Development of four different devices to turn over introducer sheaths

被引:0
作者
Borggrefe, Jan [1 ]
Knabe, Helmuth [1 ]
Hoffmann, Beata [1 ]
Heller, Martin [1 ]
Schaefer, Jost Philipp [1 ]
机构
[1] Univ Klinikum Schleswig Holstein, Radiol Diagnost Klin, Kiel, Germany
关键词
FEMORAL-ARTERY CATHETERIZATION; TECHNICAL NOTE; RETROGRADE; CONVERSION; INTERVENTION;
D O I
10.1016/j.jvs.2012.10.065
中图分类号
R61 [外科手术学];
学科分类号
摘要
Introduction: We developed four prototype sheath-turning auxiliary devices (STADs) and evaluated them in an in vitro study setup designed to enable the change of catheter direction in endovascular interventions. Methods: Four different prototypes, A through D, of STADs were designed and created by modifying commercially available dilators and catheters. All STADs work with different anchor-like tips to ensure fixation inside the vessel at the puncture site. The STAD is loaded into the introducer sheath, retracted with the introducer sheath, and turned at the puncture site. The STADs were tested in an in vitro vascular study setup using bovine veins. Success rates and procedure times were calculated, and the handling, reliability, and overall performance were evaluated. The maximum soft tissue thickness (STTmax) applicable was tested using bovine vessels with 7-mm thickness surrounded by a soft tissue phantom consisting of chicken breast. A retrospective cross-sectional observation in 108 patients from our center was performed to provide mean STTmax at the common femoral artery in patients for comparison. Results: The success rate ranged between 75% for prototype D and 90% for prototypes A and C. The procedure time averaged 60 seconds (range, 25-165 seconds). The mean handling was rated 2.4 (good) for prototype A, 2.0 (good) for prototype B, 2.6 (satisfactory) for prototype C, and 3.5 (poor) for prototype D. Mean reliability was rated 3.4 (satisfactory) for prototype A, 2.0 (good) for prototype B, 1.6 (good) for prototype C, and 2.4 (good) for prototype D. Mean overall performance was rated 2.0 (good) for prototype C, 2.6 (satisfactory) for prototype B, 3.3 (poor) for prototype D, and 3.4 (poor) for prototype A. In the cross-sectional patient observation, the mean STTmax was 3.3 cm (range, 0,5-13 cm) with a 95% confidence interval of the distribution including an STTmax of up to 8 cm. The STTmax was <= 5 cm in 100 of 108 patients (93%). The applicable STTmax for prototype A was 1 cm (8 of 10 successful cases), 3 cm for prototype B (9 of 10 successful cases), 5 cm for prototype C (8 of 10 successful cases), and 3 cm for prototype D (7 of 10 successful cases). Conclusions: All four STAD prototypes offered the ability of turning the sheaths at the puncture site in an in vitro vascular study setup. In the future, this concept may allow routine clinical performance of turning maneuvers at the groin vascular access site.
引用
收藏
页码:194 / 200
页数:7
相关论文
共 12 条
[1]   A NEW CATHETER CONFIGURATION FOR SELECTIVE ANTEGRADE CATHETERIZATION OF THE SUPERFICIAL FEMORAL-ARTERY - TECHNICAL NOTE [J].
BOHNDORF, K ;
GUNTHER, RW .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1991, 14 (02) :129-131
[2]  
Criado FJ, 1998, J ENDOVASC SURG, V5, P236, DOI 10.1583/1074-6218(1998)005<0236:OEIIIS>2.0.CO
[3]  
2
[4]  
Dormandy J, 1999, Semin Vasc Surg, V12, P96
[6]   An improved reversal technique from retrograde to antegrade femoral artery cannulation [J].
Hartnell, G .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1998, 21 (06) :512-513
[7]   TECHNIQUE FOR CONVERSION OF A RETROGRADE INTO AN ANTEGRADE FEMORAL-ARTERY CATHETERIZATION [J].
KADIR, S ;
BAASSIRI, A ;
BARTH, KH .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1981, 136 (02) :430-431
[8]  
Meyerson Shari L, 2002, Vasc Endovascular Surg, V36, P137, DOI 10.1177/153857440203600208
[9]   Conversion of a Retrograde Access to an Antegrade Access for Bilateral Femorocrural Intervention in One Session [J].
Neumann, C. ;
Dorffner, R. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2012, 184 (05) :450-454
[10]   CATHETER FOR CONVERSION OF RETROGRADE TO ANTEGRADE FEMORAL-ARTERY CATHETERIZATION [J].
PATEL, YD .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1990, 154 (01) :179-180