Inter-laboratory comparison of different rapid methods for the detection of bacterial contamination in platelet concentrates

被引:16
|
作者
Vollmer, T. [1 ]
Hinse, D. [1 ]
Schottstedt, V. [2 ]
Bux, J. [2 ]
Tapernon, K. [3 ]
Sibrowski, W. [3 ]
Knabbe, C. [1 ]
Dreier, J. [1 ]
机构
[1] Ruhr Univ Bochum, Univ Klin, Inst Lab & Transfus Med, Herz & Diabet Zentrum N Rhein Westfalen, D-32545 Bad Oeynhausen, Germany
[2] DRK Blutspendedienst W, Zentrallabor Hagen, Germany
[3] Univ Klinikum Munster, Inst Transfus Med & Transplantat Immunol, Munster, Germany
关键词
BactiFlow; nucleic acid amplification techniques; Pan Genera detection; platelet bacteria screening; rapid methods; 6 YEARS EXPERIENCE; TECHNOLOGY; PCR;
D O I
10.1111/j.1423-0410.2011.01572.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Bacterial contamination of platelet concentrates still represents a major risk in transfusion medicine, and a variety of screening methods have been available to improve the safety of PCs. In the present study, the analytical quality of three different rapid screening methods (BactiFlow flow cytometry, Pan Genera Detection Assay, 23S rRNA RT-PCR) was evaluated in an inter-laboratory comparison in three different German blood services. Methods Samples were inoculated with different bacteria [Staphylococcus aureus, Staphylococcus epidermidis, Escherichia coli (two strains), Klebsiella pneumoniae (two strains), Enterobacter aerogenes (one strain), Serratia marcescens (one strain)] at different counts (4.5 X 1034.5 X 108 CFU/ml) alternating with negative samples in one transfusion facility. Samples were blinded with a random order for each screening method, shipped to partners and analysed immediately after receipt with different rapid screening methods. Results The inter-laboratory comparison revealed that the BactiFlow assay and 23S rRNA RT-PCR-screening detected all samples correctly (positive: 12/12, negative: 8/8). The Pan Genera Detection Assay test detected only four of the positive samples. Four of the non-detected positive samples were below the assays detection limit. Another four inoculated samples with comparatively high bacteria counts were detected false negative (E. coli (two strains): 9.87 X 105 and 2.10 X 107 CFU/ml, respectively, K. pneumoniae: 4.79 X 106 CFU/ml, S. aureus: 6.03 X 105 CFU/ml). All rapid screening methods revealed no false-positive results. Conclusions Both BactiFlow and 23S rRNA RT-PCR demonstrated a high sensitivity to detecting bacterial contamination in PCs. The Pan Genera Detection Assay had some shortcomings regarding sensitivity, especially for the detection of Gram-negative strains.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 47 条
  • [1] Three different bacterial detection systems for platelet concentrates under inter-laboratory conditions
    Bian GuoHui
    Yang Chunhui
    He Miao
    Wang Hong
    Liu Jiaxin
    Cao Ye
    Yang Hong
    Liu Zhong
    Li Wuping
    TRANSFUSION AND APHERESIS SCIENCE, 2013, 49 (03) : 600 - 607
  • [2] Bench Test for the Detection of Bacterial Contamination in Platelet Concentrates Using Rapid and Cultural Detection Methods with a Standardized Proficiency Panel
    Vollmer, Tanja
    Knabbe, Cornelius
    Geilenkeuser, Wolf-Jochen
    Schmidt, Michael
    Dreier, Jens
    TRANSFUSION MEDICINE AND HEMOTHERAPY, 2015, 42 (04) : 220 - 225
  • [3] Establishment of a proficiency panel for an external quality assessment programme for the detection of bacterial contamination in platelet concentrates using rapid and cultural detection methods
    Vollmer, T.
    Schmidt, M.
    Hourfar, K.
    Schottstedt, V.
    Pichl, L.
    Gubbe, K.
    Knabbe, C.
    Dreier, J.
    VOX SANGUINIS, 2016, 110 (04) : 336 - 343
  • [4] The Pan Genera Detection Immunoassay: a Novel Point-of-Issue Method for Detection of Bacterial Contamination in Platelet Concentrates
    Vollmer, Tanja
    Hinse, Dennis
    Kleesiek, Knut
    Dreier, Jens
    JOURNAL OF CLINICAL MICROBIOLOGY, 2010, 48 (10) : 3475 - 3481
  • [5] Development and application of MLVA methods as a tool for inter-laboratory surveillance
    Nadon, C. A.
    Trees, E.
    Ng, L. K.
    Nielsen, E. Moller
    Reimer, A.
    Maxwell, N.
    Kubota, K. A.
    Gerner-Smidt, P.
    EUROSURVEILLANCE, 2013, 18 (35) : 10 - 19
  • [6] Detection of bacterial contamination in platelet concentrates by a sensitive flow cytometric assay (BactiFlow): a multicentre validation study
    Vollmer, T.
    Dreier, J.
    Schottstedt, V.
    Bux, J.
    Tapernon, K.
    Sibrowski, W.
    Kleesiek, K.
    Knabbe, C.
    TRANSFUSION MEDICINE, 2012, 22 (04) : 262 - 271
  • [7] Comparison of different methods of bacterial detection in blood components
    Schmidt, M.
    ISBT SCIENCE SERIES, VOL 4, NO 1, STATE OF THE ART PRESENTATIONS, 2009, 4 (01): : 80 - 86
  • [8] Applications of real-time PCR in the screening of platelet concentrates for bacterial contamination
    Mohammadi, Tamimount
    Savelkoul, Paul H. M.
    Pietersz, Ruby N. I.
    Reesink, Henk W.
    EXPERT REVIEW OF MOLECULAR DIAGNOSTICS, 2006, 6 (06) : 865 - 872
  • [9] Inter-laboratory validation of PCR-based HPV detection in pathology specimens
    Petersen, Iver
    Schewe, Christiane
    Schluens, Karsten
    Dietel, Manfred
    Speich, Norbert
    Schmitt, Christoph
    Bollmann, Magdolna
    Sotlar, Karl
    Bueltmann, Burkhard
    Dours-Zimmermann, Maria T.
    Padberg, Barbara
    Zimmermann, Dieter R.
    VIRCHOWS ARCHIV, 2007, 451 (03) : 701 - 716
  • [10] The Italian inter-laboratory study on the detection of Pseudomonas syringae pv. actinidiae
    Loreti, Stefania
    Pucci, Nicoletta
    Gallelli, Angela
    Minardi, Paola
    Ardizzi, Stefano
    Balestra, Giorgio Mariano
    Mazzaglia, Angelo
    Taratufolo, Maria Claudia
    PHYTOPATHOLOGIA MEDITERRANEA, 2014, 53 (01) : 159 - 167