Combined local impedance and contact force for radiofrequency ablation assessment

被引:39
作者
Garrott, Kara [1 ]
Laughner, Jacob [1 ]
Gutbrod, Sarah [1 ]
Sugrue, Alan [2 ]
Shuros, Allan [1 ]
Sulkin, Matt [1 ]
Yasin, Omar [2 ]
Bush, Jamie [1 ]
Pottinger, Nathan [1 ]
Meyers, Jason [3 ]
Kapa, Suraj [2 ]
机构
[1] Boston Sci Corp, St Paul, MN USA
[2] Mayo Clin, Dept Cardiovasc Med, Coll Med, 200 First St SW, Rochester, MN 55905 USA
[3] Iowa Heart Ctr, Heart Rhythm Ctr, W Des Moines, IA USA
关键词
Ablation; Contact force; Electroanatomic mapping; Impedance; Radiofrequency; CATHETER ABLATION; LESION FORMATION; TEMPERATURE; SAFETY; MODEL; SIZE; FLOW;
D O I
10.1016/j.hrthm.2020.03.016
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND The combination of contact force (CF) and local impedance (LI) may improve tissue characterization and lesion prediction during radiofrequency (RF) ablation. OBJECTIVE The purpose of this study was to evaluate the utility of LI combined with CF in assessing RF ablation efficacy. METHODS An LI catheter with CF sensing was evaluated in swine (n = 11) and in vitro (n = 14). The relationship between LI and CF in different tissue types was evaluated in vivo. Discrete lesions were created in vitro and in vivo at a range of forces, powers, and durations. Finally, an intercaval line was created in 3 groups at 30 W: 30s, Delta 20 Omega, and Delta 30 Omega. In the Delta 20 Omega and Delta 30 Omega groups, the user ablated until a 20 or 30 Omega LI drop. In the 30s group, the user was blinded to LI. RESULTS In vivo, distinction in LI was found between the blood pool and the myocardium (blood pool: 122 +/- 7.02 Omega; perpendicular contact: 220 +/- 29 Omega; parallel contact: 207 +/- 31 Omega). LI drop correlated with lesion depth both in vitro (R = 0.84) and in vivo (R = 0.79), informing sufficient lesion creation (LI drop >20 Omega) and warning of excessive heating (LI drop >65 Omega). When creating an intercaval line, the total RF time was significantly reduced when using LI guidance (6.4 +/- 2 minutes in Delta 20 Omega and 8.1 +/- 1 minutes in Delta 30 Omega) compared with a standard 30-second workflow (18 +/- 7 minutes). Acute conduction block was achieved in all Delta 30 Omega and 30s lines. CONCLUSION The addition of LI to CF provides feedback on both electrical and mechanical loads. This provides information on tissue type and catheter-tissue coupling; provides feedback on whether volumetric tissue heating is inadequate, sufficient, or excessive; and reduces ablation time.
引用
收藏
页码:1371 / 1380
页数:10
相关论文
共 21 条
[1]   The effects of electrode-tissue contact on radiofrequency lesion generation [J].
Avitall, B ;
Mughal, K ;
Hare, J ;
Helms, R ;
Krum, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (12) :2899-2910
[2]   Flow effect on lesion formation in RF cardiac catheter ablation [J].
Cao, H ;
Vorperian, VR ;
Tungjitkusolmun, S ;
Tsai, JZ ;
Haemmerich, D ;
Bin Choy, Y ;
Webster, JG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2001, 48 (04) :425-433
[3]   Determinants of Acute and Late Pulmonary Vein Reconnection in Contact Force-Guided Pulmonary Vein Isolation Identifying the Weakest Link in the Ablation Chain [J].
El Haddad, Milad ;
Taghji, Philippe ;
Phlips, Thomas ;
Wolf, Michael ;
Demolder, Anthony ;
Choudhury, Rajin ;
Knecht, Sebastien ;
Vandekerckhove, Yves ;
Tavernier, Rene ;
Nakagawa, Hiroshi ;
Duytschaever, Mattias .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2017, 10 (04)
[4]   A novel assessment of local impedance during catheter ablation: initial experience in humans comparing local and generator measurements [J].
Gunawardene, Melanie ;
Muenkler, Paula ;
Eickholt, Christian ;
Akbulak, Ruken Oe ;
Jularic, Mario ;
Klatt, Niklas ;
Hartmann, Jens ;
Dinshaw, Leon ;
Jungen, Christiane ;
Moser, Julia M. ;
Merbold, Lydia ;
Willems, Stephan ;
Meyer, Christian .
EUROPACE, 2019, 21 :34-42
[5]   TISSUE HEATING DURING RADIOFREQUENCY CATHETER ABLATION - A THERMODYNAMIC MODEL AND OBSERVATIONS IN ISOLATED PERFUSED AND SUPERFUSED CANINE RIGHT VENTRICULAR FREE WALL [J].
HAINES, DE ;
WATSON, DD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (06) :962-976
[6]   A three-dimensional finite element model of radiofrequency ablation with blood flow and its experimental validation [J].
Jain, MK ;
Wolf, PD .
ANNALS OF BIOMEDICAL ENGINEERING, 2000, 28 (09) :1075-1084
[7]   First clinical use of novel ablation catheter incorporating local impedance data [J].
Martin, Claire A. ;
Martin, Ruairidh ;
Gajendragadkar, Parag R. ;
Maury, Philippe ;
Takigawa, Masateru ;
Cheniti, Ghassen ;
Frontera, Antonio ;
Kitamura, Takeshi ;
Duchateau, Josselin ;
Vlachos, Konstantinos ;
Bourier, Felix ;
Lam, Anna ;
Lord, Stephen ;
Murray, Stephen ;
Shephard, Ewen ;
Pambrun, Thomas ;
Denis, Arnaud ;
Derval, Nicolas ;
Hocini, Meleze ;
Haissaguerre, Michel ;
Jais, Pierre ;
Sacher, Frederic .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2018, 29 (09) :1197-1206
[8]   Assessment of Catheter Tip Contact Force Resulting in Cardiac Perforation in Swine Atria Using Force Sensing Technology [J].
Perna, Francesco ;
Heist, E. Kevin ;
Danik, Stephan B. ;
Barrett, Conor D. ;
Ruskin, Jeremy N. ;
Mansour, Moussa .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2011, 4 (02) :218-224
[9]   Improving procedural and one-year outcome after contact force-guided pulmonary vein isolation: the role of interlesion distance, ablation index, and contact force variability in the 'CLOSE'-protocol [J].
Phlips, Thomas ;
Taghji, Philippe ;
El Haddad, Milad ;
Wolf, Michael ;
Knecht, Sebastien ;
Vandekerckhove, Yves ;
Tavernier, Rene ;
Duytschaever, Mattias .
EUROPACE, 2018, 20 :F419-F427
[10]   Feasibility, Efficacy, and Safety of Radiofrequency Ablation of Atrial Fibrillation Guided by Monitoring of the Initial Impedance Decrease as a Surrogate of Catheter Contact [J].
Reichlin, Tobias ;
Lane, Christopher ;
Nagashima, Koichi ;
Nof, Eyal ;
Chopra, Nagesh ;
Ng, Justin ;
Barbhaiya, Chirag ;
Tadros, Tomas ;
John, Roy M. ;
Stevenson, William G. ;
Michaud, Gregory F. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2015, 26 (04) :390-396