Externalized conductors and electrical dysfunction in transvenous ventricular leads: Results of the Cardiac Lead Assessment Study

被引:0
|
作者
Freedman, Roger A. [1 ]
Curtis, Anne B. [2 ]
Delgado, Stephanie M. [3 ]
Lee, Li-Yin [3 ,4 ]
机构
[1] Univ Utah, Dept Med, Salt Lake City, UT USA
[2] Univ Buffalo, Jacobs Sch Med & Biomed Sci, Buffalo, NY USA
[3] Cardiac Rhythm Management, Sylmar, CA USA
[4] Appl Mol Transport, 1 Tower Pl,Suite 850, San Francisco, CA 94080 USA
来源
HEART RHYTHM O2 | 2022年 / 3卷 / 02期
关键词
Electrical dysfunction; Externalized conductor; Highvoltage lead; Implantable cardioverter-defibrillator; Lead insulation; RIATA LEAD; DEFIBRILLATOR LEADS; SILICONE LEADS; FAILURE; PERFORMANCE; PREVALENCE; INSULATION;
D O I
10.1016/j.hroo.2022.01.007
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND The Cardiac Lead Assessment Study (CLAS) was a large prospective, multicenter, international postmarket surveillance study conducted at 45 sites. OBJECTIVE The purpose of CLAS was to examine the prevalence and incidence of externalized conductors and electrical dysfunction in subjects with selected St. Jude Medical defibrillator and left ventricular leads. METHODS Cinefluoroscopy was used to determine the presence of externalized conductors at enrollment and at 12-, 24-, and 36month follow-up visits. Lead electrical measurements were collected systematically. RESULTS The study enrolled 2216 subjects with a total of 2847 study leads. The prevalence of externalized conductors through 36 months for Riata leads was 30.9%, Riata ST leads 12.6%, Durata leads 0.5%, and QuickSite/QuickFlex leads 4.7%. The prevalence of electrical dysfunction through 36 months for Riata was 4.0%, Riata ST 3.3%, Durata 2.4%, and QuickSite/QuickFlex 0.3%. In Riata and Riata ST leads with externalized conductors, there was a low risk of electrical dysfunction. None of the Durata or QuickSite/QuickFlex leads with externalized conductors developed electrical dysfunction. There was no evidence of an electrical short in a highvoltage shocking circuit leading to failed shock. CONCLUSION A high prevalence of externalized conductors was found in Riata and Riata ST defibrillator leads, with a higher risk of externalization for 8F Riata leads than for 7F Riata ST leads. The 98% reduction in prevalence of externalized conductors in Durata leads compared to Riata/Riata ST leads confirms that the design improvements culminating in Durata leads significantly improved abrasion resistance and durability.
引用
收藏
页码:160 / 168
页数:9
相关论文
共 50 条
  • [1] Fluoroscopic and Electrical Assessment of a Series of Defibrillation Leads: Prevalence of Externalized Conductors
    Kodoth, Vivek N.
    Hodkinson, Emily C.
    Noad, Rebecca L.
    Ashfield, Kyle P.
    Cromie, Nicholas A.
    McEneaney, David J.
    Wilson, Carol M.
    Roberts, Michael J. D.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (12): : 1498 - 1504
  • [2] Prevalence of externalized conductors in Riata and Riata ST silicone leads: Results from the prospective, multicenter Riata Lead Evaluation Study
    Hayes, David
    Freedman, Roger
    Curtis, Anne B.
    Niebauer, Mark
    Kay, G. Neal
    Dinerman, Jay
    Beau, Scott
    HEART RHYTHM, 2013, 10 (12) : 1778 - 1782
  • [3] Results of transvenous lead extraction of coronary sinus leads in patients with cardiac resynchronization therapy
    Starck, Christoph T.
    Caliskan, Etem
    Klein, Holger
    Steffel, Jan
    Schoenrath, Felix
    Falk, Volkmar
    CHINESE MEDICAL JOURNAL, 2013, 126 (24) : 4703 - 4706
  • [4] Generator Pocket Adhesions of Cardiac Leads: Classification and Correlation with Transvenous Lead Extraction Results
    Biefer, Hector Rodriguez Cetina
    Huerlimann, David
    Gruenenfelder, Juerg
    Salzberg, Sacha P.
    Steffel, Jan
    Falk, Volkmar
    Starck, Christoph T.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2013, 36 (09): : 1111 - 1116
  • [5] Fluoroscopic and Electrical Assessment of Implantable Cardioverter Defibrillator Leads: A Prospective Observational Study
    McKeag, Nicholas A.
    Hodkinson, Emily C.
    Noad, Rebecca L.
    Kodoth, Vivek N.
    Ashfield, Kyle
    Wilson, Carol M.
    McEneaney, David J.
    Roberts, Michael J. D.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2014, 37 (11): : 1538 - 1543
  • [6] Contribution of ethylenetetrafluoroethylene (ETFE) insulation to the electrical performance of RiataA® silicone leads having externalized conductors
    Fischer, Avi
    Klehn, Russell
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2013, 37 (02) : 141 - 145
  • [7] Prevalence and Presentation of Externalized Conductors and Electrical Abnormalities in Riata Defibrillator Leads After Fluoroscopic Screening Report From the Netherlands Heart Rhythm Association Device Advisory Committee
    Theuns, Dominic A. M. J.
    Elvan, Arif
    de Voogt, Willem
    de Cock, Carel C.
    van Erven, Lieselot
    Meine, Mathias
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2012, 5 (06) : 1059 - 1063
  • [8] Prevalence and management of electrical lead abnormalities in cardiac implantable electronic device leads
    Roberts, Hilary
    Matheson, Kara
    Sapp, John
    Gardner, Martin
    Gray, Chris
    AbdelWahab, Amir
    Lee, David
    MacIntyre, Ciorsti
    Parkash, Ratika
    HEART RHYTHM O2, 2023, 4 (07): : 417 - 426
  • [9] Assessment of the impact of organisational model of transvenous lead extraction on the effectiveness and safety of procedure: an observational study
    Tulecki, Lukasz
    Jachec, Wojciech
    Polewczyk, Anna
    Czajkowski, Marek
    Targonska, Sylwia
    Tomkow, Konrad
    Karpeta, Kamil
    Nowosielecka, Dorota
    Kutarski, Andrzej
    BMJ OPEN, 2022, 12 (12):
  • [10] Post myocardial infarction, left ventricular dysfunction, and the expanding role of cardiac implantable electrical devices
    Naccarelli, GV
    CLINICAL CARDIOLOGY, 2005, 28 (11) : I51 - I57