Non-Invasive Optical Sensor Based Approaches for Monitoring Virus Culture to Minimize BSL3 Laboratory Entry

被引:1
作者
Ragupathy, Viswanath [1 ]
Setty, Mohan Kumar Hayuri Giri [1 ]
Kostov, Yordan [2 ,3 ]
Ge, Xudong [2 ,3 ]
Uplekar, Shaunak [2 ,3 ]
Hewlett, Indira [1 ]
Rao, Govind [2 ,3 ]
机构
[1] US FDA, LMV, DETTD, OBRR,CBER, Silver Spring, MD 20993 USA
[2] Univ Maryland Baltimore Cty, Ctr Adv Sensor Technol, Baltimore, MD 21250 USA
[3] Univ Maryland Baltimore Cty, Dept Chem Biochem & Environm Engn, Baltimore, MD 21250 USA
关键词
virus cell culture; BSL3; use; monitoring; optical sensor; MAMMALIAN-CELL CULTURE; DISSOLVED-OXYGEN; HIV; SYSTEM; PANEL; PH;
D O I
10.3390/s150714864
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
High titers of infectious viruses for vaccine and diagnostic reference panel development are made by infecting susceptible mammalian cells. Laboratory procedures are strictly performed in a Bio-Safety Level-3 (BSL3) laboratory and each entry and exit involves the use of disposable Personnel Protective Equipment (PPE) to observe cell culture conditions. Routine PPE use involves significant recurring costs. Alternative non-invasive optical sensor based approaches to remotely monitor cell culture may provide a promising and cost effective approach to monitor infectious virus cultures resulting in lower disruption and costs. We report here the monitoring of high titer cultures of Human Immunodeficiency Virus-1 (HIV-1) and Herpes Simplex Virus-2 (HSV-2) remotely with the use of optical oxygen sensors aseptically placed inside the cell culture vessel. The replacement of culture media for cell and virus propagation and virus load monitoring was effectively performed using this fluorescent sensor and resulted in half the number of visits to the BSL3 lab (five versus ten).
引用
收藏
页码:14864 / 14870
页数:7
相关论文
共 15 条
[1]   A Unique Noninvasive Approach to Monitoring Dissolved O2 and CO2 in Cell Culture [J].
Chatterjee, Madhubanti ;
Ge, Xudong ;
Uplekar, Shaunak ;
Kostov, Yordan ;
Croucher, Leah ;
Pilli, Manohar ;
Rao, Govind .
BIOTECHNOLOGY AND BIOENGINEERING, 2015, 112 (01) :104-110
[2]  
Chermann J C, 1983, Antibiot Chemother (1971), V32, P48
[3]   Validation of an optical sensor-based high-throughput bioreactor system for mammalian cell culture [J].
Ge, XD ;
Hanson, M ;
Shen, H ;
Kostov, Y ;
Brorson, KA ;
Frey, DD ;
Moreira, AR ;
Rao, G .
JOURNAL OF BIOTECHNOLOGY, 2006, 122 (03) :293-306
[4]   A Completely Noninvasive Method of Dissolved Oxygen Monitoring in Disposable Small-Scale Cell Culture Vessels Based on Diffusion Through Permeable Vessel Walls [J].
Gupta, Priyanka A. ;
Ge, Xudong ;
Kostov, Yordan ;
Rao, Govind .
BIOTECHNOLOGY PROGRESS, 2014, 30 (01) :172-177
[5]   Bioprocess monitoring [J].
Harms, P ;
Kostov, Y ;
Rao, G .
CURRENT OPINION IN BIOTECHNOLOGY, 2002, 13 (02) :124-127
[6]   RAPID AND SENSITIVE VIRAL CULTURE METHOD FOR HUMAN IMMUNODEFICIENCY VIRUS TYPE-1 [J].
JACKSON, JB ;
COOMBS, RW ;
SANNERUD, K ;
RHAME, FS ;
BALFOUR, HH .
JOURNAL OF CLINICAL MICROBIOLOGY, 1988, 26 (07) :1416-1418
[7]   ONLINE MONITORING OF HYBRIDOMA CELL-GROWTH USING A LASER TURBIDITY SENSOR [J].
KONSTANTINOV, KB ;
PAMBAYUN, R ;
MATANGUIHAN, R ;
YOSHIDA, T ;
PERUSICH, CM ;
HU, WS .
BIOTECHNOLOGY AND BIOENGINEERING, 1992, 40 (11) :1337-1342
[8]   Alterations of intracellular pH homeostasis in apoptosis: origins and roles [J].
Lagadic-Gossmann, D ;
Huc, L ;
Lecureur, V .
CELL DEATH AND DIFFERENTIATION, 2004, 11 (09) :953-961
[9]  
Manak M, 2012, AIDS RES HUM RETROV, V28, P594, DOI [10.1089/AID.2011.0271, 10.1089/aid.2011.0271]
[10]   Estimating sexual transmission of HIV from persons aware and unaware that they are infected with the virus in the USA [J].
Marks, Gary ;
Crepaz, Nicole ;
Janssen, Robert S. .
AIDS, 2006, 20 (10) :1447-1450