Non-Positive Corrections and Variance Models for Iterative Post-Log Reconstruction of Extremely Low-Dose CT Data

被引:0
|
作者
Kim, Soo Mee [1 ]
Lee, Tzu-Cheng [2 ]
Kinahan, Paul E. [2 ]
机构
[1] Korea Inst Ocean Sci & Technol, ICT R&D Unit, Busan 49111, South Korea
[2] Univ Washington, Dept Radiol, Seattle, WA 98195 USA
基金
美国国家卫生研究院; 新加坡国家研究基金会;
关键词
CT iterative reconstruction; Low-dose CT imaging; Statistical post-log CT model; IMAGE QUALITY; ABDOMINAL CT; REDUCTION; PATIENT;
D O I
10.3938/jkps.77.177
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
In extremely low-dose protocols to reduce radiation dose to patients, computed tomography (CT) images suffer from increased bias and low signal-to-noise ratio in measurements. In this study, we consider three different non-positive corrections, flip, truncation and mean-preserving filter (MPF), affecting the measurement mean, propose a new variance expression for weights in weighted least-squares (WLS) reconstruction, and evaluate the impact on changes in the mean and variance of measurements. We simulated 1000 polychromatic CT sinograms of a chest phantom, including realistic levels of quantum and electronic noises. For the simulated scenario of 80 kVp and 0.5 mAs, compared to the conventional threshold and flip methods, the mean-preserving filter reduced the bias in post-log sinogram values by up to five times. Simple weights in WLS reconstruction that neglected the effect of non-positive correction limited improvements in the image quality. The advanced variance estimates considering electronic noise and the effect of pre-processing on the variance change made both WLS and penalized WLS reconstructions improve. Although the image quality improvement from a WLS reconstruction based on a Gaussian post-log distribution is inherently limited, the proposed method for estimating the post-log variance including electronic noise and the effect of pre-corrections from a single measurement leads to some improvements in variance estimates for post-log CT data and showed the feasibility of post-log iterative reconstruction for extremely low-dose CT imaging.
引用
收藏
页码:177 / 185
页数:9
相关论文
共 50 条
  • [1] Non-Positive Corrections and Variance Models for Iterative Post-Log Reconstruction of Extremely Low-Dose CT Data
    Soo Mee Kim
    Tzu-Cheng Lee
    Paul E. Kinahan
    Journal of the Korean Physical Society, 2020, 77 : 177 - 185
  • [2] Comparison between Pre-log and Post-log Statistical Models in Low-Dose CT Iterative Reconstruction
    Fu, Lin
    Kim, Soo Mee
    Alessio, Adam M.
    Kinahan, Paul E.
    De Man, Bruno
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [3] Analysis of Statistical Models for Iterative Reconstruction of Extremely Low-Dose CT Data
    Kim, Soo Mee
    Alessio, Adam M.
    Perlmutter, David S.
    Thibault, Jean-Baptiste
    De Man, Bruno
    Kinahan, Paul E.
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [4] Comparison Between Pre-Log and Post-Log Statistical Models in Ultra-Low-Dose CT Reconstruction
    Fu, Lin
    Lee, Tzu-Cheng
    Kim, Soo Mee
    Alessio, Adam M.
    Kinahan, Paul E.
    Chang, Zhiqian
    Sauer, Ken
    Kalra, Mannudeep K.
    De Man, Bruno
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2017, 36 (03) : 707 - 720
  • [5] An iterative reconstruction method for sparse-projection data for low-dose CT
    Huang, Ying
    Wan, Qian
    Chen, Zixiang
    Hu, Zhanli
    Cheng, Guanxun
    Qi, Yulong
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2021, 29 (05) : 797 - 812
  • [6] Reduction of CT Exposure for Low-Dose CT Imaging by Iterative CT Reconstruction
    Grosser, O. S.
    Czuczwara, D.
    Wissel, H.
    Laatz, K.
    Ulrich, G.
    Furth, C.
    Ruf, J.
    Amthauer, H.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S501 - S501
  • [7] DIAGNOSTIC PERFORMANCE OF LOW-DOSE NON ENHANCE CT WITH ITERATIVE RECONSTRUCTION IN DIAGNOSIS OF UROLITHIASIS
    Choe, Jung Won
    Do Kim, Kyung
    Moon, Young Tae
    Kwon, Jong Kyou
    Chang, In Ho
    Kim, Jung Noon
    Choi, Jae Duck
    Ahn, Seung Hyun
    Kim, Min Su
    Lee, So Yeon
    Ryu, Jae Hyung
    Park, Sang Ho
    JOURNAL OF UROLOGY, 2014, 191 (04): : E902 - E902
  • [8] Fast iterative adaptive reconstruction in low-dose CT imaging
    Cheng, Lin
    Chen, Yunqiang
    Fang, Tong
    Tyan, Jason
    2006 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING, ICIP 2006, PROCEEDINGS, 2006, : 889 - +
  • [9] Investigation of iterative image reconstruction in low-dose breast CT
    Bian, Junguo
    Yang, Kai
    Boone, John M.
    Han, Xiao
    Sidky, Emil Y.
    Pan, Xiaochuan
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (11): : 2659 - 2685
  • [10] Iterative reconstruction of low-dose CT based on differential sparse
    Lu, Siyu
    Yang, Bo
    Xiao, Ye
    Liu, Shan
    Liu, Mingzhe
    Yin, Lirong
    Zheng, Wenfeng
    BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2023, 79