Traceable Measurement of Optical Path Length of Gas Cell Based on Tunable Diode Laser Absorption Spectroscopy

被引:2
作者
Long Jiang-xiong [1 ,2 ]
Zhang Yu-jun [1 ]
Shao Li [1 ]
Ye Qing [1 ,2 ]
He Ying [3 ]
You Kun [3 ]
Sun Xiao-quan [1 ,2 ]
机构
[1] Natl Univ Def Technol, State Key Lab Pulsed Power Laser Technol, Hefei 230037, Peoples R China
[2] Adv Laser Technol Lab Anhui Prov, Hefei 230037, Peoples R China
[3] Chinese Acad Sci, Hefei Inst Phys Sci, Hefei 230031, Peoples R China
关键词
Laser absorption spectroscopy; TDLAS; Traceability; Uncertainty; Optical path length measurement;
D O I
10.3964/j.issn.1000-0593(2022)11-3461-06
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In the context of carbon neutrality, reliable and accurate quantification of atmospheric greenhouse gas is of great importance. Thus, it is necessary to develop gas analyzers which can provide analytes amount-of-substance fraction results traceable to the international system of units (SI). Tunable diode laser absorption spectroscopy (TDLAS) is a common way to measure gas concentration. According to Beer-Lambert Law, one of the necessary conditions to realize the direct traceability of the measured concentration of the instrument to the SI is the directly traceable gas cell optical path length (OPL). The gas cell OPL directly affects the measurement uncertainty of the gas concentration. Accurate measurement is conducive to developing gas analyzers whose measurement results can be directly traced. In this paper, the directly traceable measurement of a gas cell OPL with three reflection points is carried out. The direct measurement result obtained by measuring three parts of the optical path of the gas cell using a calibrated meter is 81.21 +/- 0.80 cm. The large measurement uncertainty (0. 80 cm) is a careful consideration, estimated considering positioning error and the measurement error caused by the possible misalignment of the three optical paths and the measurement path. In order to reduce the OPL measurement uncertainty, a TDLAS gas cell OPL measurement system is built in this paper. A 1 576 nm distributed feedback laser with a ramp sweep voltage loading on the laser controller is used to measure the absorption spectrum of standard high-purity carbon dioxide (CO2 , 99. 999%) in the gas cell to be measured near 6 344. 68 cm(-1) . A traceable pressure and temperature sensors are used to measure the pressure and gas temperature in the gas cell. The key parameter of line strength of the P 4e branch of CO2 at 30012-00001 transition band, which has relative uncertainty of 0. 15% , is from the National Institute of Standards and Technology. A quadratic speed-dependent Voigt line profile is used to fitting the acquired absorbance at various pressure from 36 to 75 Torr. The uncertainty and the transfer process of various input parameters are comprehensively analyzed. A generalized linear regression (GLR) is applied to absorbance data with uncertainty at various pressure. The slope from the GLR yield the optical path length of the gas cell L = 81.61 +/- 0.42 cm with a standard uncertainty of 0. 51%. This uncertainty range falls within one of the direct measurement results. The optical path structure of the gas cell used in this paper is a simplified multi-reflection gas cell with long OPL. The system can be equally applied to the traceable measurement for OPL of the multi-reflection gas cell.
引用
收藏
页码:3461 / 3466
页数:6
相关论文
共 16 条
[1]  
[Anonymous], NIST Uncertainty Machine
[2]   Speed-dependent Voigt profile for water vapor in infrared remote sensing applications [J].
Boone, Chris D. ;
Walker, Kaley A. ;
Bernath, Peter F. .
JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 2007, 105 (03) :525-532
[3]   Mid-Infrared Tunable Laser-Based Broadband Fingerprint Absorption Spectroscopy for Trace Gas Sensing: A Review [J].
Du, Zhenhui ;
Zhang, Shuai ;
Li, Jinyi ;
Gao, Nan ;
Tong, Kebin .
APPLIED SCIENCES-BASEL, 2019, 9 (02)
[4]   The HITRAN2016 molecular spectroscopic database [J].
Gordon, I. E. ;
Rothman, L. S. ;
Hill, C. ;
Kochanov, R. V. ;
Tan, Y. ;
Bernath, P. F. ;
Birk, M. ;
Boudon, V. ;
Campargue, A. ;
Chance, K. V. ;
Drouin, B. J. ;
Flaud, J. -M. ;
Gamache, R. R. ;
Hodges, J. T. ;
Jacquemart, D. ;
Perevalov, V. I. ;
Perrin, A. ;
Shine, K. P. ;
Smith, M. -A. H. ;
Tennyson, J. ;
Toon, G. C. ;
Tran, H. ;
Tyuterev, V. G. ;
Barbe, A. ;
Csaszar, A. G. ;
Devi, V. M. ;
Furtenbacher, T. ;
Harrison, J. J. ;
Hartmann, J. -M. ;
Jolly, A. ;
Johnson, T. J. ;
Karman, T. ;
Kleiner, I. ;
Kyuberis, A. A. ;
Loos, J. ;
Lyulin, O. M. ;
Massie, S. T. ;
Mikhailenko, S. N. ;
Moazzen-Ahmadi, N. ;
Mueller, H. S. P. ;
Naumenko, O. V. ;
Nikitin, A. V. ;
Polyansky, O. L. ;
Rey, M. ;
Rotger, M. ;
Sharpe, S. W. ;
Sung, K. ;
Starikova, E. ;
Tashkun, S. A. ;
Vander Auwera, J. .
JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 2017, 203 :3-69
[5]   A portable sensor for in-situ measurement of ammonia based on near-infrared laser absorption spectroscopy [J].
Guo, Xinqian ;
Zheng, Fei ;
Li, Chuanliang ;
Yang, Xiaofei ;
Li, Ning ;
Liu, Shuping ;
Wei, Jilin ;
Qiu, Xuanbing ;
He, Qiusheng .
OPTICS AND LASERS IN ENGINEERING, 2019, 115 :243-248
[6]  
Hanson R. K., 2016, SPECTROSCOPY OPTICAL, DOI DOI 10.1007/978-3-319-23252-2
[7]   HITRANonline: An online interface and the flexible representation of spectroscopic data in the HITRAN database [J].
Hill, Christian ;
Gordon, Iouli E. ;
Kochanov, Roman V. ;
Barrett, Lorenzo ;
Wilzewski, Jonas S. ;
Rothman, Laurence S. .
JOURNAL OF QUANTITATIVE SPECTROSCOPY & RADIATIVE TRANSFER, 2016, 177 :4-14
[8]  
InternationalOrganizationforStandardization, CMS RENDER LIVE SITE
[9]  
JCGM, JCGM
[10]   On-line monitoring of odorant in natural gas mixtures of different composition by the infrared absorption spectroscopy method [J].
Kireev, S. V. ;
Shnyrev, S. L. .
LASER PHYSICS LETTERS, 2018, 15 (03)