A comprehensive theory of thermoluminescence (TL) supralinearity and sensitisation, the Unified Interaction Model (UNIM) is described, based on both radiation absorption stage and recombination stage mechanisms. The UNIM can be applied to both heavy charged particles and isotropically ionising gamma rays and electrons, in a unified and self-consistent conceptual and mathematical formalism. The model is applied to explain the unique features of gamma induced supralinearity and sensitisation of peaks 4 and 5 in LiF:Mg,Ti, especially the strictly linear, then supralinear behaviour and the dependence of the supralinearity on ionisation density (gamma ray energy and particle type). A model is proposed for peak 4, based on its mainly 'hole-trapping' characteristics, which explains its diametrically opposed behaviour in sensitised material, i.e. a decreased sensitivity and supralinear dose response. All features arise from a localised trapping entity (the track for HCPs), spatially correlated trapping centres and luminescent centres (TC/LC pairs) for gamma rays and electrons, which dominate the dose response at low dose and are nor subject to intra-track competitive processes, thus leading to linear dose-response behaviour. The decrease of the supralinearity with decreasing gamma ray energy (increasing ionisation density) arises from the increasing probability of the TC/LC,pair to capture simultaneously an electron-hole pair, leading to geminate recombination not subject to competitive processes. The UNIM is shown to be capable of yielding excellent fits to the experimental data with many of the variable parameters of the model strongly constrained by ancillary measurements.
引用
收藏
页码:29 / 34
页数:6
相关论文
共 12 条
[1]  
BRAUNLICH P, 1979, TOPICS APPL PHYSICS, V37
[2]   NEW EVIDENCE AGAINST THE PRESENCE OF PEAK-IV IN THE TL AND PTTL GLOW CURVES OF HEAVILY ANNEALED TLD-100 [J].
DELGADO, A ;
ROS, JMG ;
PORTILLO, JC .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1988, 21 (06) :1015-1018
[3]   STUDY OF THE LONG-TERM STABILITY OF PEAKS 4 AND 5 IN TLD-100 - CORRELATION WITH ISOTHERMAL DECAY MEASUREMENTS AT ELEVATED-TEMPERATURES [J].
HOROWITZ, YS ;
BENSHACHAR, B ;
YOSSIAN, D .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1993, 26 (09) :1475-1481
[4]   The unified interaction model applied to the gamma-induced supralinearity and sensitisation of peaks 4 and 5 in LiF:Mg,Ti (TLD-100) [J].
Horowitz, YS ;
Mahajna, S ;
Oster, L ;
Weizman, Y ;
Satinger, D ;
Yossian, D .
RADIATION PROTECTION DOSIMETRY, 1998, 78 (03) :169-193
[5]   The dependence of the dose response supralinearity of peak 5 in TLD-100 on recombination temperature [J].
Horowitz, YS ;
Satinger, D ;
Oster, L .
RADIATION PROTECTION DOSIMETRY, 1999, 84 (1-4) :99-102
[6]   The track interaction model for alpha particle induced thermoluminescence supralinearity: Dependence of the supralinearity on the vector properties of the alpha particle radiation field [J].
Horowitz, YS ;
Rosenkrantz, M ;
Mahajna, S ;
Yossian, D .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1996, 29 (01) :205-217
[7]  
HOROWITZ YS, 1984, THERMOLUMINESCENCE T, V2, P3
[8]   HEAVY CHARGED-PARTICLE THERMO-LUMINESCENCE DOSIMETRY - TRACK STRUCTURE-THEORY AND EXPERIMENTS [J].
KALEFEZRA, J ;
HOROWITZ, YS .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1085-1100
[9]   The unified interaction model applied to the gamma ray induced supralinearity and sensitization of peak 5 in LiF:Mg,Ti (TLD-100) [J].
Mahajna, S ;
Horowitz, YS .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 1997, 30 (18) :2603-2619
[10]   THERMOLUMINESCENCE AND COLOR CENTER CORRELATIONS IN DOSIMETRY LIF [J].
MAYHUGH, MR ;
CHRISTY, RW ;
JOHNSON, NM .
JOURNAL OF APPLIED PHYSICS, 1970, 41 (07) :2968-&