Computer-aided Alignment of off-axis three-mirror imaging spectrometer system

被引:5
|
作者
Gong Dun [1 ]
Wang Hong [1 ]
Tian Tie-yin [1 ]
机构
[1] Chinese Acad Sci, Key Lab Opt Syst Adv Mfg Technol, Changchun Inst Opt Fine Mech & Phys, Changchun 130033, Peoples R China
来源
INTERNATIONAL SYMPOSIUM ON PHOTOELECTRONIC DETECTION AND IMAGING 2013: IMAGING SPECTROMETER TECHNOLOGIES AND APPLICATIONS | 2013年 / 8910卷
关键词
Off-axis three mirror imaging spectrometer system; Computer-aided alignment; Sensitivity matrix; Merit Function;
D O I
10.1117/12.2033021
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The space camera is one of the most advanced optical equipments in obtaining earth surface information from space. With the development of the optical design, manufacture and alignment technology, the performance of the optical camera is moving forward quickly. In space-to-ground remote sensing field, the resolution of ground remote sensing images have become higher and higher. Off-axis three-mirror system becomes an advanced optical system structure of space camera at present, because of its merits of large field, long focal length, no obstruction, high modulated transfer function, wide spectrum, good image quality, small cubage and light weight, etc. As it is difficult to align and test, the alignment of off-axis three mirror system is time-consuming, so it is important and necessary of studying the computer-aided alignment of the complex optical system. Aiming at cutting the time of alignment, Computer-aided alignment technology is applied to this system. According to relation between fringe Zernike coefficient and Seidel aberration, wavefront aberration in the exit pupil of optical system is instead with fringe Zernike coefficient. A series of Zernike coefficient expressing the incorrect parameters of system are obtained by using multi-fields ZYGO interferometer auto-collimating interference test, which are transformed into geometric aberrations as the corrected object. Incorrect parameters of the system are determined by sensitivity matrix resulted from optical design software. These incorrect parameters are defined to be variables; the Zernike coefficient are defined to be the optimized target in merit function. The system is optimized by applying optical design software to receive the incorrect parameters result. Consequently, the adjusted result is brought into the optical design software to verify the right selection of incorrect parameters. The practical experiments are also given, result with RMS value lower than 0.04 lambda is acquired by using this alignment technology which has proved the effectiveness of above methodology of guiding-alignment.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] The computer-aided alignment study of three-mirror off-axis field bias optical system
    Pang Zhi-hai
    Fan Xue-wu
    Ma Zhen
    Chen Qin-fang
    6TH INTERNATIONAL SYMPOSIUM ON ADVANCED OPTICAL MANUFACTURING AND TESTING TECHNOLOGIES: OPTICAL TEST AND MEASUREMENT TECHNOLOGY AND EQUIPMENT, 2012, 8417
  • [2] Study on computer-aided alignment method of a three-mirror off-axis aspherical optical system
    Zhao Xi-ting
    Jiao Wen-chun
    Liao Zhi-bo
    Wang Yun
    Chen Jia-yi
    5TH INTERNATIONAL SYMPOSIUM ON ADVANCED OPTICAL MANUFACTURING AND TESTING TECHNOLOGIES: OPTICAL TEST AND MEASUREMENT TECHNOLOGY AND EQUIPMENT, 2010, 7656
  • [3] Optical design of off-axis three-mirror anastigmatic system for imaging spectrometer
    Wang, Meiqin
    Wang, Zhonghou
    Bai, Jiaguang
    Wang, M. (wangmeiqin.1984@163.com), 1600, Chinese Society of Astronautics (41): : 167 - 172
  • [4] Computer-aided Alignment of Off-axis Two-mirror Reflective System
    Bin, Wang
    Fan, Wu
    Tang, Ye Yu
    6TH INTERNATIONAL SYMPOSIUM ON ADVANCED OPTICAL MANUFACTURING AND TESTING TECHNOLOGIES: OPTICAL TEST AND MEASUREMENT TECHNOLOGY AND EQUIPMENT, 2012, 8417
  • [5] Design and Alignment of Folded Off-Axis Three-Mirror Optical System
    Chen Taixi
    Wu Yanxiong
    Song Shaoman
    Li Shuxian
    LASER & OPTOELECTRONICS PROGRESS, 2021, 58 (17)
  • [6] Application of computer-aided alignment to off-axis cassegrain system
    Luo, Miao
    Zhu, Yong-Tian
    Guangxue Jishu/Optical Technique, 2008, 34 (04): : 514 - 517
  • [7] Design of telecentric off-axis three-mirror system of imaging spectrometer with wide field-of-view
    Liu X.
    Xiang Y.
    Guangxue Xuebao/Acta Optica Sinica, 2011, 31 (06):
  • [8] Research on Computer-aided alignment for an afocal off-axis optical system
    Zhao Xi-ting
    Ji Yi
    Yue Peng-yuan
    Dong Xin
    Zhang Chao
    Liu Hui
    Huang Yang
    Zhang Zhi-fei
    Jiao Wen-chun
    9TH INTERNATIONAL SYMPOSIUM ON ADVANCED OPTICAL MANUFACTURING AND TESTING TECHNOLOGIES: LARGE MIRRORS AND TELESCOPES, 2018, 10837
  • [9] Design of wide field-of-view off-axis three-mirror telescope for imaging spectrometer
    Chen, W. (xue_ciomp@163.com), 1600, Chinese Optical Society (42):
  • [10] Design of Off-axis Three-mirror System with Zernike Surface
    Gong Guangbiao
    Wei Jun
    Shen Weimin
    6TH INTERNATIONAL SYMPOSIUM ON PRECISION ENGINEERING MEASUREMENTS AND INSTRUMENTATION, 2010, 7544