PurposeFlow instability has been shown to contribute to the risk of future cardiovascular and cerebrovascular events. Nonetheless, it is challenging to noninvasively detect and identify flow instability in blood vessels. Here, we present a new framework called Doppler ultrasound bandwidth imaging (DUBI) that uses high-frame-rate ultrasound and Doppler bandwidth analysis principles to assess flow instability within an image view. MethodsDoppler ultrasound bandwidth imaging seeks to estimate the instantaneous Doppler bandwidth based on autoregressive modeling at every pixel position of data frames acquired from high-frame-rate plane wave pulsing. This new framework is founded upon the principle that flow instability naturally gives rise to a wide range of flow velocities over a sample volume, and such velocity range in turn yields a larger Doppler bandwidth estimate. The ability for DUBI to map unstable flow was first tested over a range of fluid flow conditions (ranging from laminar to turbulent) with a nozzle-flow phantom. As a further demonstration, DUBI was applied to assess flow instability in healthy and stenosed carotid bifurcation phantoms. ResultsNozzle-flow phantom results showed that DUBI can effectively detect and visualize the difference in Doppler bandwidth magnitude (increased from 2.1 to 5.2kHz) at stable and unstable flow regions in an image view. Receiver operating characteristic analysis also showed that DUBI can achieve optimal sensitivity and specificity of 0.72 and 0.83, respectively. In the carotid phantom experiments, differences were observed in the spatiotemporal dynamics of Doppler bandwidth over a cardiac cycle. Specifically, as the degree of stenosis increased (from 50% to 75%), DUBI showed an increase in Doppler bandwidth magnitude from 1.4kHz in the healthy bifurcation to 7.7kHz at the jet tail located downstream from a 75% stenosis site, thereby indicating flow perturbation in the stenosed bifurcations. ConclusionDUBI can detect unstable flow. This new technique can provide useful hemodynamic information that may aid clinical diagnosis of atherosclerosis.
引用
收藏
页码:1620 / 1633
页数:14
相关论文
共 50 条
  • [41] An Efficient Algorithm Based on Spectrum Migration for High Frame Rate Ultrasound Imaging
    Feng, Shuai
    Jin, Liu
    Wang, Yadan
    Zhao, Wei
    Peng, Hu
    Qiao, Heyuan
    CMES-COMPUTER MODELING IN ENGINEERING & SCIENCES, 2021, 126 (02): : 739 - 754
  • [42] High Frame Rate Contrast Enhanced Ultrasound Imaging of Lymphatic Vessel Phantom
    Zhu, Jiaqi
    Lin, Shengtao
    Harput, Sevan
    Toulemonde, Matthieu
    Leow, Chee Hau
    Tang, Meng-Xing
    2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [43] ApodNet: Learning for High Frame Rate Synthetic Transmit Aperture Ultrasound Imaging
    Chen, Yinran
    Liu, Jing
    Luo, Xiongbiao
    Luo, Jianwen
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2021, 40 (11) : 3190 - 3204
  • [44] Transcranial ultrasound perfusion imaging:: Implementation of a low MI and a high frame rate
    Harrer, J. U.
    Hornen, S.
    Valaikiene, J.
    Oertel, M. F.
    Kloetzsch, C.
    Schlachetzki, F.
    ULTRASCHALL IN DER MEDIZIN, 2007, 28 (04): : 380 - 386
  • [45] High Frame-Rate Ultrasound Imaging Using Deep Learning Beamforming
    Ghani, Muhammad Usman
    Meral, F. Can
    Vignon, Francois
    Robert, Jean-luc
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 295 - 298
  • [46] High-Frame-Rate Coherence Imaging of the Heart with Ultrasound Diverging Waves
    Matrone, Giulia
    Spairani, Edoardo
    Matrone, Benedetta
    Ramalli, Alessandro
    PROCEEDINGS OF THE 2020 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2020,
  • [47] High-frame rate vector flow imaging of the carotid bifurcation
    Alfredo Goddi
    Chandra Bortolotto
    Ilaria Fiorina
    Maria Vittoria Raciti
    Marianna Fanizza
    Elena Turpini
    Giulia Boffelli
    Fabrizio Calliada
    Insights into Imaging, 2017, 8 : 319 - 328
  • [48] A new technique for improved spatial resolution in high frame rate color Doppler imaging
    Kirkhorn, J
    Bjærum, S
    Olstad, B
    Kristoffersen, K
    Torp, H
    2003 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2, 2003, : 1947 - 1950
  • [49] High-frame rate vector flow imaging of the carotid bifurcation
    Goddi, Alfredo
    Bortolotto, Chandra
    Fiorina, Ilaria
    Raciti, Maria Vittoria
    Fanizza, Marianna
    Turpini, Elena
    Boffelli, Giulia
    Calliada, Fabrizio
    INSIGHTS INTO IMAGING, 2017, 8 (03): : 319 - 328
  • [50] Elasticity Imaging using conventional and high-frame rate ultrasound imaging: Experimental study
    Park, Suhyun
    Aglyamov, Salavat R.
    Emelianov, Stanislav Y.
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2007, 54 (11) : 2246 - 2256