Human-Derived D-Dimer for External Quality Assessment Results of Four Surveys in Ontario

被引:6
作者
Crowther, Mark A. [1 ]
Bates, Shannon M. [2 ,3 ]
Keeney, Michael [4 ]
Raby, Anne [5 ]
Flynn, Greg [5 ]
机构
[1] McMaster Univ, St Josephs Hosp, Hamilton Reg Lab Med Program, Hamilton, ON L8N 4A6, Canada
[2] McMaster Univ, Hamilton, ON, Canada
[3] Henderson Res Ctr, Hamilton, ON, Canada
[4] London Hlth Sci Ctr, London, ON, Canada
[5] Qual Management Program Lab Serv, Toronto, ON, Canada
关键词
Human-derived D-dimer; External quality assessment; Validated quality control material;
D O I
10.1309/AJCPUBFRQ8Q6ZWTY
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
External quality assessment (EQA) of D-dimer assays has been limited by a lack of standardized human-derived testing material. In 2006 and 2007, the Quality Management Program-Laboratory Services, Toronto, Canada, investigated the use of commercially prepared lyophilized human plasma spiked with human-derived D-dimer components manufactured by Affinity Biologicals, Hamilton, Canada. Four surveys were performed. Participants reported the level or presence of D-dimer using quantitative or qualitative methods. Participants performing quantitative testing provided their unit of measure and reference interval. Results were considered correct if they fell within the range appropriate for each sample (normal/negative or abnormal/positive). Overall, survey results were excellent, with 4.0% (95% confidence interval [CI], 1.3%-9.1%), 0.8% (CI, 0.0%-1.5%), 2.3% (CI, 0.5%-6.6%), and 2.3% (CI, 0.4%-6.6%) of participants reporting an incorrect result in the first, second, third, and fourth surveys, respectively. A commercially prepared D-dimer is a suitable material for EQA testing.
引用
收藏
页码:805 / 810
页数:6
相关论文
共 6 条
  • [1] de Maat MPM, 2000, SEMIN THROMB HEMOST, V26, P625
  • [2] JENNINGS I, 2005, J THROMB HAEMOST S1, V3, pP1898
  • [3] Emerging technologies and quality assurance: The United Kingdom National External Quality Assessment Scheme perspective
    Jennings, Ian
    Kitchen, Dianne P.
    Woods, Timothy A. L.
    Kitchen, Steven
    Walker, Isobel D.
    [J]. SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2007, 33 (03) : 243 - 249
  • [4] An evaluation of D-dimer in the diagnosis of pulmonary embolism - A randomized trial
    Kearon, Clive
    Ginsberg, Jeffrey S.
    Douketis, James
    Turpie, Alexander G.
    Bates, Shannon M.
    Lee, Agnes Y.
    Crowther, Mark A.
    Weitz, Jeffrey I.
    Brill-Edwards, Patrick
    Wells, Philip
    Anderson, David R.
    Kovacs, Michael J.
    Linkins, Lori-Ann
    Julian, Jim A.
    Bonilla, Laura R.
    Gent, Michael
    [J]. ANNALS OF INTERNAL MEDICINE, 2006, 144 (11) : 812 - 821
  • [5] RABY A, 2005, J THROMB HAEMOST S1, V3, pP443
  • [6] Performance of quantitative D-dimer methods: results of the Italian external quality assessment scheme
    Tripodi, A.
    Chantarangkul, V.
    [J]. JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2007, 5 (01) : 184 - 185