Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects

被引:33
作者
Halkiadakis, Ioannis [1 ]
Stratos, Aimilianos [1 ]
Stergiopoulos, George [1 ]
Patsea, Eleni [1 ]
Skouriotis, Sotiris [1 ]
Mitropoulos, Panagiotis [1 ]
Papaconstantinou, Dimitrios [1 ,2 ]
Georgopoulos, Gerasimos [1 ,2 ]
机构
[1] Athens Eye Hosp, Athens 15124, Greece
[2] Univ Athens, Sch Med, Dept Ophthalmol, GR-11527 Athens, Greece
关键词
Goldmann applanation tonometry; Self-tonometry; Ocular hypertension; Rebound tonometry; Central corneal thickness; GOLDMANN APPLANATION TONOMETER; CORNEAL THICKNESS; GLAUCOMA MANAGEMENT; CLINICAL-EVALUATION; DYNAMIC CONTOUR; AGREEMENT;
D O I
10.1007/s00417-011-1875-6
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
To compare Icare ONE rebound self-tonometer (ICRBT) measurements with Goldman applanation tonometry (GAT). A trained examiner instructed each of 60 normal subjects on use of the ICRBT. Each subject then took two measurements of his/her own pressure using the ICRBT. Finally, a different examiner, who was masked to the earlier readings, measured IOP by GAT. Bland-Altman limits of agreement (LOA), intraclass correlation coefficients (ICCs), Kappa values, and paired t-test were used to assess the agreement between the two methods. Pearson's correlation coefficient was used for correlation analysis. All of the subjects were able to obtain correct measurements with ICRBT after three attempts. The mean intraocular pressure with ICRBT and GAT measurements were 16.0 +/- 3.3 mmHg and 13.7 +/- 2.5 mmHg respectively. The mean difference between patient's ICRBT and technician's GAT measurements was 2.3 mmHg (p < 0.001). In 63 % (38/60) of the cases the IOP difference (ICRBT - GAT) was within +/- 3 mmHg. The weighted Kappa for the IOP measurements of the two methods was 0.49 (95% CI: 0.30-0.68, p < 0.001), indicating acceptable agreement. A significantly positive correlation was found between ICRBT IOP measurements and central corneal thickness (CCT) (r = 0.48, p < 0.001). In addition, the difference in IOP measurements (ICRBT - GAT) between the two methods was positively correlated with CCT (r = 0.31, p = 0.015), indicating that greater thickness is associated with greater differences between the two methods. The ICRBT was reliable in the hands of normal subjects, and may be used for self-monitoring of IOP. ICRBT measurements generally overestimated GAT measurements.
引用
收藏
页码:1207 / 1211
页数:5
相关论文
共 31 条
[1]   Comparison of the ICare® rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists [J].
Abraham, L. M. ;
Epasinghe, N. C. R. ;
Selva, D. ;
Casson, R. .
EYE, 2008, 22 (04) :503-506
[2]   MEASUREMENT IN MEDICINE - THE ANALYSIS OF METHOD COMPARISON STUDIES [J].
ALTMAN, DG ;
BLAND, JM .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES D-THE STATISTICIAN, 1983, 32 (03) :307-317
[3]   Evaluation of the ICare Rebound Tonometer as a Home Intraocular Pressure Monitoring Device [J].
Asrani, Sanjay ;
Chatterjee, Ashmita ;
Wallace, David K. ;
Santiago-Turla, Cecilia ;
Stinnett, Sandra .
JOURNAL OF GLAUCOMA, 2011, 20 (02) :74-79
[4]   Intraocular pressure measurements throughout the 24 h [J].
Bagga, Harmohina ;
Liu, John H. K. ;
Weinreb, Robert N. .
CURRENT OPINION IN OPHTHALMOLOGY, 2009, 20 (02) :79-83
[5]  
Balkrishnan Rajesh, 2003, Am J Geriatr Pharmacother, V1, P75, DOI 10.1016/S1543-5946(03)90003-1
[6]   Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients [J].
Brusini, P ;
Salvetat, ML ;
Zeppieri, M ;
Tosoni, C ;
Parisi, L .
JOURNAL OF GLAUCOMA, 2006, 15 (03) :213-217
[7]   Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure [J].
Cervino, A. .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2006, 90 (12) :1444-1446
[8]   The influence of cornea properties on rebound tonometry [J].
Chui, Wan-sang ;
Lam, Andrew ;
Chen, Davie ;
Chiu, Roger .
OPHTHALMOLOGY, 2008, 115 (01) :80-84
[9]   Method for the noninvasive measurement of intraocular pressure in mice [J].
Danias, J ;
Kontiola, AI ;
Filippopoulos, T ;
Mittag, T .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (03) :1138-1141
[10]   Clinical evaluation of rebound tonometer [J].
Davies, LN ;
Bartlett, H ;
Mallen, EAH ;
Wolffsohn, JS .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2006, 84 (02) :206-209