Significant reduction of radiation exposure to operator and staff during cardiac interventions by analysis of radiation leakage and improved lead shielding

被引:101
作者
Kuon, E
Schmitt, M
Dahm, JB
机构
[1] Klin Frank Schweiz, Dept Cardiol, D-91320 Ebermannstadt, Germany
[2] Univ Erlangen Nurnberg, Ctr Diagnost Radiol, Erlangen, Germany
[3] Ernst Moritz Arndt Univ Greifswald, Dept Internal Med, Greifswald, Germany
关键词
D O I
10.1016/S0002-9149(01)02161-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The objectives of this study were to disclose and to reduce occupational radiation leakage in invasive cardiology. prospectively, we analyzed various dose parameters for 330 coronary procedures. We used a Rando phantom to measure scatter entrance skin air kerma to the operator (S-ESAK-O) during fluoroscopy for all standard tube angulations, and to plot isodose lines for 0degrees/0degrees-posterior anterior angulation. The patient's measured dose area product due to diagnostic catheterization and elective percutaneous transluminal coronary angioplasty was 6.2 and 10.4 Gycm(2), which represents 11% and 13% of currently typical values, respectively. With use of 0.5- and 1.0-mm overcouch and undercouch shielding, it was possible to reduce the mean of 4,686 nSv/Gycm(2) to 677 and 277 nSv/Gycm(2), respectively. Closure of radiation leakage up to 897 muSv/hour at the operator's gonadal height (80 to 105 cm), not heretofore described, was achieved by an additional 1.0-mm, lead-equivalent undercouch-top and overcouch-flop adjacent to the table, down to a S-ESAKO/dose area product level of 47.5 nSv/Gycm(2). With use of a 0.5-mm lead apron, collar, glasses, foot-switch shield and 1.0-mm lead cover around the patient's thighs, the operator received a mean S-ESAK-O of 8.5, while his forehead, eyes, thyroid, chest, gonads, and hands were exposed to 68.2, 1.2, 1.2, 1.2, 0.8, and 58.2 nSv/Gycm(2), respectively. In conclusion, radiation-attenuating intervention techniques and improved lead protection can effectively contribute to a new state of the art in invasive cardiology, with reduction of operator radiation exposure to 0.8% of typical S-ESAK-O levels in advanced catheterization laboratories. (C) 2002 by Excerpta Medica, Inc.
引用
收藏
页码:44 / 49
页数:6
相关论文
共 30 条
[1]   EFFECTIVENESS OF GLASS LENSES IN REDUCING EXPOSURE TO EYES [J].
AGARWAL, SK ;
FRIESEN, EJ ;
HUDDLESTON, AL ;
RAO, RP .
RADIOLOGY, 1978, 129 (03) :810-811
[2]  
Aldridge HE, 1997, CAN J CARDIOL, V13, P459
[3]   RADIATION EXPOSURE TO PERSONNEL DURING CARDIAC CATHETERIZATION [J].
ARDRAN, GM ;
FURSDON, PS .
RADIOLOGY, 1973, 106 (03) :517-518
[4]   RADIATION EXPOSURE TO OPERATOR PERFORMING CARDIAC ANGIOGRAPHY WITH U-ARM SYSTEMS [J].
BALTER, S ;
SONES, FM ;
BRANCATO, R .
CIRCULATION, 1978, 58 (05) :925-932
[5]  
Becker C, 1999, ROFO-FORTSCHR RONTG, V170, P99
[6]   RADIATION EXPOSURE TO ANGIOGRAPHER DURING CORONARY ARTERIOGRAPHY USING U ARM IMAGE AMPLIFIER [J].
BEGG, FR ;
HANS, LF .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1975, 1 (03) :261-265
[7]   Patient radiation doses during cardiac catheterization procedures [J].
Betsou, S ;
Efstathopoulos, EP ;
Katritsis, D ;
Faulkner, K ;
Panayiotakis, G .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (846) :634-639
[8]   REDUCTION IN RADIATION EXPOSURE DURING CORONARY ANGIOGRAPHY [J].
CARPENTER, ML ;
SINGER, PR ;
JAIN, A ;
DEHMER, GJ .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1990, 21 (03) :195-197
[9]  
*CEC COUNC DIR, 1997, EURATOM AMTSBLATT L, V180, P22
[10]   Factors affecting patient radiation exposure during routine coronary angiography in a tertiary referral centre [J].
Clark, AL ;
Brennan, AG ;
Robertson, LJ ;
McArthur, JD .
BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (866) :184-189