 First St SW, Rochester, MN 55905 USA
[3] Iowa Heart Ctr, Heart Rhythm Ctr, W Des Moines, IA USA
关键词
Ablation; Contact force; Electroanatomic mapping; Impedance; Radiofrequency; CATHETER ABLATION; LESION FORMATION; TEMPERATURE; SAFETY; MODEL; SIZE; FLOW;
D O I
10.1016/j.hrthm.2020.03.016
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND The combination of contact force (CF) and local impedance (LI) may improve tissue characterization and lesion prediction during radiofrequency (RF) ablation. OBJECTIVE The purpose of this study was to evaluate the utility of LI combined with CF in assessing RF ablation efficacy. METHODS An LI catheter with CF sensing was evaluated in swine (n = 11) and in vitro (n = 14). The relationship between LI and CF in different tissue types was evaluated in vivo. Discrete lesions were created in vitro and in vivo at a range of forces, powers, and durations. Finally, an intercaval line was created in 3 groups at 30 W: 30s, Delta 20 Omega, and Delta 30 Omega. In the Delta 20 Omega and Delta 30 Omega groups, the user ablated until a 20 or 30 Omega LI drop. In the 30s group, the user was blinded to LI. RESULTS In vivo, distinction in LI was found between the blood pool and the myocardium (blood pool: 122 +/- 7.02 Omega; perpendicular contact: 220 +/- 29 Omega; parallel contact: 207 +/- 31 Omega). LI drop correlated with lesion depth both in vitro (R = 0.84) and in vivo (R = 0.79), informing sufficient lesion creation (LI drop >20 Omega) and warning of excessive heating (LI drop >65 Omega). When creating an intercaval line, the total RF time was significantly reduced when using LI guidance (6.4 +/- 2 minutes in Delta 20 Omega and 8.1 +/- 1 minutes in Delta 30 Omega) compared with a standard 30-second workflow (18 +/- 7 minutes). Acute conduction block was achieved in all Delta 30 Omega and 30s lines. CONCLUSION The addition of LI to CF provides feedback on both electrical and mechanical loads. This provides information on tissue type and catheter-tissue coupling; provides feedback on whether volumetric tissue heating is inadequate, sufficient, or excessive; and reduces ablation time.
引用
收藏
页码:1371 / 1380
页数:10
相关论文
共 50 条
  • [11] Local impedance and contact force guidance to predict successful cavotricuspid isthmus ablation with a zero-fluoroscopy approach
    Melero-Polo, Jorge
    Cabrera-Ramos, Mercedes
    Alfonso-Almazan, Jose Manuel
    Marin-Garcia, Isabel
    Montilla-Padilla, Isabel
    Ruiz-Arroyo, Jose Ramon
    Lopez-Rodriguez, Guillermo
    Ramos-Maqueda, Javier
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2024, 10
  • [12] Contact force and force-time integral in atrial radiofrequency ablation predict transmurality of lesions
    Squara, Fabien
    Latcu, Decebal Gabriel
    Massaad, Youssef
    Mahjoub, Marouane
    Bun, Sok-Sithikun
    Saoudi, Nadir
    EUROPACE, 2014, 16 (05): : 660 - 667
  • [13] Real-time lesion assessment using a novel combined ultrasound and radiofrequency ablation catheter
    Wright, Matthew
    Harks, Erik
    Deladi, Szabolcs
    Suijver, Freek
    Barley, Maya
    van Dusschoten, Anneke
    Fokkenrood, Steven
    Zuo, Fei
    Sacher, Frederic
    Hocini, Meleze
    Haissaguerre, Michel
    Jais, Pierre
    HEART RHYTHM, 2011, 8 (02) : 304 - 312
  • [14] Impedance fluctuation and steam pop occurrence during radiofrequency current ablation: An experimental in vitro model
    Olszewski, Rafal
    Ptaszynski, Pawel
    Cygankiewicz, Iwona
    Kaczmarek, Krzysztof
    ADVANCES IN CLINICAL AND EXPERIMENTAL MEDICINE, 2021, 30 (10): : 1051 - 1056
  • [15] The effect of contact force during radiofrequency catheter ablation using a vibrating catheter: New cooling method for catheter ablation
    Yu, Kaihong
    Yamashita, Tetsui
    Shingyochi, Shigeaki
    Matsumoto, Kazuo
    Komiya, Atsuki
    Ohta, Makoto
    TECHNOLOGY AND HEALTH CARE, 2019, 27 (06) : 589 - 601
  • [16] Economic impact of atrial fibrillation ablation with radiofrequency contact force catheter versus cryoballoon catheter
    Pollak, Scott J.
    Goldstein, Laura
    Daskiran, Mehmet
    Kalsekar, Iftekhar
    Khanna, Rahul
    JOURNAL OF COMPARATIVE EFFECTIVENESS RESEARCH, 2019, 8 (04) : 251 - 264
  • [17] Contact force guided radiofrequency current application at developing myocardium
    Backhoff, David
    Mueller, Matthias J.
    Betz, Teresa
    Arnold, Andreas
    Schneider, Heike
    Paul, Thomas
    Krause, Ulrich
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (08) : 1757 - 1766
  • [18] Near-Field Ultrasound Imaging During Radiofrequency Catheter Ablation Tissue Thickness and Epicardial Wall Visualization and Assessment of Radiofrequency Ablation Lesion Formation and Depth
    Haines, David E.
    Wright, Matthew
    Harks, Erik
    Deladi, Szabolcs
    Fokkenrood, Steven
    Brink, Rob
    Belt, Harm
    Kolen, Alexander F.
    Mihajlovic, Nenad
    Zuo, Fei
    Rankin, Darrell
    Stoffregen, William
    Cockayne, Debra
    Cefalu, Joseph
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2017, 10 (12)
  • [19] Higher power achieves greater local impedance drop, shorter ablation time, and more transmural lesion formation in comparison to lower power in local impedance guided radiofrequency ablation of atrial fibrillation
    Yamashita, Shu
    Mizukami, Akira
    Ono, Maki
    Hiroki, Jiro
    Miyakuni, Shota
    Ueshima, Daisuke
    Matsumura, Akihiko
    Miyazaki, Shinsuke
    Sasano, Tetsuo
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2023, 34 (09) : 1869 - 1877
  • [20] Forcing Through Fat: Effects of Epicardial Adipose on Lesion Formation With Radiofrequency Ablation With Increasing Contact Force
    Zipse, Matthew
    Borne, Ryan
    Zheng Lijun
    Schuller, Joseph
    Edward, Justin
    Tzou, Wendy
    Sauer, William
    Nguyen, Duy
    CIRCULATION, 2018, 138