Failure of a novel silicone-polyurethane copolymer (Optim™) to prevent implantable cardioverter-defibrillator lead insulation abrasions

被引:47
作者
Hauser, Robert G. [1 ]
Abdelhadi, Raed H. [1 ]
McGriff, Deepa M. [1 ]
Retel, Linda Kallinen [1 ]
机构
[1] Minneapolis Heart Inst Fdn, Minneapolis, MN 55407 USA
来源
EUROPACE | 2013年 / 15卷 / 02期
关键词
Implantable cardioverter-defibrillator; Lead; Insulation; Complications; Failure; ICD LEADS;
D O I
10.1093/europace/eus245
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The purpose of this study was to determine if Optim, a unique copolymer of silicone and polyurethane, protects Riata ST Optim and Durata implantable cardioverter-defibrillator (ICD) leads (SJM, St Jude Medical Inc., Sylmar, CA, USA) from abrasions that cause lead failure. We searched the US Food and Drug Administrations (FDAs) Manufacturers and User Device Experience (MAUDE) database on 13 April 2012 using the simple search terms oRiata ST Optim abrasion analysis' and oDurata abrasion analysis'. Lead implant time was estimated by subtracting 3 months from the reported lead age. The MAUDE search returned 15 reports for Riata ST Optim and 37 reports for Durata leads, which were submitted by SJM based on its analyses of returned leads for clinical events that occurred between December 2007 and January 2012. Riata ST Optim leads had been implanted 29.1 11.7 months. Eight of 15 leads had can abrasions and three abrasions were caused by friction with another device, most likely another lead. Four of these abrasions resulted in high-voltage failures and one death. One failure was caused by an internal insulation defect. Durata leads had been implanted 22.2 10.6 months. Twelve Durata leads had can abrasions, and six leads had abrasions caused by friction with another device. Of these 18 can and other device abrasions, 13 (72) had electrical abnormalities. Low impedances identified three internal insulation abrasions. Riata ST Optim and Durata ICD leads have failed due to insulation abrasions. Optim did not prevent these abrasions, which developed 4 years after implant. Studies are needed to determine the incidence of these failures and their clinical implications.
引用
收藏
页码:278 / 283
页数:6
相关论文
共 15 条
  • [1] [Anonymous], 2011, DUR LEAD REL
  • [2] [Anonymous], DUR LEAD REL
  • [3] ICD Leads and Postmarketing Surveillance
    Carlson, Mark D.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2012, 366 (10) : 967 - 967
  • [4] Incidence and implications of abrasion of implantable cardioverter-defibrillator leads
    DeLurgio, DB
    Sathavorn, C
    Mera, F
    Leon, A
    Walter, PF
    Langberg, JJ
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1997, 79 (10) : 1409 - &
  • [5] Performance of the St. Jude Medical Riata Leads
    Epstein, Andrew E.
    Baker, James H.
    Beau, Scott L.
    Deering, Thomas F.
    Greenberg, Steven M.
    Goldman, Daniel S.
    [J]. HEART RHYTHM, 2009, 6 (02) : 204 - 209
  • [6] ICD leads:: Design and chronic dysfunctions
    Gradaus, R
    Breithardt, G
    Böcker, D
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (02): : 649 - 657
  • [7] Deaths caused by the failure of Riata and Riata ST implantable cardioverter-defibrillator leads
    Hauser, Robert G.
    Abdelhadi, Raed
    McGriff, Deepa
    Retel, Linda Kallinen
    [J]. HEART RHYTHM, 2012, 9 (08) : 1227 - 1235
  • [8] Riata implantable cardioverter-defibrillator lead failure: Analysis of explanted leads with a unique insulation defect
    Hauser, Robert G.
    McGriff, Deepa
    Retel, Linda Kallinen
    [J]. HEART RHYTHM, 2012, 9 (05) : 742 - 749
  • [9] Jenney C., 2005, Heart Rhythm, V2, pS318, DOI [10.1016/j.hrthm.2005.02.1004, DOI 10.1016/J.HRTHM.2005.02.1004]
  • [10] Kleeman T, CIRCULATION, V115, P2474