Possibility of gated silicon drift detector detecting hard X-ray

被引:0
作者
Matsuura, Hideharu [1 ]
Fukushima, Shinya [1 ]
Sakurai, Shungo [1 ]
Ishikawa, Shohei [1 ]
Takeshita, Akinobu [1 ]
Hidaka, Atsuki [1 ]
机构
[1] Osaka Electrocommun Univ, Neyagawa, Osaka 5728530, Japan
来源
HARD X-RAY, GAMMA-RAY, AND NEUTRON DETECTOR PHYSICS XVII | 2015年 / 9593卷
关键词
Gated silicon drift detector; Silicon drift detector; Hard X-ray; Thick silicon X-ray detector; HIGH-RESOLUTION; ROOM-TEMPERATURE; SPECTROSCOPY; PERFORMANCE; DESIGN; ARRAY;
D O I
10.1117/12.2186062
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
One of the authors has proposed a simple-structure silicon X-ray detector (gated silicon drift detector: GSDD), whose structure is much simpler than commercial silicon drift detectors (SDDs). SDDs contain multiple built-in metal-oxide-semiconductor field-effect transistors (MOSFETs) or implanted resistors, whose fabrication processes lower the yield rate of detectors, and also require at least two high-voltage sources. On the other hand, GSDDs do not contain built-in MOSFETs or implanted resistors. Moreover, GSDDs require only one high-voltage source. Therefore, GSDDs greatly reduce the cost of the X-ray detection system. We fabricated prototype GSDDs that contained 0.625-mm-thick Si substrates with an active area of 18 mm2, operated by Peltier cooling and a single voltage source. Its energy resolution at 5.9 keV from an 55Fe source was 145 eV at -38 degrees C and -90 V. Thicker Si substrates are required to enhance its absorption of X-rays. To detect X-ray photons with energies up to 77 keV for X-ray absorbance higher than 15%, we simulate the electric potential distribution in GSDDs with Si thicknesses from 0.625 to 3.0 mm. We obtain an adequate electric potential distribution in the thicknesses of up to 3.0 mm, and the capacitance of the GSDD remains small and its X-ray count rate remain high. The high reverse bias required in the 3-mm-thick GSDD was a third of that in a 3-mm-thick pin diode.
引用
收藏
页数:11
相关论文
共 54 条
[1]  
Anderson B.L., 2005, Fundamentals of semiconductor devices
[2]   ROOM-TEMPERATURE X-RAY SPECTROSCOPY WITH A SILICON DIODE DETECTOR AND AN ULTRA-LOW NOISE PREAMPLIFIER [J].
BERTUCCIO, G ;
PULLIA, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :1704-1709
[3]   Laboratory and test beam results from a large-area silicon drift detector [J].
Bonvicini, V ;
Busso, L ;
Giubellino, P ;
Gregorio, A ;
Idzik, M ;
Kolojvari, A ;
Montano, LM ;
Nouais, D ;
Petta, C ;
Rashevsky, A ;
Randazzo, N ;
Reito, S ;
Tosello, F ;
Vacchi, A ;
Vinogradov, L ;
Zampa, N .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 439 (2-3) :476-482
[4]   Performance of a Thin-window Silicon Drift Detector X-Ray Fluorescence Spectrometer [J].
Carini, Gabriella A. ;
Chen, Wei ;
De Geronimo, Gianluigi ;
Gaskin, Jessica A. ;
Keister, Jeffrey W. ;
Li, Zheng ;
Ramsey, Brian D. ;
Rehak, Pavel ;
Siddons, David P. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) :2843-2849
[5]   A new silicon drift detector with reduced lateral diffusion [J].
Castoldi, A ;
Rehak, P ;
Holl, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 377 (2-3) :375-380
[6]   LARGE AREA CYLINDRICAL SILICON DRIFT DETECTOR [J].
CHEN, W ;
KRANER, H ;
LI, Z ;
REHAK, P ;
GATTI, E ;
LONGONI, A ;
SAMPIETRO, M ;
HOLL, P ;
KEMMER, J ;
FASCHINGBAUER, U ;
SCHMITT, B ;
WORNER, A ;
WURM, JP .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (04) :619-628
[7]  
Christophersen M., 2008, P IEEE NUCL SCI S WA, P2727
[8]   Silicon drift detectors with enlarged sensitive areas [J].
Eggert, T ;
Boslau, O ;
Goldstrass, P ;
Kemmer, J .
X-RAY SPECTROMETRY, 2004, 33 (04) :246-252
[9]   A monolithic array of silicon drift detectors for high-resolution gamma-ray imaging [J].
Fiorini, C. ;
Longoni, A. ;
Perotti, F. ;
Labanti, C. ;
Rossi, E. ;
Lechner, P. ;
Soltau, H. ;
Strüder, Lothar .
IEEE Transactions on Nuclear Science, 2002, 49 II (03) :995-1000
[10]   A new detection system for x-ray microanalysis based on a silicon drift detector with Peltier cooling [J].
Fiorini, C ;
Kemmer, J ;
Lechner, P ;
Kromer, K ;
Rohde, M ;
Schulein, T .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (06) :2461-2465