Cone-beam Computed Tomography-guided Stereotactic Liver Punctures: A Phantom Study

被引:10
作者
Toporek, Grzegorz [1 ]
Wallach, Daphne [1 ]
Weber, Stefan [1 ]
Bale, Reto [2 ]
Widmann, Gerlig [2 ]
机构
[1] Univ Bern, ARTORG Ctr Biomed Engn Res, CH-3008 Bern, Switzerland
[2] Med Univ Innsbruck, Sect Microinvas Therapy, Dept Radiol, A-6020 Innsbruck, Austria
关键词
Cone-beam computed tomography; Radiological interventions; Ablations; PERCUTANEOUS RADIOFREQUENCY ABLATION; HEPATOCELLULAR-CARCINOMA; NAVIGATION SYSTEM; IMAGE REGISTRATION; PLANNING CT; SOFT-TISSUE; WORKFLOW; DESIGN;
D O I
10.1007/s00270-013-0635-x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Images from computed tomography (CT), combined with navigation systems, improve the outcomes of local thermal therapies that are dependent on accurate probe placement. Although the usage of CT is desired, its availability for time-consuming radiological interventions is limited. Alternatively, three-dimensional images from C-arm cone-beam CT (CBCT) can be used. The goal of this study was to evaluate the accuracy of navigated CBCT-guided needle punctures, controlled with CT scans. Five series of five navigated punctures were performed on a nonrigid phantom using a liver specific navigation system and CBCT volumetric dataset for planning and navigation. To mimic targets, five titanium screws were fixed to the phantom. Target positioning accuracy (TPECBCT) was computed from control CT scans and divided into lateral and longitudinal components. Additionally, CBCT-CT guidance accuracy was deducted by performing CBCT-to-CT image coregistration and measuring TPECBCT-CT from fused datasets. Image coregistration was evaluated using fiducial registration error (FRECBCT-CT) and target registration error (TRECBCT-CT). Positioning accuracies in lateral directions pertaining to CBCT (TPECBCT = 2.1 +/- A 1.0 mm) were found to be better to those achieved from previous study using CT (TPECT = 2.3 +/- A 1.3 mm). Image coregistration error was 0.3 +/- A 0.1 mm, resulting in an average TRE of 2.1 +/- A 0.7 mm (N = 5 targets) and average Euclidean TPECBCT-CT of 3.1 +/- A 1.3 mm. Stereotactic needle punctures might be planned and performed on volumetric CBCT images and controlled with multidetector CT with positioning accuracy higher or similar to those performed using CT scanners.
引用
收藏
页码:1629 / 1637
页数:9
相关论文
共 32 条
[11]   Percutaneous radiofrequency ablation of liver metastases in potential candidates for resection the - "Test-of-time" approach [J].
Livraghi, T ;
Solbiati, L ;
Meloni, F ;
Ierace, T ;
Goldberg, SN ;
Gazelle, GS .
CANCER, 2003, 97 (12) :3027-3035
[12]  
Maier-Hein L, 2007, LECT NOTES COMPUT SC, V4792, P42
[13]   In vivo accuracy assessment of a needle-based navigation system for CT-guided radiofrequency ablation of the liver [J].
Maier-Hein, Lena ;
Tekbas, Aysun ;
Seitel, Alexander ;
Pianka, Frank ;
Mueller, Sascha A. ;
Satzl, Stefanie ;
Schawo, Simone ;
Radeleff, Boris ;
Tetzlaff, Ralf ;
Franz, Alfred M. ;
Mueller-Stich, Beat P. ;
Wolf, I. ;
Kauczor, H. -U. ;
Schmied, B. M. ;
Meinzer, H. -P. .
MEDICAL PHYSICS, 2008, 35 (12) :5385-5396
[14]  
McCarley James R, 2010, Semin Intervent Radiol, V27, P255, DOI 10.1055/s-0030-1261783
[15]   Computer-assisted navigation system for interventional CT-guided procedures: Results of phantom and clinical studies [J].
Meier-Meitinger, M. ;
Nagel, M. ;
Kalender, W. ;
Bautz, W. A. ;
Baum, U. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2008, 180 (04) :310-317
[16]   Percutaneous radiofrequency ablation guided by contrast-enhanced harmonic sonography with artificial pleural effusion for hepatocellular carcinoma in the hepatic dome [J].
Minami, Y ;
Kudo, M ;
Kawasaki, T ;
Chung, H ;
Ogawa, C ;
Shiozaki, H .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 182 (05) :1224-1226
[17]   Cone-Beam Computed Tomography-Guided Percutaneous Radiologic Gastrostomy [J].
Moehlenbruch, Markus ;
Nelles, Michael ;
Thomas, Daniel ;
Willinek, Winfried ;
Gerstner, Andreas ;
Schild, Hans H. ;
Wilhelm, Kai .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2010, 33 (02) :315-320
[18]   C-Arm Cone Beam CT for Hepatic Tumor Ablation Under Real-Time 3D Imaging [J].
Morimoto, Manabu ;
Numata, Kazushi ;
Kondo, Masaaki ;
Nozaki, Akito ;
Hamaguchi, Shingo ;
Takebayashi, Shigeo ;
Tanaka, Katsuaki .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (05) :W452-W454
[19]  
Oliveira-Santos Thiago, 2011, Information Processing in Computer-Assisted Interventions. Proceedings of the Second International Conference, IPCAI 2011, P156, DOI 10.1007/978-3-642-21504-9_15
[20]   A navigation system for percutaneous needle interventions based on PET/CT images: Design, workflow and error analysis of soft tissue and bone punctures [J].
Oliveira-Santos, Thiago ;
Klaeser, Bernd ;
Weitzel, Thilo ;
Krause, Thomas ;
Nolte, Lutz-Peter ;
Peterhans, Matthias ;
Weber, Stefan .
COMPUTER AIDED SURGERY, 2011, 16 (05) :203-219