Incomplete neointimal coverage of sirolimus-eluting stents - Angioscopic findings

被引:341
作者
Kotani, Jun-ichi [1 ]
Awata, Masaki
Nanto, Shinsuke
Uematsu, Masaaki
Oshima, Fusako
Minamiguchi, Hitoshi
Mintz, Gary S.
Nagata, Seiki
机构
[1] Kansai Rosai Hosp, Div Cardiovasc, Amagasaki, Hyogo, Japan
[2] Cardiovasc Res Fdn, New York, NY USA
关键词
D O I
10.1016/j.jacc.2005.11.092
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES The goal of this study was to use angioscopy to investigate the amount of neointimal coverage after sirolimus-eluting stent (SES) implantation. BACKGROUND Sirolimus-eluting stents reduce intimal hyperplasia. METHODS We used angioscopy to evaluate 37 consecutive stented coronary artery lesions (15 SES and 22 bare-metal stents [BMS]) in 25 patients (18 men, 7 women) at 3 to 6 months after stent implantation. Angioscopic evaluation focused on: 1) neointimal coverage of stent struts, and 2) the existence of thrombi. The degree of neointimal coverage was classified as grade 0 when there was no neointimal coverage (similar to immediately after the implantation); grade 1 when stent struts bulged into the lumen, but were covered and still translucently visible; grade 2 when stent struts were visible but not clearly seen (not translucent); and grade 3 when stent struts were not visible because they were embedded in the neointima. RESULTS Thrombi were identified in eight stented segments, tended to be more common with SES (p = 0.14), but were not seen on angiography. Three of the 15 SES (20%) had grade 0 neointimal coverage, and only 2 SES (13.3%) had complete coverage (grades 2/3). In contrast, all 22 BMS showed complete intimal coverage (grades 2/3). Thrombi were more common in stents with incomplete neointimal coverage (p = 0.09). CONCLUSIONS The SES had incomplete neointimal coverage three to six months after implantation, and this was associated with subclinical thrombus formation. (J Am Coll Cardiol 2006;47:2108-11) (c) 2006 by the American College of Cardiology Foundation.
引用
收藏
页码:2108 / 2111
页数:4
相关论文
共 17 条
[1]   Remodeling of in-stent neointima, which became thinner and transparent over 3 years - Serial angiographic and angioscopic follow-up [J].
Asakura, M ;
Ueda, Y ;
Nanto, S ;
Hirayama, A ;
Adachi, T ;
Kitakaze, M ;
Hori, M ;
Kodama, K .
CIRCULATION, 1998, 97 (20) :2003-2006
[2]   Time course of stent endothelialization after intravascular radiation therapy in rabbit iliac arteries [J].
Cheneau, E ;
John, MC ;
Fournadjiev, J ;
Chan, RC ;
Kim, HS ;
Leborgne, L ;
Pakala, R ;
Yazdi, H ;
Ajani, AE ;
Virmani, R ;
Waksman, R .
CIRCULATION, 2003, 107 (16) :2153-2158
[3]   Pathological mechanisms of fatal late coronary stent thrombosis in humans [J].
Farb, A ;
Burke, AP ;
Kolodgie, FD ;
Virmani, R .
CIRCULATION, 2003, 108 (14) :1701-1706
[4]   Five-year clinical follow-up after intracoronary radiation - Results of a randomized clinical trial [J].
Grise, MA ;
Massullo, V ;
Jani, S ;
Popma, JJ ;
Russo, RJ ;
Schatz, RA ;
Guarneri, EM ;
Steuterman, S ;
Cloutier, DA ;
Leon, MB ;
Tripuraneni, P ;
Teirstein, PS .
CIRCULATION, 2002, 105 (23) :2737-2740
[5]   Sirolimus-eluting stent implanted in human coronary artery for 16 months - Pathological findings [J].
Guagliumi, G ;
Farb, A ;
Musumeci, G ;
Valsecchi, O ;
Tespili, M ;
Motta, T ;
Virmani, R .
CIRCULATION, 2003, 107 (09) :1340-1341
[6]   EXPERIMENTAL VALIDATION OF GEOMETRIC AND DENSITOMETRIC CORONARY MEASUREMENTS ON THE NEW-GENERATION CARDIOVASCULAR ANGIOGRAPHY ANALYSIS SYSTEM (CAAS-II) [J].
HAASE, J ;
ESCANED, J ;
VANSWIJNDREGT, EM ;
OZAKI, Y ;
GRONENSCHILD, E ;
SLAGER, CJ ;
SERRUYS, PW .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1993, 30 (02) :104-114
[7]   Incidence, predictors, and outcome of thrombosis after successful implantation of drug-eluting stents [J].
Iakovou, I ;
Schmidt, T ;
Bonizzoni, E ;
Ge, L ;
Sangiorgi, GM ;
Stankovic, G ;
Airoldi, F ;
Chieffo, A ;
Montorfano, M ;
Carlino, M ;
Michev, I ;
Corvaja, N ;
Briguori, C ;
Gerckens, U ;
Grube, E ;
Colombo, A .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2005, 293 (17) :2126-2130
[8]   Stent thrombosis after successful sirolimus-eluting stent implantation [J].
Jeremias, A ;
Sylvia, B ;
Bridges, J ;
Kirtane, AJ ;
Bigelow, B ;
Pinto, DS ;
Ho, KKL ;
Cohen, DJ ;
Garcia, LA ;
Cutlip, DE ;
Carrozza, JP .
CIRCULATION, 2004, 109 (16) :1930-1932
[9]   Late thrombosis in drug-eluting coronary stents after discontinuation of antiplatelet therapy [J].
McFadden, EP ;
Stabile, E ;
Regar, E ;
Cheneau, E ;
Ong, ATL ;
Kinnaird, T ;
Suddath, WO ;
Weissman, NJ ;
Torguson, R ;
Kent, KM ;
Pichard, AD ;
Satler, LF ;
Waksman, R ;
Serruys, PW .
LANCET, 2004, 364 (9444) :1519-1521
[10]   Intravascular ultrasound assessment of neointima distribution and the length of stent that was free of intravascular ultrasound-detectable intimal hyperplasia in paclitaxel-eluting stents [J].
Mintz, GS ;
Hong, MK ;
Raizner, AE ;
Lee, CW ;
Kim, JJ ;
Escolar, E ;
Fearnot, NE ;
Park, SW ;
Park, SJ ;
Weissman, NJ .
AMERICAN JOURNAL OF CARDIOLOGY, 2005, 95 (01) :107-109